Time-of-Flight Information Improved the Detectability of Subcentimeter Spheres Using a Clinical PET/CT Scanner.

Naoki Hashimoto1, Keishin Morita1, Yuji Tsutsui1,2, Kazuhiko Himuro2, Shingo Baba2, Masayuki Sasaki3.   

Abstract

Recent advancements in clinical PET/CT scanners have improved the detectability of small lesions. However, the ideal reconstruction parameters for detecting small lesions have not yet been sufficiently clarified. The purpose of this study was to investigate the detectability of subcentimeter spheres using a clinical PET/CT scanner.
Methods: We used a clinical PET/CT scanner to obtain the data of a National Electrical Manufacturers Association body phantom consisting of 6 small spheres (inner diameters, 4.0, 5.0, 6.2, 7.9, 10, and 37 mm) containing 18F solution. The background activity was 2.65 kBq/mL, and the sphere-to-background ratio was 8. The PET data obtained for 2 and 120 min were reconstructed using ordered-subsets expectation maximization (OSEM), OSEM + point-spread function (PSF), and OSEM + time-of-flight (TOF) with voxel sizes of 2.04 × 2.04 × 2.00 mm (2-mm voxels) and 4.07 × 4.07 × 3.99 mm (4-mm voxels). A gaussian filter was not used. The image quality was evaluated by visual assessment, as well as by physical assessment of the detectability index and recovery coefficients.
Results: According to the visual assessment, the detectability of the spheres improved using TOF and a longer acquisition. Using the OSEM+TOF model, the smallest visually detected spheres were 5 mm in diameter with a 120-min acquisition and 6 mm in diameter with a 2-min acquisition. According to physical assessment, the detectability of spheres 10 mm or smaller using the OSEM+TOF image was superior to that using the OSEM image. In addition, the detectability of each hot sphere and recovery coefficient with 2-mm voxels was superior to that with 4-mm voxels. Although OSEM+PSF images showed less background noise, detectability and the recovery coefficient were not improved for spheres 8 mm or smaller.
Conclusion: The TOF model with 2-mm voxels improved the detectability of subcentimeter hot spheres on a clinical PET/CT scanner.
© 2018 by the Society of Nuclear Medicine and Molecular Imaging.

Keywords:  PET/CT; instrumentation; point-spread function; small voxel size; subcentimeter sphere; time-of-flight

Mesh:

Year:  2018        PMID: 29599404     DOI: 10.2967/jnmt.117.204735

Source DB:  PubMed          Journal:  J Nucl Med Technol        ISSN: 0091-4916


  2 in total

1.  A NIM PET/CT phantom for evaluating the PET image quality of micro-lesions and the performance parameters of CT.

Authors:  Shujie Lu; Peng Zhang; Chengwei Li; Jie Sun; Wenli Liu; Pu Zhang
Journal:  BMC Med Imaging       Date:  2021-11-08       Impact factor: 1.930

2.  Evaluation of Reconstruction Algorithms to Validate the NEMA Phantom Results in Clinical Scenario - A Comparative Study Using Time-of-Flight versus Non-Time-of-Flight Positron Emission Tomography Imaging.

Authors:  Ajay Kumar; Pearl Jacob; Ankit Watts; Anwin Joseph; Harneet Kaur; Monika Hooda; Amritjyot Kaur; Baljinder Singh
Journal:  Indian J Nucl Med       Date:  2022-07-08
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.