Non-invasive oscillometric measurement of systolic, mean and diastolic blood pressure in infants with congenital heart defects after operation. A comparison with direct blood pressure measurements.

V Pilossoff, J G Schöber, D Peters, K Bühlmeyer.   

Abstract

The accuracy of indirect oscillometric blood pressure measurements has been evaluated in 32 infants up to 11 months of age undergoing heart surgery. In a number of 1029 simultaneous measurements the indirect blood pressure was compared with the direct value obtained from a radial artery catheter. Cuffs of different sizes were applied. The main results were as follows: Regarding the ratio of cuff width to arm circumference, the best correlation between oscillometric and direct blood pressure measurements was obtained with a ratio of 0.38-0.41. The value of indirect blood pressure measurements depends on the absolute height of the blood pressure. In low blood pressure there is a tendency to underestimate and in high blood pressure there is a tendency to overestimate by the oscillometric blood pressure measurement. By applying an appropriate cuff size and by using our diagrams it should be possible to derive a direct value for the blood pressure on the basis of indirect oscillometric blood pressure measurements.

Entities:  

Mesh:

Year:  1985        PMID: 4076248     DOI: 10.1007/BF00441772

Source DB:  PubMed          Journal:  Eur J Pediatr        ISSN: 0340-6199            Impact factor:   3.183


  23 in total

1.  A study of instruments in preparation for a blood pressure survey of children.

Authors:  L S Webber; A W Voors; T A Foster; G S Berenson
Journal:  Circulation       Date:  1977-10       Impact factor: 29.690

2.  Indirect blood pressure measurement in newborns.

Authors:  H S Dweck; D W Reynolds; G Cassady
Journal:  Am J Dis Child       Date:  1974-04

3.  Indirect measurement of blood pressure in neonates and infants utilizing an automatic noninvasive oscillometric monitor.

Authors:  R H Friesen; J L Lichtor
Journal:  Anesth Analg       Date:  1981-10       Impact factor: 5.108

4.  An assessment of the Dinamap 845.

Authors:  P Hutton; J Dye; C Prys-Roberts
Journal:  Anaesthesia       Date:  1984-03       Impact factor: 6.955

Review 5.  Hypertension in childhood: a review.

Authors:  L G McLain
Journal:  Am Heart J       Date:  1976-11       Impact factor: 4.749

6.  Hazard of automatic noninvasive blood pressure monitoring.

Authors:  A Showman; E K Betts
Journal:  Anesthesiology       Date:  1981-12       Impact factor: 7.892

7.  Clinical evaluation of Dinamap 845 automated blood pressure recorder.

Authors:  J H Silas; A T Barker; L E Ramsay
Journal:  Br Heart J       Date:  1980-02

8.  [Simultaneous measurement of indirect (Doppler-ultrasonic-technique) and direct arterial systolic blood pressure of the human radial artery (author's transl)].

Authors:  M Köhler; B Lösse
Journal:  Z Kardiol       Date:  1979-08

9.  Noninvasive determination of systolic, diastolic and end-systolic blood pressure in neonates, infants and young children: comparison with central aortic pressure measurements.

Authors:  S D Colan; A Fujii; K M Borow; D MacPherson; S P Sanders
Journal:  Am J Cardiol       Date:  1983-10-01       Impact factor: 2.778

10.  Evaluating indirect blood pressure measurement techniques: a comparison of three systems in infants and children.

Authors:  R F Reder; I Dimich; M L Cohen; L Steinfeld
Journal:  Pediatrics       Date:  1978-09       Impact factor: 7.124

View more
  4 in total

1.  Captopril in children with dilated cardiomyopathy: acute and long-term effects in a prospective study of hemodynamic and hormonal effects.

Authors:  H Stern; J Weil; T Genz; W Vogt; K Bühlmeyer
Journal:  Pediatr Cardiol       Date:  1990-01       Impact factor: 1.655

2.  Systolic blood pressure in babies of less than 32 weeks gestation in the first year of life. Northern Neonatal Nursing Initiative.

Authors: 
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  1999-01       Impact factor: 5.747

3.  Use of pulse oximetry for blood pressure measurement after cardiac surgery.

Authors:  A J Movius; S L Bratton; G K Sorensen
Journal:  Arch Dis Child       Date:  1998-05       Impact factor: 3.791

4.  Pulse oximetry vs non-invasive blood pressure/oscillometry to record blood pressure in neonates: A prospective observational study.

Authors:  Reenu Raju; Mathai Isac
Journal:  J Family Med Prim Care       Date:  2022-06-30
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.