Literature DB >> 7706513

Effect of thromboplastin and coagulometer interaction on the precision of the International Normalised Ratio.

D W Pi1, J M Raboud, C Filby, C J Carter.   

Abstract

AIMS: To examine the magnitude of thromboplastin and coagulometer interactions on the precision of International Normalised Ratio (INR) values when the manufacturers' recommended instrument specific International Sensitivity Index (ISI) values are adopted for the INR calculation.
METHODS: The variability of INR values obtained from four automated phototopical coagulometers frequently used in North American laboratories was studied. When used with five commercial thromboplastins of moderate to high sensitivity (ISI values 0.92-1.97), 20 prothrombin time results were generated for each of a population of 98 patients on established warfarin treatment.
RESULTS: The mean INR values of the patients ranged from 2.05 to 2.81, depending on which reagent/coagulometer combination was used. Within patient variation increased as the median INR value increased. The mean coefficient of variation of within patient INR values was 10%; the mean coefficient of variation of the prothrombin time results in seconds and prothrombin time ratio were 21 and 18%, respectively.
CONCLUSIONS: There was considerable bias in the estimated ISI values of the thromboplastins compared with the manufacturers' instrument specific ISI value. Despite this apparent imperfection, our study clearly showed that the INR is preferable to other prothrombin time reporting formats for assessing the degree of anticoagulation for patients on warfarin treatment.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7706513      PMCID: PMC502253          DOI: 10.1136/jcp.48.1.13

Source DB:  PubMed          Journal:  J Clin Pathol        ISSN: 0021-9746            Impact factor:   3.411


  12 in total

1.  Assessment of value of calibrated lyophilised plasmas to determine International Sensitivity Index for coagulometers.

Authors:  K Clarke; D A Taberner; J M Thomson; J A Morris; L Poller
Journal:  J Clin Pathol       Date:  1992-01       Impact factor: 3.411

2.  Determination of the mean normal prothrombin time for assessment of international normalized ratios. Usefulness of lyophilized plasma.

Authors:  R H Peters; A M van den Besselaar; F M Olthuis
Journal:  Thromb Haemost       Date:  1991-10-01       Impact factor: 5.249

3.  The dependence of the International Sensitivity Index on the coagulometer used to perform the prothrombin time.

Authors:  M J Ray; I R Smith
Journal:  Thromb Haemost       Date:  1990-06-28       Impact factor: 5.249

4.  Automation and prothrombin time: a United Kingdom field study of two widely used coagulometers.

Authors:  J M Thomson; D A Taberner; L Poller
Journal:  J Clin Pathol       Date:  1990-08       Impact factor: 3.411

5.  Survey of prothrombin time in National External Quality Assessment Scheme exercises (1980-87).

Authors:  L Poller; D A Taberner; J M Thomson; K V Darby
Journal:  J Clin Pathol       Date:  1988-04       Impact factor: 3.411

6.  Effect of international sensitivity index (ISI) of thromboplastins on precision of international normalised ratios (INR)

Authors:  D A Taberner; L Poller; J M Thomson; K V Darby
Journal:  J Clin Pathol       Date:  1989-01       Impact factor: 3.411

7.  Failure of the International Normalized Ratio to generate consistent results within a local medical community.

Authors:  V L Ng; J Levin; L Corash; E L Gottfried
Journal:  Am J Clin Pathol       Date:  1993-06       Impact factor: 2.493

8.  Effect of the choice of WHO International Reference Preparation for thromboplastin on International Normalised Ratios.

Authors:  L Poller; D A Taberner; J M Thomson; J Morris; R S Mibashan; N K Shinton
Journal:  J Clin Pathol       Date:  1993-01       Impact factor: 3.411

9.  Effect of automation on prothrombin time test in NEQAS surveys.

Authors:  L Poller; J M Thomson; D A Taberner
Journal:  J Clin Pathol       Date:  1989-01       Impact factor: 3.411

10.  Imprecision of prothrombin time monitoring of oral anticoagulation. A survey of hospital laboratories.

Authors:  J E Ansell
Journal:  Am J Clin Pathol       Date:  1992-08       Impact factor: 2.493

View more
  3 in total

1.  Accuracy and precision of the TAS analyser for near-patient INR testing by non-pathology staff in the community.

Authors:  P G Cachia; E McGregor; S Adlakha; P Davey; B M Goudie
Journal:  J Clin Pathol       Date:  1998-01       Impact factor: 3.411

Review 2.  Clinical utilization of the international normalized ratio (INR).

Authors:  R S Riley; D Rowe; L M Fisher
Journal:  J Clin Lab Anal       Date:  2000       Impact factor: 2.352

3.  Adaptive force sonorheometry for assessment of whole blood coagulation.

Authors:  F William Mauldin; Francesco Viola; Theresa C Hamer; Eman M Ahmed; Shawna B Crawford; Doris M Haverstick; Michael B Lawrence; William F Walker
Journal:  Clin Chim Acta       Date:  2010-01-22       Impact factor: 3.786

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.