Linze Li1,2, Changqing Jiang1, Wanru Duan3,4, Zhiyan Wang1, Feng Zhang1, Changgeng He1, Tiangang Long1, Luming Li1,5,6,7.
Abstract
Neural electrode interfaces are essential to the stimulation safety and recording quality of various bioelectronic therapies. The recently proposed hierarchical platinum-iridium (Pt-Ir) electrodes produced by femtosecond lasers have exhibited superior electrochemical performance in vitro, but their in vivo performance is still unclear. In this study, we explored the electrochemical performance, biological response, and tissue adhesion of hierarchical Pt-Ir electrodes by implantation in adult rat brains for 1, 8, and 16 weeks. Regular smooth Pt-Ir electrodes were used as a control. The results showed that the electrochemical performance of both electrodes decreased and leveled off during implantation. However, after 16 weeks, the charge storage capacity of hierarchical electrodes stabilized at ~16.8 mC/cm2, which was 15 times that of the smooth control electrodes (1.1 mC/cm2). Moreover, the highly structured electrodes had lower impedance amplitude and cutoff frequency values. The similar histological response to smooth electrodes indicated good biocompatibility of the hierarchically structured Pt-Ir electrodes. Given their superior in vivo performance, the femtosecond laser-treated Pt-Ir electrode showed great potential for neuromodulation applications.Entities:
Keywords: Electrical and electronic engineering; Nanostructures
Year: 2022 PMID: 36065436 PMCID: PMC9440118 DOI: 10.1038/s41378-022-00433-8
Source DB: PubMed Journal: Microsyst Nanoeng ISSN: 2055-7434 Impact factor: 8.006
Fig. 1Study overview.
a Smooth and hierarchical Pt-Ir electrode surfaces (sPt-Ir and hPt-Ir). PU polyurethane. b Fixation base of the electrode[17]. c X-ray photo of the implanted electrode. d Experimental design. WE working electrode, RE reference electrode, CE counter electrode, CV cyclic voltammetry, EIS electrochemical impedance spectrum, HE hematoxylin–eosin staining, IHC immunohistochemistry, SEM scanning electron microscopy.
Groups of experimental animals.
| Electrode type | Animal number | ||
|---|---|---|---|
| 1 week | 8 weeks | 16 weeks | |
| sPt-Ir | 6 | 6 | 6 |
| hPt-Ir | 6 | 6 | 6 |
Fig. 2Weekly body weight.
Rats implanted for a 1 week, b 8 weeks and c 16 weeks.
Fig. 3Electrochemical performance results.
a Cyclic voltammetry curves of sPt-Ir and hPt-Ir at 0 and 16 weeks. b In vitro and in vivo charge storage capacity (CSC). The inset shows the trend of normalized CSC varying with implantation time compared to the in vitro value. c In vitro and in vivo electrochemical impedance spectra. An asterisk (*) denotes p < 0.05 (one-way ANOVA).
Fig. 4Histological results.
a HE, Iba-1, GFAP, and NeuN staining of the tissue surrounding the electrodes at 1, 8, and 16 weeks after implantation. b Normalized neurons count with increasing distance from the electrode interface.
Fig. 5Scanning electron microscopy of electrodes surfaces.
The sPt-Ir and hPt-Ir electrodes surfaces a as processed and after b 1 week, c 8 weeks, and d 16 weeks of implantation.