Ajay Kumar1, Pearl Jacob1, Ankit Watts1, Anwin Joseph1, Harneet Kaur1, Monika Hooda1, Amritjyot Kaur1, Baljinder Singh1.
Abstract
Objectives: The objective is to standardize the reconstruction parameters for the time-of-flight (TOF) versus non-TOF positron-emission tomography/computed tomography (PET/CT) imaging data and validation of the same in a clinical setting.Entities:
Keywords: Image contrast; NEMA IQ phantom; positron emission tomography Imaging; signal-to-noise ratio; time-of-flight/non-time-of-flight reconstruction
Year: 2022 PMID: 35982821 PMCID: PMC9380809 DOI: 10.4103/ijnm.ijnm_137_21
Source DB: PubMed Journal: Indian J Nucl Med ISSN: 0974-0244
The visual scores of 10.0-mm lesion for various iterations of lutetium-yttrium oxyorthosilicate time-offlight and (lutetium-yttrium oxyorthosilicate non-time-offlight and bismuth germanate non-time-of-flight systems
| Iteration number | LYSO PET system Visual score | BGO PET system Visual score non-TOF | |
|---|---|---|---|
|
| |||
| TOF | non-ToF | ||
| 1 | 2 | 0 | 0 |
| 2 | 3 | 1 | 1 |
| 3 | 2 | 3 | 2 |
| 4 | 2 | 2 | 1 |
| 5 | 2 | 1 | 1 |
| 6 | 2 | 1 | 1 |
| 7 | 2 | 1 | 1 |
| 8 | 2 | 1 | 1 |
| 9 | 1 | 1 | 1 |
| 10 | 1 | 1 | 1 |
PET: Positron-emission tomography, TOF: Time of flight, LYSO: Lutetium-yttrium oxyorthosilicate, BGO: Bismuth germanate
Figure 1Transaxial images of NEMA NU-2 IQ phantom on visual analysis demonstrating best image contrast for 10.0-mm sphere (smallest sphere) at 2nd (Ist row), 3rd (2nd row), and 3rd (3rd row) iteration for lutetium–yttrium oxyorthosilicate time-of-flight, lutetium–yttrium oxyorthosilicate non-time-of-flight and bismuth germanate non-time-of-flight systems, respectively
The signal-to-noise ratio values obtained using lutetium-yttrium oxyorthosilicate (time-of-flight and non-timeof- flight) and bismuth germanate (non-time-of-flight) positron-emission tomography systems for 10-mm and 13-mm hot spheres at different iterations
| Iterations | LYSO PET system SNR | BGO non-TOF SNR | ||||
|---|---|---|---|---|---|---|
|
|
| |||||
| 10 mm | 13 mm | 10 mm Non-TOF | 13 mm Non-TOF | |||
|
|
| |||||
| TOF | Non-TOF | TOF | Non-TOF | |||
| 1 | 18.19 | 4.92 | 49.43 | 14.85 | 4.08 | 17.15 |
| 2 | 20.66 | 7.08 | 53.32 | 16.86 | 8.49 | 22.63 |
| 3 | 14.39 | 10.80 | 35.43 | 25.80 | 8.42 | 17.19 |
| 4 | 10.23 | 9.83 | 24.89 | 24.30 | 7.02 | 13.25 |
| 5 | 9.98 | 9.37 | 22.61 | 23.51 | 6.04 | 11.07 |
| 6 | 8.00 | 6.62 | 17.43 | 17.10 | 5.98 | 10.47 |
| 7 | 7.46 | 4.74 | 15.61 | 11.86 | 5.94 | 10.07 |
| 8 | 7.33 | 4.68 | 14.91 | 11.43 | 5.63 | 9.23 |
| 9 | 6.17 | 4.35 | 12.26 | 10.77 | 5.32 | 8.54 |
| 10 | 6.41 | 4.14 | 12.21 | 10.70 | 4.58 | 7.34 |
PET: Positron-emission tomography, TOF: Time of flight, LYSO: Lutetium-yttrium oxyorthosilicate, BGO: Bismuth germinate, SNR: Signal-to-noise ratio
The contrast recovery coefficient/background variability ratios obtained in lutetium-yttrium oxyorthosilicate (time-of-flight and non-time-of-flight) and bismuth germanate (non-time-of-flight) positron-emission tomography systems of 10-mm and 13-mm hot spheres at different iterations
| Iterations | LYSO PET system CRC/BV ratio | BGO non-TOF CRC/BV ratio | ||||
|---|---|---|---|---|---|---|
|
|
| |||||
| 10 mm | 13 mm | 10 mm Non-TOF | 13 mm Non-TOF | |||
|
|
| |||||
| TOF | Non-TOF | TOF | Non-TOF | |||
| 1 | 2.66 | 1.67 | 5.44 | 3.88 | 0.666 | 3.17 |
| 2 | 3.21 | 2.27 | 6.48 | 4.23 | 1.58 | 4.03 |
| 3 | 3.11 | 2.34 | 6.05 | 4.35 | 1.49 | 3.89 |
| 4 | 2.82 | 2.31 | 5.23 | 4.33 | 1.49 | 3.896 |
| 5 | 2.86 | 2.30 | 5.38 | 4.25 | 1.49 | 3.35 |
| 6 | 2.56 | 2.24 | 5.05 | 4.13 | 1.49 | 3.93 |
| 7 | 2.52 | 2.17 | 4.82 | 3.98 | 1.40 | 3.33 |
| 8 | 2.44 | 2.12 | 4.7 | 3.87 | 1.32 | 3.07 |
| 9 | 2.42 | 2.09 | 4.61 | 3.75 | 1.36 | 3.45 |
| 10 | 2.33 | 2.06 | 4.46 | 3.65 | 1.22 | 2.64 |
PET: Positron-emission tomography, TOF: Time of flight, LYSO: Lutetium-yttrium oxyorthosilicate, BGO: Bismuth germinate, CRC/BV: Contrast recovery coefficient/background variability
Figure 2Bar diagram presenting signal -to-noise ratio values (Y-axis) for 10.0-mm and 13.0-mm hot spheres in lutetium–yttrium oxyorthosilicate time-of-flight lutetium–yttrium oxyorthosilicate non-time-of-flight and bismuth germanate non-time-of-flight systems at 2nd and 3rd iteration (X-axis), respectively
Figure 3Bar diagram presenting contrast recovery coefficient/background variability values (Y-axis) for 10-mm and 13-mm hot spheres in lutetium–yttrium oxyorthosilicate time-of-flight, lutetium–yttrium oxyorthosilicate nontime-of-flight, and bismuth germanate nontime-of-flight systems at 2nd and 3rd iteration (X-axis), respectively
Figure 4The visual image analysis in an 18-year-old NHL patient showing better lesion (right inguinal lymph node) contrast at 2/12 iteration (a) in lutetium–yttrium oxyorthosilicate time-of-flight positron-emission tomography and at 3/12 iteration (b) in lutetium–yttrium oxyorthosilicate non-time-of-flight positron-emission tomography system. The bismuth germanate non-time-of-flight positron-emission tomography imaging in a 64 years patient of NHL showing better lesion (spleen) contrast at 2/14 (c) iteration
Figure 5[18F]-fludeoxyglucose positron-emission tomography MIP images reconstructed in time-of-flight (a) and nontime-of-flight mode (b) of a 62-year-old male patient diagnosed with lymphoma showing nearly identical lesions on both the positron-emission tomography systems. A subcentimetric axillary lymph node (<10.0-mm) seen (arrow) more distinctly on lutetium–yttrium oxyorthosilicate time-of-flight system (a) compared to nontime-of-flight counterpart (b)
The signal-to-noise ratio values of the smallest identifiable lesion for lutetium-yttrium oxyorthosilicate time-of-flight and lutetium-yttrium oxyorthosilicate nontime- of-flight systems in six patients using 2/12 and 3/12 iteration/subset reconstruction algorithms
| Patient number | SNR (TOF) 2/12 | SNR (non-TOF) 3/12 |
|---|---|---|
| 1 | 15.170 | 13.607 |
| 2 | 19.901 | 18.185 |
| 3 | 3.192 | 0.538 |
| 4 | 7.924 | 7.997 |
| 5 | 18.928 | 16.34 |
| 6 | 14.251 | 13.882 |
| Mean±SD | 13.228±6.49 | 11.758±6.48 |
SNR: Signal-to-noise ratio, TOF: Time of flight, SD: Standard deviation
The signal-to-noise ratio values of the smallest identifiable lesion for bismuth germanate non-time-offlight system in 10 patients in using 2/14 and 3/14 subset/ iteration reconstruction algorithms
| Patient number | SNR value for 2/14 | SNR value for 3/14 |
|---|---|---|
| 1 | 0.676 | 0.9302 |
| 2 | 9.244 | 9.181 |
| 3 | 15.79 | 9.966 |
| 4 | 5.50 | 4.141 |
| 5 | 27.35 | 17.841 |
| 6 | 51.30 | 46.675 |
| 7 | 11.37 | 7.657 |
| 8 | 14.51 | 13.331 |
| 9 | 18.67 | 12.495 |
| 10 | 44.98 | 25.506 |
| Mean±SD | 19.93±16.60 | 14.70±13.16 |
SNR: Signal-to-noise ratio, SD: Standard deviation