Accuracy and precision of the TAS analyser for near-patient INR testing by non-pathology staff in the community.

P G Cachia1, E McGregor, S Adlakha, P Davey, B M Goudie.   

Abstract

AIMS: To assess the accuracy and precision of INR measurement by trained practice and district nursing staff using the Thrombolytic Assessment System (TAS) analyser.
METHODS: Seventeen nurses from four practices were trained to measure INR using the TAS analyser on citrated capillary blood samples. Quality control (QC) consisted of: daily internal QC using normal and abnormal commercial plasmas; monthly local external QC scheme using fresh citrated venous blood; and registration of all analysers in the NEQAS (national external quality assessment scheme) main users scheme.
RESULTS: Analysis of internal QC results demonstrated satisfactory interanalyser and intra-analyser precision with no evidence of analytical drift in any of the four practice analysers over an eight month period. Local and national external QC results confirmed the interanalyser precision but INR was underestimated by the TAS analysers compared with the CA 1000 using either Diagen rabbit brain thromboplastin or Innovin, and with other NEQAS users.
CONCLUSIONS: The TAS analyser has many features to commend it for use by nonpathology staff to determine INR. Local internal and external QC and entry into the NEQAS main users group are possible because the TAS analyses citrated plasma or blood. The TAS analyser underestimates INR when the geometric mean normal prothrombin time (GMNPT) is determined by conventional methods. A local correction factor can be introduced by adjusting the normal PT to give INR results comparable with the local laboratory. This is particularly desirable when INRs are measured using both near-patient and laboratory analytical systems on different occasions.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9577376      PMCID: PMC500435          DOI: 10.1136/jcp.51.1.68

Source DB:  PubMed          Journal:  J Clin Pathol        ISSN: 0021-9746            Impact factor:   3.411


  6 in total

1.  Evaluation of the Ciba Corning Biotrack 512 coagulation monitor for the control of oral anticoagulation.

Authors:  I Jennings; R J Luddington; T Baglin
Journal:  J Clin Pathol       Date:  1991-11       Impact factor: 3.411

Review 2.  ABC of atrial fibrillation. Antithrombotic treatment for atrial fibrillation.

Authors:  G Y Lip; G D Lowe
Journal:  BMJ       Date:  1996-01-06

3.  The value of plasma calibrants in correcting coagulometer effects on international normalized ratios. An international multicenter study.

Authors:  L Poller; D A Triplett; J Hirsh; J Carroll; K Clarke
Journal:  Am J Clin Pathol       Date:  1995-03       Impact factor: 2.493

4.  Effect of thromboplastin and coagulometer interaction on the precision of the International Normalised Ratio.

Authors:  D W Pi; J M Raboud; C Filby; C J Carter
Journal:  J Clin Pathol       Date:  1995-01       Impact factor: 3.411

5.  Cost-effectiveness of warfarin and aspirin for prophylaxis of stroke in patients with nonvalvular atrial fibrillation.

Authors:  B F Gage; A B Cardinalli; G W Albers; D K Owens
Journal:  JAMA       Date:  1995-12-20       Impact factor: 56.272

6.  Thromboplastin related differences in the determination of international normalised ratio: a cause for concern? Steering Committee of the UK National External Quality Assessment Scheme in Blood Coagulation.

Authors:  S Kitchen; I D Walker; T A Woods; F E Preston
Journal:  Thromb Haemost       Date:  1994-09       Impact factor: 5.249

  6 in total
  4 in total

Review 1.  Oral anticoagulant therapy: Antithrombotic Therapy and Prevention of Thrombosis, 9th ed: American College of Chest Physicians Evidence-Based Clinical Practice Guidelines.

Authors:  Walter Ageno; Alexander S Gallus; Ann Wittkowsky; Mark Crowther; Elaine M Hylek; Gualtiero Palareti
Journal:  Chest       Date:  2012-02       Impact factor: 9.410

Review 2.  Accuracy, precision, and quality control for point-of-care testing of oral anticoagulation.

Authors:  A M van den Besselaar
Journal:  J Thromb Thrombolysis       Date:  2001-09       Impact factor: 2.300

Review 3.  Clinical utilization of the international normalized ratio (INR).

Authors:  R S Riley; D Rowe; L M Fisher
Journal:  J Clin Lab Anal       Date:  2000       Impact factor: 2.352

4.  Reliability of international normalised ratios from two point of care test systems: comparison with conventional methods.

Authors:  Leon Poller; Michelle Keown; Nikhil Chauhan; Anton M H P van den Besselaar; Armando Tripodi; Caroline Shiach; Jorgen Jespersen
Journal:  BMJ       Date:  2003-07-05
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.