Optic nerve head analyzer and Heidelberg retina tomograph: relative error and reproducibility of topographic measurements in a model eye with simulated cataract.

P Janknecht1, J Funk.   

Abstract

BACKGROUND: We measured the relative error and reproducibility of the optic nerve head analyzer (ONHA) and the Heidelberg retina tomograph (HRT) in a model eye with a cataract that was simulated by Bangerter foils.
METHODS: There were two artificial discs and one retinal elevation (the latter could not be analyzed by the ONHA) that could be inserted into the model eye. The relative error of the parameter 'cup area' ('cup volume') of the ONHA for the measurement of artificial disc no. 1 changed from 1.1% (5.4%) without Bangerter foil to 7.9% (7.6%) with Bangerter foil 0.6. The standard deviation of the ONHA increased from 0.059 mm2 (0.1 mm3) without Bangerter foil to 0.142 mm2 (0.121 mm3) with Bangerter foil 0.6. With the smaller artificial disc no. 2, no measurements with Bangerter foils were obtained.
RESULTS: Relative error and reproducibility of the parameter 'volume below (above) surface' of the HRT did not show any consistent change with increasing intensity of the simulated cataract. With artificial disc no. 1, the relative error without Bangerter foil was 14.6%, while the worst relative error with one of the Bangerter foils 0.8 to 0.4 was 16.4%. The corresponding values for the standard deviation were 0.019 mm3 and 0.033 mm3. With the smaller artificial disc no. 2, the relative error without Bangerter foil was 6.3%, while the worst relative error with one of the Bangerter foils 0.8 to 0.2 was 18.3%. The corresponding values for the standard deviation were 0.016 mm3 and 0.017 mm3. The relative error in measuring a retinal elevation without a Bangerter foil was 2.3%, with a Bangerter foil 11.2-18.0%. The standard deviation was 0.068 mm3 without Bangerter foil and 0.013-0.023 mm3 with Bangerter foils.
CONCLUSION: Our data support the assumption that the HRT is able to measure fundus structures even in the case of opaque optical media. The HRT is superior to the ONHA in this regard.

Entities:  

Mesh:

Year:  1995        PMID: 8537028     DOI: 10.1007/bf00183434

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  10 in total

1.  Accuracy of topographic measurements in a model eye with the laser tomographic scanner.

Authors:  A W Dreher; R N Weinreb
Journal:  Invest Ophthalmol Vis Sci       Date:  1991-10       Impact factor: 4.799

2.  Reproducibility of topographic measurements of the normal and glaucomatous optic nerve head with the laser tomographic scanner.

Authors:  A W Dreher; P C Tso; R N Weinreb
Journal:  Am J Ophthalmol       Date:  1991-02-15       Impact factor: 5.258

3.  Variability and reproducibility of optic disk topographic measurements with the Rodenstock Optic Nerve Head Analyzer.

Authors:  K I Bishop; E B Werner; T Krupin; D M Kozart; S R Beck; F A Nunan; M B Wax
Journal:  Am J Ophthalmol       Date:  1988-12-15       Impact factor: 5.258

4.  Reproducibility of topographic measurements of the optic nerve head with laser tomographic scanning.

Authors:  F E Kruse; R O Burk; H E Völcker; G Zinser; U Harbarth
Journal:  Ophthalmology       Date:  1989-09       Impact factor: 12.079

5.  Accuracy of topographic measurements with the Optic Nerve Head Analyzer.

Authors:  M B Shields; J S Tiedeman; K N Miller; D Hickingbotham; A R Ollie
Journal:  Am J Ophthalmol       Date:  1989-03-15       Impact factor: 5.258

6.  [Determination of spatial optic disk data with the fundus analyzer].

Authors:  F Dannheim; U Klingbeil
Journal:  Fortschr Ophthalmol       Date:  1986

7.  Reproducibility of topographic measurements with the optic nerve head analyzer.

Authors:  M B Shields; J F Martone; A R Shelton; A R Ollie; J MacMillan
Journal:  Am J Ophthalmol       Date:  1987-12-15       Impact factor: 5.258

8.  Reproducibility of optic disc measurements with computerized analysis of stereoscopic video images.

Authors:  J Caprioli; U Klingbeil; M Sears; B Pope
Journal:  Arch Ophthalmol       Date:  1986-07

9.  Reliability of optic disk topographic measurements recorded with a video-ophthalmograph.

Authors:  F S Mikelberg; G R Douglas; M Schulzer; T N Cornsweet; K Wijsman
Journal:  Am J Ophthalmol       Date:  1984-07-15       Impact factor: 5.258

10.  [Improved reproducibility of computer-assisted structural analysis of the optic papilla].

Authors:  J Funk; R Steeb
Journal:  Klin Monbl Augenheilkd       Date:  1991-06       Impact factor: 0.700

  10 in total
  6 in total

1.  Optic nerve head morphometry in healthy adults using confocal laser scanning tomography.

Authors:  M M Hermann; I Theofylaktopoulos; N Bangard; C Jonescu-Cuypers; S Coburger; M Diestelhorst
Journal:  Br J Ophthalmol       Date:  2004-06       Impact factor: 4.638

Review 2.  Test-retest variability in structural parameters measured with glaucoma imaging devices.

Authors:  Makoto Araie
Journal:  Jpn J Ophthalmol       Date:  2012-11-09       Impact factor: 2.447

3.  The Effect of Bangerter Filters on Visual Acuity and Contrast Sensitivity With External Noise.

Authors:  Pan Zhang; Hanlin Wang; Weicong Ren; Huanhuan Guo; Jiayi Yang; Jiayu Tao; Zhijie Yang; Ying Li; Lijun Chen; Yajing Zhang; Di Wu
Journal:  Front Neurosci       Date:  2022-05-13       Impact factor: 5.152

4.  The effect of Bangerter filters on optotype acuity, Vernier acuity, and contrast sensitivity.

Authors:  Naomi V Odell; David A Leske; Sarah R Hatt; Wendy E Adams; Jonathan M Holmes
Journal:  J AAPOS       Date:  2008-08-15       Impact factor: 1.220

5.  Alterations induced by bangerter filters on the visual field: a frequency doubling technology and standard automated perimetry study.

Authors:  Costantino Schiavi; Filippo Tassi; Alessandro Finzi; Mauro Cellini
Journal:  J Ophthalmol       Date:  2015-01-20       Impact factor: 1.909

6.  Relationship between pattern electroretinogram and optic disc morphology in glaucoma.

Authors:  Soo Ji Jeon; Hae-Young Lopilly Park; Kyoung In Jung; Chan Kee Park
Journal:  PLoS One       Date:  2019-11-07       Impact factor: 3.240

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.