INTRODUCTION: The detection of possible root canal perforations caused during a metal post placement is frequently difficult to diagnose. The aim of this study was to investigate the accuracy of apex locators to diagnose such perforations.Entities:
Mesh:
Year: 2014 PMID: 25533476 PMCID: PMC4311483 DOI: 10.1186/s13005-014-0057-2
Source DB: PubMed Journal: Head Face Med ISSN: 1746-160X Impact factor: 2.151
Figure 1Fixation device showing the root tooth placed in a plastic tube in an agar-agar solution and with both apex locator electrodes (left) and a maxillary molar with a perforated root canal through a metal post (right).
All methods excluded perforations when they were not present (100% specificity; 95% confidence interval [78%; 100%])
|
|
|
|
|---|---|---|
| ProPex II | 93% [68%;100%] | 100% [78%; 100%] |
| Elements Apex Locator | 93% [68%;100%] | 100% [78%; 100%] |
| Apex NRG | 100% [78%; 100%] | 100% [78%; 100%] |
| RayPex 5 | 100% [78%; 100%] | 100% [78%; 100%] |
| RayPex 6 | 93% [68%;100%] | 100% [78%; 100%] |
The Apex NRG and RayPex 5 also detected all perforations (100% sensitivity; 95% confidence interval [78%; 100%]). The ProPex II, Elements Apex Locator, and RayPex 6 detected 14 of 15 perforations (93% sensitivity; 95% confidence interval [68%;100%]).
Results of the inter-device agreement which proved to be high
|
|
|
|
|---|---|---|
| Pro Pex II | Elements Apex Locator | 0.86 [0.69; 1.00] |
| Pro Pex II | Apex NRG | 0.93 [0.81; 1.00] |
| Pro Pex II | RayPex 5 | 0.93 [0.81; 1.00] |
| Pro Pex II | RayPex 6 | 0.86 [0.69; 1.00] |
| Elements Apex Locator | Apex NRG | 0.93 [0.81; 1.00] |
| Elements Apex Locator | RayPex 5 | 0.93 [0.81; 1.00] |
| Elements Apex Locator | RayPex 6 | 0.86 [0.69; 1.00] |
| Apex NRG | RayPex 5 | 1.00 |
| Apex NRG | RayPex 6 | 0.93 [0.81; 1.00] |
| RayPex 5 | Raypex 6 | 0.93 [0.81; 1.00] |