| Literature DB >> 36163514 |
E Soons1, T Rath2, Y Hazewinkel3, W A van Dop3, D Esposito4, P A Testoni4, P D Siersema3.
Abstract
BACKGROUND AND AIMS: Colonoscopy aims to early detect and remove precancerous colorectal polyps, thereby preventing development of colorectal cancer (CRC). Recently, computer-aided detection (CADe) systems have been developed to assist endoscopists in polyp detection during colonoscopy. The aim of this study was to investigate feasibility and safety of a novel CADe system during real-time colonoscopy in three European tertiary referral centers.Entities:
Keywords: Colorectal adenoma; Colorectal polyp; Computer-aided detection (CADe) system; Detection; Endoscopy
Mesh:
Year: 2022 PMID: 36163514 PMCID: PMC9560918 DOI: 10.1007/s00384-022-04258-9
Source DB: PubMed Journal: Int J Colorectal Dis ISSN: 0179-1958 Impact factor: 2.796
Fig. 1An example video output of a CADe system. Printed with permission of Pentax Medical EMEA
Fig. 2Flow diagram of inclusion
Baseline characteristics of demographic and endoscopic data. CADe computer aided detection system; y years; SD standard deviation; BMI body mass index; iFOBT immunochemical fecal occult blood test; BBPS Boston bowel preparation scale; ‡ missing for one case; † adequate BBPS score if sum ≥ 6 and every segment ≥ 2
| Age (y), mean ± SD | 62.3 ± 12.1 |
| BMI, mean ± SD | 25.9 ± 4.9 |
| Sex (female), | 30 (33.3) |
Indication for colonoscopy, Diagnostic Screening (non-iFOBT) Surveillance | 38 (42.2) 12 (13.3) 40 (44.4) |
| Hereditary cancer syndrome | 4 (4.4) |
| Smoking, | 14 (15.7) |
| Family history of CRC‡, | 19 (21.3) |
Endoscope, EC38-i10 EC-3890Fi | 80 (88.9) 10 (11.1) |
| General anesthesia (no), | 59 (65.6) |
| Total BBPS score, mean ± SD | 7.1 ± 1.9 |
| BBPS rank adequate†, | 73 (81.1) |
| Procedure time (min), mean ± SD | 23.0 ± 9.3 |
| Inspection time excluding biopsy/polypectomy time (min), mean ± SD | 10.8 ± 4.3 |
Polyp characteristics. CADe, computer-aided detection system; SSL, sessile serrated lesion; HP, hyperplastic polyp
Polyps removed per patient, 0 1 2 3 ≥ 4 | 40 (44.4) 27 (30.0) 16 (17.8) 3 (3.3) 4 (4.4) |
Location, Cecum Ascending Transverse Descending Sigmoid Rectum | 10 (11.5) 24 (27.6) 10 (11.5) 16 (18.4) 17 (19.5) 10 (11.5) |
Size, 0–5 mm 6–10 mm 10–20 mm > 20 mm | 62 (71.3) 17 (19.5) 7 (8.0) 1 (1.1) |
Shape (Paris classification), Ip Is IIa IIb | - 43 (49.4) 35 (40.2) 9 (10.3) |
PA-conclusion, Adenoma SSL HP Other | 41 (47.1) 12 (13.8) 23 (26.4) 11 (12.6) |
Detection rates and median number of detected polyps. CADe, computer-aided detection system; PDR, polyp detection ratio; ADR, adenoma detection ratio; SDR, sessile serrated lesion detection ratio; PPC, number of polyps per colonoscopy; APC, number of adenomas per colonoscopy; SPC, number of sessile serrated lesions per colonoscopy; SD, standard deviation
| PDR, | 50 (55.6) |
| ADR, | 26 (28.9) |
| SDR, | 10 (11.1) |
PPC Mean (SD) | 0.97 (1.19) |
APC Mean (SD) | 0.46 (0.88) |
SPC Mean (SD) | 0.13 (0.40) |
(Severe) adverse events
|
| |
|---|---|
Adverse event None Major bleeding Perforation Bradycardia | 88 (97.8) 1 (1.1) - 1 (1.1) |
| Severe adverse event | - |
Feasibility parameters of CADe system. CADe computer aided detection system; IQR interquartile range
| Number of additional polyps discovered with CADe system, median (range) | 0 (0–1) |
| Number of polyps missed with CADe system, median (range) | 0 (0–1) |
Frequency of false positives, 0–10 10–20 20–50 | 42 (46.7) 39 (43.3) 9 (10.0) |
| User friendliness, median (IQR) | 8.0 (2) |
Study characteristics and outcomes of other CADe system studies. RCT, randomized controlled trial; PDR, polyp detection rate; ADR, adenoma detection rate; SDR, sessile serrated lesion detection rate; NR, not reported; ‡characteristics and outcomes of interventional (CADe) arm only; †screening/surveillance reported as one category; ¤0–5 mm and 6–10 mm reported as one category; ◊6–10 mm and > 10 mm reported as one category; *significant difference compared to control group
| China | China | Italy | China | China | China | China | Germany, Italy, Netherlands | |
| RCT | RCT | RCT | RCT | RCT | RCT | RCT, tandem | Prospective cohort | |
| All levels | NR | Experienced | Experienced | All levels | Experienced | Experienced | Expert | |
| 393 | 508 | 341 | 308 | 522 | 484 | 184 | 90 | |
Screening Surveillance Diagnostic | 21.16 - 78.84 | 5.91 - 94.09 | 22.6 25.2 22.3 | 37.34† - 62.66 | 7.66 - 92.34 | 17 - 83 | 31.52 10.33 58.15 | 13.3 44.4 42.2 |
PDR ADR SDR | 47.07* 29.01* 0.76 | 43.65* 39.10* NR | NR 54.8* 7.0 | 38.3* 28.9* NR | 45.02* 29.12* NR | 52* 34* NR | 55.98* 34.78 NR | 55.6 28.9 11.1 |
0–5 mm 6–10 mm > 10 mm | 77.66* 19.68 2.66 | 66.40* 25.20 8.40 | 92¤* 11 | 70.62* 29.38◊* - | 80.12* 16.67* 3.21 | 81* 17 2 | NR | 71.3 19.5 9.1 |
Pedicle Sessile Flat Laterally spreading tumor | 10.11 89.89* NR - | 10.70 37.04* 52.36* NR | NR | 10.17 42.37* 47.46* NR | 9.84 35.34* 54.82* NR | 5 94* NR 1 | NR | - 49.4 50.5 NR |