Literature DB >> 26160760

Are titanium-on-titanium TiAl6V4 modular necks safe in total hip arthroplasty for non-overweight patients? Results of a prospective series at a minimum follow-up of 7 years.

Matthieu Ollivier1,2, Sébastien Parratte3,4, Alexandre Galland1,2, Alexandre Lunebourg1,2, Jean-Noel Argenson1,2.   

Abstract

Using extramedullar modularity in total hip arthroplasty has been proposed as an option to optimize the restoration of hip biomechanics. To avoid the problems that were observed with cobalt-chrome modular neck, titanium modular necks have been developed. The goals of our study were to evaluate the safety of titanium-on-titanium TiAl6V4 modular neck system. Hundred patients with a mean age of 69.6 ± 10.6 (42-86 years) and mean BMI of 25.07 ± 4.86 (17-38 kg/m(2)) suffering from primary or secondary arthritis of the hip were prospectively included. At a minimum of 7 years, no fracture of the modular neck was observed. No patient required a revision. No sign of loosening was found in the radiological analysis. Our study shows that titanium-on-titanium TiAl6V4 modular neck system can safely be used with good midterm clinical and radiological results for non-overweight patients. These results should be confirmed at longer follow-up.

Entities:  

Keywords:  Cementless; Modular; Survival; THA; Titanium neck

Mesh:

Substances:

Year:  2015        PMID: 26160760     DOI: 10.1007/s00590-015-1665-7

Source DB:  PubMed          Journal:  Eur J Orthop Surg Traumatol        ISSN: 1633-8065


  28 in total

1.  Mechanisms of failure of modular prostheses.

Authors:  J P Collier; M B Mayor; R E Jensen; V A Surprenant; H P Surprenant; J L McNamar; L Belec
Journal:  Clin Orthop Relat Res       Date:  1992-12       Impact factor: 4.176

Review 2.  Modular revision stems in total hip arthroplasty.

Authors:  Richard E Jones
Journal:  Clin Orthop Relat Res       Date:  2004-03       Impact factor: 4.176

Review 3.  Modularity of the femoral component in total hip arthroplasty.

Authors:  Anand Srinivasan; Edward Jung; Brett Russell Levine
Journal:  J Am Acad Orthop Surg       Date:  2012-04       Impact factor: 3.020

4.  "Modes of failure" of cemented stem-type femoral components: a radiographic analysis of loosening.

Authors:  T A Gruen; G M McNeice; H C Amstutz
Journal:  Clin Orthop Relat Res       Date:  1979-06       Impact factor: 4.176

5.  Proximal component modularity in THA--at what cost? An implant retrieval study.

Authors:  A M Kop; C Keogh; E Swarts
Journal:  Clin Orthop Relat Res       Date:  2012-07       Impact factor: 4.176

6.  The John Charnley Award: Metal-on-metal hip resurfacing versus large-diameter head metal-on-metal total hip arthroplasty: a randomized clinical trial.

Authors:  Donald S Garbuz; Michael Tanzer; Nelson V Greidanus; Bassam A Masri; Clive P Duncan
Journal:  Clin Orthop Relat Res       Date:  2009-08-21       Impact factor: 4.176

7.  Anatomical ceramic arthroplasty (AN.C.A.): preliminary experience with a new cementless prosthesis.

Authors:  A Toni; A Sudanese; D Ciaroni; D Dallari; T Greggi; A Giunti
Journal:  Chir Organi Mov       Date:  1990 Jan-Mar

8.  Modular titanium alloy neck adapter failures in hip replacement--failure mode analysis and influence of implant material.

Authors:  Thomas M Grupp; Thomas Weik; Wilhelm Bloemer; Hanns-Peter Knaebel
Journal:  BMC Musculoskelet Disord       Date:  2010-01-04       Impact factor: 2.362

9.  Feasibility of 4 patient-reported outcome measures in a registry setting.

Authors:  Aksel Paulsen; Alma B Pedersen; Søren Overgaard; Ewa M Roos
Journal:  Acta Orthop       Date:  2012-08-20       Impact factor: 3.717

10.  Hip disability and osteoarthritis outcome score (HOOS)--validity and responsiveness in total hip replacement.

Authors:  Anna K Nilsdotter; L Stefan Lohmander; Maria Klässbo; Ewa M Roos
Journal:  BMC Musculoskelet Disord       Date:  2003-05-30       Impact factor: 2.362

View more
  2 in total

1.  Femoral neck preservation with a short hip stem produced with powder manufacturing: mid-term results of a consecutive case series.

Authors:  Marco Schiraldi; Manuel Bondi; Lodovico Renzi Brivio
Journal:  Eur J Orthop Surg Traumatol       Date:  2019-01-28

2.  Exchangeable Femoral Neck (Dual-Modular) THA Prostheses Have Poorer Survivorship Than Other Designs: A Nationwide Cohort of 324,108 Patients.

Authors:  Sandrine Colas; Assia Allalou; Antoine Poichotte; Philippe Piriou; Rosemary Dray-Spira; Mahmoud Zureik
Journal:  Clin Orthop Relat Res       Date:  2017-02-13       Impact factor: 4.176

  2 in total