| Literature DB >> 7500892 |
Abstract
The principle of active acoustic screening, recently introduced by Mansfield et al. (1) to reduce acoustic noise in gradient coils for MRI, is applied to the design of transverse distributed gradient coils of the fingerprint design. The results indicate that for complete Lorentz force balancing and for exact satisfaction of Kirchoff's law, it is necessary to have a coil system wound on four cylinders if full active magnetic screening of the assembly is required.Mesh:
Year: 1995 PMID: 7500892 DOI: 10.1002/mrm.1910340331
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668