Literature DB >> 8376421

Complications related to modularity of total hip components.

R L Barrack1, D W Burke, S D Cook, H B Skinner, W H Harris.   

Abstract

We report complications from the use of modular components in 20 hip replacements in 18 patients. Fifteen complications (in 13 patients) were related to failure of a modular interface after operation. Femoral head detachment from its trunnion was seen in 6 hips from trauma (3), reduction of a dislocation (2), and normal activity (1). In one case the base of the trunnion fractured below an extra-long modular head. In seven other hips the modular polyethylene liner dislodged from its shell, causing severe damage to the shell in four cases with extensive metallosis. In one other hip an asymmetrical polyethylene liner rotated, resulting in impingement of the femoral component and recurrent dislocation. Operative errors were seen in five cases: implantation of a trial acetabular component in one; and mismatching between the size of the femoral head and the acetabular component in the others. Surgeons who use hip replacements with modular components should be aware of the potential for operative error and of the importance of early treatment for postoperative mechanical failure.

Entities:  

Mesh:

Year:  1993        PMID: 8376421

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  19 in total

1.  Stem-Sleeve Junction Failure of a Modular Femoral Hip System: a Retrieval Analysis.

Authors:  Feras Waly; Fahad H Abduljabbar; Trevor Gascoyne; Thomas R Turgeon; Olga Huk
Journal:  HSS J       Date:  2015-07-09

Review 2.  Total hip arthroplasty instability in Italy.

Authors:  Francesco Falez; Matteo Papalia; Fabio Favetti; Gabriele Panegrossi; Filippo Casella; Gianluca Mazzotta
Journal:  Int Orthop       Date:  2016-12-20       Impact factor: 3.075

3.  Are titanium-on-titanium TiAl6V4 modular necks safe in total hip arthroplasty for non-overweight patients? Results of a prospective series at a minimum follow-up of 7 years.

Authors:  Matthieu Ollivier; Sébastien Parratte; Alexandre Galland; Alexandre Lunebourg; Jean-Noel Argenson
Journal:  Eur J Orthop Surg Traumatol       Date:  2015-07-10

4.  A stature-specific concept for uncemented, primary total hip arthroplasty.

Authors:  Georg W Omlor; Hannah Ullrich; Knut Krahmer; Alexander Jung; Günther Aldinger; Peter Aldinger
Journal:  Acta Orthop       Date:  2010-02       Impact factor: 3.717

5.  Clinical biomechanics of wear in total hip arthroplasty.

Authors:  John J Callaghan; Douglas R Pedersen; Richard C Johnston; Thomas D Brown
Journal:  Iowa Orthop J       Date:  2003

6.  [Titanium modular revision prosthesis stem in revision hip prosthesis].

Authors:  A Schuh; U Holzwarth; G Zeiler
Journal:  Orthopade       Date:  2004-01       Impact factor: 1.087

7.  Fixation strength of taper connection at head-neck junction in retrieved carbon fiber-reinforced PEEK hip stems.

Authors:  Ichiro Nakahara; Masaki Takao; Shunichi Bandoh; Nobuhiko Sugano
Journal:  J Artif Organs       Date:  2014-09-05       Impact factor: 1.731

8.  Stem modularity alone is not effective in reducing dislocation rate in hip revision surgery.

Authors:  Dario Regis; Andrea Sandri; Pietro Bartolozzi
Journal:  J Orthop Traumatol       Date:  2009-11-18

9.  Fracture of the Modular Neck in Total Hip Arthroplasty.

Authors:  A Hernandez; A Gargallo-Margarit; V Barro; I Gallardo-Calero; A Sallent
Journal:  Case Rep Orthop       Date:  2015-07-22

10.  Spontaneous Dissociation of Anatomic Medullary Locking A Plus (AML A Plus) Femoral Component at the Head-Neck Interface.

Authors:  Ketan Pande; Juzaily Fekry Leong; Ngai Nung Lo
Journal:  J Orthop Case Rep       Date:  2015 Jul-Sep
View more