Abstract
Background: Computed tomography (CT) contributes to 60% of the collective dose in medical imaging. Literature has demonstrated that patient dose varies across regions and countries. Establishing diagnostic reference levels (DRLs) contributes to the optimization of clinical practices and radiation protection. Purpose: To survey the dose indices (CTDIvol and dose-length product) for frequently performed CT examinations from the chosen hospitals in Norway and Canada and to determine local DRLs (LDRLs) based on the collected data. Material andEntities:
Keywords: CT dosimetry; diagnostic reference levels
Year: 2022 PMID: 36225897 PMCID: PMC9549116 DOI: 10.1177/20584601221131477
Source DB: PubMed Journal: Acta Radiol Open
Characteristics of the CT scanners included in the survey.
| Hospital | Manufacturer | Model | Install year | Detector
configuration[ | Iterative reconstruction |
|---|---|---|---|---|---|
| Norway | |||||
| 1 | Siemens | DSSomatom force | 2019 | 2 · 96 × 0.6 mm | ADMIRE |
| Siemens | DSSomatom Force | 2017 | 2 · 96 × 0.6 mm | ADMIRE | |
| Canon | Aquilion One | 2015 | 320 × 0.5 mm | AIDR | |
| 2 | Siemens | DSSomatom Force | 2020 | 2 · 96 × 0.6 mm | ADMIRE |
| Siemens | DSSomatom Drive | 2016 | 2 · 64 × 0.6 mm | ADMIRE | |
| Siemens | DSDefinition Flash | 2012 | 2 · 64 × 0.6 mm | SAFIRE | |
| GE | Revolution | 2018 | 256 × 0.625 mm | ASiR-V, DLIR | |
| GE | Revolution Apex | 2018 | 256 × 0.625 mm | ASiR-V, DLIR | |
| Canada | |||||
| 3 | Siemens | Definition Edge | 2018 | 2 · 64 × 0.6 mm | SAFIRE |
| 4 | Siemens | DSDefinition Flash | 2011 | 2 · 64 × 0.6 mm | SAFIRE |
| 5 | GE | Optima CT660 | 2013 | 64 × 0.625 | ASiR |
| 6 | GE | Optima CT660 | 2012 | 64 × 0.625 | ASiR |
Superscript DS indicates a dual-source.
aNotation of the detector configuration: the number of active detector rows x the detector element thickness; an additional factor of 2 indicates that a z-flying focal spot is used to double the number of slices acquired in one rotation.
Acquisition parameters for CT examinations on each scanner.
| Hospital | Scanner | Tube voltage | AEC settings | Slice (mm) | Pitch | Rotation time (s) |
|---|---|---|---|---|---|---|
| Head | ||||||
| 1 | Somatom Force2019 | 120 Ref kV, Care kV | 450 Ref mAs | 0.6 | 0.55 | 1 |
| Somatom Force2017 | ||||||
| [ | 120 kV | Fixed 410 mA | 0.5 | – | 1 | |
| 2 | Somatom force2020 | 120 kV | 390 Ref mAs | 0.6 | 0.55 | 1 |
| Somatom Drive2016 | ||||||
| Definition Flash2012 | ||||||
| Revolution2018 | 120 kV | min/max: 100/340 mA, NI = 6 | 0.625 | 0.516 | 0.6 | |
| Revolution Apex2018 | ||||||
| 3 | Definition Edge2018 | 120 Ref kV, Care kV | 320 Ref mAs | 0.6 | 0.55 | 1 |
| 4 | Definition Flash2011 | 120 Ref kV, Care kV | 350 Ref mAs | 0.6 | 0.55 | 1 |
| 5 | Optima CT6602013 | 120 kV | Fixed 380 mA | 0.625 | 0.516 | 0.6 |
| 6 | Optima CT6602012 | 120 kV | Fixed 200 mA | 0.625 | 0.531 | 1 |
| Thorax | ||||||
| 1 | Somatom Force2019 | 120 Ref kV, Care kV | 130 Ref mAs | 0.6 | 2.5 | 0.25 |
| Somatom Force2017 | ||||||
| [ | 120 kV | min/max: 100/750 mA, SD = 10 | 0.5 | 0.813 | 0.275 | |
| 2 | Somatom Force2020 | 120 Ref kV, Care kV | 120 Ref mAs | 0.6 | 0.6 | 0.5 |
| Somatom Drive2016 | ||||||
| Definition Flash2012 | ||||||
| Revolution2018 | 80–120 kV, kV Assist | min/max: 80/500 mA, NI = 32 | 0.625 | 0.508 | 0.5 | |
| Revolution Apex2018 | ||||||
| 3 | Definition Edge2018 | 120 Ref kV, Care kV | 110 Ref mAs | 0.6 | 0.6 | 0.33 |
| 4 | Definition Flash2011 | 120 Ref kV, Care kV | 70 Ref mAs | 0.6 | 0.6 | 0.5 |
| 5 | Optima CT6602013 | 120 kV | min/max: 50/500 mA, NI = 13.88 | 0.625 | 1.35 | 0.8 |
| 6 | Optima CT6602012 | 120 kV | min/max:100/500 mA, NI = 26.00 | 0.625 | 0.984 | 0.6 |
| Abdomen and pelvis | ||||||
| 1 | Somatom Force2019 | Tube A: 90 kV | 220 Ref mAs | 0.6 | 0.6 | 0.5 |
| Somatom Force2017 | ||||||
| Aquilion One2015 | Auto kV | min/max: 200/900 mA, SD = 7.5 | 0.5 | 0.637 | 0.5 | |
| 2 | Somatom Force2020 | 120 Ref kV | 210 Ref mAs | 0.6 | 0.6 | 0.5 |
| Somatom Drive2016 | ||||||
| Definition Flash2012 | ||||||
| Revolution2018 | 100–120 kV, kV Assist | min/max: 150/900 mA, NI = 29 | 0.625 | 0.508 | 0.5 | |
| Revolution Apex2018 | ||||||
| 3 | Definition Edge2018 | 120 Ref kV, Care kV | 210 Ref mAs | 0.6 | 0.6 | 0.5 |
| 4 | Definition Flash2011 | 120 Ref kV, Care kV | 185 Ref mAs | 0.6 | 0.6 | 0.5 |
| 5 | Optima CT6602013 | 120 kV | min/max: 50/560 mA, NI = 12.31 | 0.625 | 1.375 | 0.8 |
| 6 | Optima CT6602012 | 120 kV | min/max:100/500 mA, NI = 24.00 | 0.625 | 0.984 | 0.8 |
aDynamic volume scan.
Results of statistical analysis showing the min/max range, the mean, and standard deviation for the CTDIvol and DLP values for all examinations on each scanner.
| Norway | |||||||
|---|---|---|---|---|---|---|---|
| CT scanner | Examination | Mean AP thickness (cm) | Mean weight (kg) | Min/max DLP (mGy·cm) | Mean DLP (mGy·cm) | Min/max CTDIvol (mGy) | Mean CTDIvol (mGy) |
| Somatom Force2019 | Head | — | — | 790.2–1233.8 | 995.7 ± 110.5 | 29.4–65.5 | 53.04 ± 7.2 |
| Thorax | 22.66 ± 1.39 | — | 142.1–402.8 | 226.0 ± 61.5 | 4.2–9.2 | 6.18 ± 1.12 | |
| Abdomen and pelvis | — | 71.30 ± 5.91 | 227.1–1206.3 | 519.6 ± 200.0 | 7.3–22.3 | 10.04 ± 3.30 | |
| Somatom Force2017 | Head | — | — | 671.8–1187.8 | 959.2 ± 130.6 | 43.3–67.7 | 53.06 ± 4.97 |
| Thorax | 22.83 ± 1.95 | — | 112.6–300.0 | 220.3 ± 46.4 | 3.1–9.7 | 6.43 ± 1.17 | |
| Abdomen and pelvis | — | 73.27 ± 9.88 | 294.2–1008.2 | 495.0 ± 157.8 | 7.1–17.9 | 9.94 ± 2.50 | |
| Somatom Force2020 | Head | — | — | 550.6–899.2 | 733.3 ± 98.7 | 34.0–45.2 | 40.06 ± 3.53 |
| Thorax | 24.01 ± 2.79 | — | 168.9–718.1 | 327.0 ± 152.8 | 4.1–17.9 | 8.79 ± 3.74 | |
| Abdomen and pelvis | — | 67.75 ± 9.67 | 368.7–1293.2 | 708.5 ± 302.7 | 7.7–24.8 | 13.91 ± 5.30 | |
| Definition Flash2012 | Head | — | — | 741.4–1214.5 | 939.6 ± 138.7 | 45.3–63.1 | 52.95 ± 5.36 |
| Thorax | 23.24 ± 2.03 | — | 219.0–574.7 | 414.3 ± 110.9 | 5.8–19.7 | 12.52 ± 3.58 | |
| Abdomen and pelvis | — | 70.75 ± 9.22 | 476.0–912.3 | 687.4 ± 148.7 | 10.4–18.5 | 13.81 ± 2.69 | |
| Somatom Drive2016 | Head | — | — | 696.1–956.5 | 801.2 ± 71.9 | 40.4–52.8 | 46.23 ± 3.28 |
| Thorax | 22.65 ± 2.27 | — | 258.0–686.2 | 413.8 ± 132.2 | 6.8–19.8 | 11.52 ± 3.32 | |
| Abdomen and pelvis | — | 66.50 ± 9.19 | 406.4–1211.3 | 587.4 ± 178.6 | 8.7–23.0 | 11.82 ± 3.08 | |
| Revolution2018 | Head | — | — | 704.7–1352.5 | 1065.6 ± 163.6 | 34.4–59.2 | 52.43 ± 6.54 |
| Thorax | 23.81 ± 2.69 | — | 206.2–900.4 | 459.1 ± 184.9 | 5.6–21.8 | 11.62 ± 4.76 | |
| Abdomen and pelvis | — | 65.50 ± 8.57 | 433.6–1481.7 | 782.3 ± 264.3 | 10.1–28.0 | 14.13 ± 4.30 | |
| Revolution Apex2018 | Head | — | — | 632.0–1254.9 | 913.6 ± 202.0 | 32.8–59.3 | 44.53 ± 9.03 |
| Thorax | 23.71 ± 1.78 | — | 232.2–762.6 | 381.8 ± 125.5 | 5.5–19.5 | 9.74 ± 3.37 | |
| Abdomen and pelvis | — | 68.88 ± 8.69 | 520.1–1387.4 | 831.3 ± 297.7 | 10.6–25.3 | 15.32 ± 4.67 | |
| Aquilion One2015 | Head | — | — | 904.0–1214.3 | 937.3 ± 85.1 | 59.1–59.9 | 59.75 ± 0.28 |
| Thorax | 22.08 ± 1.61 | — | 102.5–427.5 | 181.7 ± 79.0 | 2.4–12.0 | 4.86 ± 2.20 | |
| Abdomen and pelvis | — | 74.80 ± 11.2 | 322.3–1217.0 | 687.3 ± 261.3 | 6.6–25.4 | 13.48 ± 5.06 | |
| Canada | |||||||
| CT scanner | Examination | Mean AP thickness (cm) | Mean weight (kg) | Min/max DLP (mGy·cm) | Mean DLP (mGy·cm) | Min/Max CTDIvol (mGy) | Mean CTDIvol (mGy) |
| Definition Edge2018 | Head | — | — | 588.3–773.0 | 638.7 ± 59.8 | 35.4–45.6 | 40.00 ± 2.98 |
| Thorax | 25.56 ± 1.97 | — | 150.2–844.6 | 363.2 ± 167.2 | 3.8–23.6 | 10.79 ± 5.19 | |
| Abdomen and pelvis | — | 70.82 ± 10.85 | 252.1–797.9 | 434.2 ± 160.4 | 5.4–16.5 | 9.28 ± 2.99 | |
| Definition Flash2011 | Head | — | — | 629.0–929.0 | 820.0 ± 80.0 | 37.9–52.9 | 45.06 ± 3.52 |
| Thorax | 25.22 ± 1.84 | — | 121.0–470.0 | 273.9 ± 78.6 | 4.2–13.7 | 8.10 ± 2.14 | |
| Abdomen and pelvis | — | 70.84 ± 8.46 | 241.0–1014.0 | 493.3 ± 182.5 | 6.3–20.2 | 10.83 ± 3.71 | |
| Optima CT6602013 | Head | — | — | 965.9–826.8 | 893.0 ± 37.7 | 43.8–44.1 | 43.96 ± 0.07 |
| Thorax | 23.99 ± 2.15 | — | 97.0–421.6 | 212.8 ± 98.7 | 2.4–12.0 | 5.33 ± 2.67 | |
| Abdomen and pelvis | — | 72.26 ± 11.25 | 121.6–924.9 | 360.4 ± 198.2 | 2.6–17.8 | 6.70 ± 3.50 | |
| Optima CT6602012 | Head | — | — | 706.1–911.7 | 789.0 ± 49.6 | 40.3–40.5 | 40.43 ± 0.06 |
| Thorax | 24.62 ± 2.72 | — | 156.9–970.9 | 414.7 ± 208.3 | 5.8–25.3 | 11.75 ± 5.67 | |
| Abdomen and pelvis | — | 73.12 ± 9.03 | 273.4–1804.2 | 761.41 ± 468.9 | 5.6–34.9 | 14.50 ± 8.60 | |
Dash (—) indicates not applicable.
Figure 1.Distribution of the median CTDIvol (a) and DLP (b) values from all individual scanners for the head examination. The first eight bars correspond to the Norwegian hospitals and the last four to the Canadian sites; horizontal lines correspond to the determined local DRLs from each country.
Figure
3.Distribution of the median CTDIvol (a) and DLP (b) values from all individual scanners for the abdomen and pelvis examinations. The first eight bars correspond to the Norwegian hospitals and the last four to the Canadian sites; horizontal lines correspond to the determined local DRLs from each country.
Figure
2.Distribution of the median CTDIvol (a) and DLP (b) values from all individual scanners for the thorax examinations. The first eight bars correspond to the Norwegian hospitals and the last four to the Canadian sites; horizontal lines correspond to the determined local DRLs from each country.
Figure 4.Comparison of the local Canadian and Norwegian DRLs determined in the current study with the national and international published data.[14,15,17–21] The DRL for the abdomen and pelvis examination has not been established in Norway.