| Literature DB >> 36071422 |
Jun Zhang1, Jie Shao2, Li Zheng2, Xia Zhao2, Shu Chen2.
Abstract
BACKGROUND: To evaluate a new method of implantable collamer lens (ICL) sizing based on ultrasound biomicroscopy (UBM) video clips.Entities:
Keywords: Intraocular; Lenses; Phakic Intraocular Lens; Refractive Errors; Ultrasound Biomicroscopy
Mesh:
Year: 2022 PMID: 36071422 PMCID: PMC9454160 DOI: 10.1186/s12886-022-02583-9
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.086
Fig. 1a Distribution of the percentage difference in toric implantable collamer lens (TICL) size between the Zhang & Zheng implantable collamer lens (ZZ ICL) formula and the STAAR formula. b Distribution of the percentage difference in TICL size between the two formulae in patients with a horizontally implanted TICL. c Distribution of the vault ranges for TICLs of differing sizes. d Distribution of the difference in vault size between the achieved value and the value predicted by the ZZ ICL formula. e Distribution of the vault prediction error for the ZZ ICL formula
Vault characteristics
| Parameter | n | Achieved postoperatively | Predicted by ZZ ICL | P |
|---|---|---|---|---|
| Vault height (µm) | 168 | 528 ± 193 (160 − 1040) | 545 ± 156 (200 − 1186) | 0.227 |
| Larger size implanteda | 17 | 404 ± 119 (240 − 650) | 513 ± 111 (316 − 690) | 0.014 |
| Smaller size implanteda | 27 | 505 ± 189 (280 − 1040) | 506 ± 157 (258 − 830) | 0.969 |
| Same size implanteda | 124 | 550 ± 196 (160 − 1040) | 558 ± 160 (200 − 1186) | 0.648 |
Data are shown as mean ± SD (range). ZZ ICL: Zhang & Zheng implantable collamer lens
aIn comparison to the size recommended by the STAAR website software
Refractive prediction errors
| Parameter (mean ± SD) | ZZ ICL | STAAR | P |
|---|---|---|---|
| Spherical equivalent prediction error (D) | 0.01 ± 0.48 | 0.16 ± 0.48 | < 0.001 |
| Absolute cylindrical prediction error (D) | 0.40 ± 0.27 | 0.82 ± 0.54 | < 0.001 |
Data are shown as mean ± SD
ZZ ICL Zhang & Zheng implantable collamer lens
Fig. 2The absolute and vector cylindrical refraction prediction error for each formula
Fig. 3Ultrasound biomicroscopy image showing a wide ciliary sulcus