PET performance measurements using the NEMA NU 2-2001 standard.

Margaret E Daube-Witherspoon1, Joel S Karp, Michael E Casey, Frank P DiFilippo, Horace Hines, Gerd Muehllehner, Vilim Simcic, Charles W Stearns, Lars-Eric Adam, Steve Kohlmyer, Vesna Sossi.   

Abstract

UNLABELLED: The NU 2-1994 standard document for PET performance measurements has recently been updated. The updated document, NU 2-2001, includes revised measurements for spatial resolution, intrinsic scatter fraction, sensitivity, counting rate performance, and accuracy of count loss and randoms corrections. The revised measurements are designed to allow testing of dedicated PET systems in both 2-dimensional and 3-dimensional modes as well as coincidence gamma cameras, conditions not considered in the original NU 2-1994 standard. In addition, the updated measurements strive toward being more representative of clinical studies, in particular, whole-body imaging.
METHODS: Performance measurements following the NU 2-1994 and NU 2-2001 standards were performed on several different PET scanners. Differences between the procedures and resulting performance characteristics, as well as the rationale for these changes, were noted.
RESULTS: Spatial resolution is measured with a point source in all 3 directions, rather than a line source, as specified previously. For the measurements of intrinsic scatter fraction, sensitivity, and counting rate performance, a 70-cm line source is now specified, instead of a 19-cm-long cylindric phantom. The longer configuration permits measurement of these performance characteristics over the entire axial field of view of all current PET scanners and incorporates the effects of activity outside the scanner. A measurement of image quality has been added in an effort to measure overall image quality under clinically realistic conditions. This measurement replaces the individual measurements of uniformity and of the accuracy of corrections for attenuation and scatter.
CONCLUSION: The changes from the NU 2-1994 standard to the NU 2-2001 standard strive toward establishing relevance with clinical studies. The tests in the updated standard also are, in general, simpler and less time-consuming to perform than those in the NU 2-1994 standard.

Mesh:

Year:  2002        PMID: 12368380

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  50 in total

Review 1.  From PET detectors to PET scanners.

Authors:  John L Humm; Anatoly Rosenfeld; Alberto Del Guerra
Journal:  Eur J Nucl Med Mol Imaging       Date:  2003-10-02       Impact factor: 9.236

2.  Evaluation of Noise Properties in PSF-Based PET Image Reconstruction.

Authors:  Shan Tong; Adam M Alessio; Paul E Kinahan
Journal:  IEEE Nucl Sci Symp Conf Rec (1997)       Date:  2009-10-24

3.  A unified Fourier theory for time-of-flight PET data.

Authors:  Yusheng Li; Samuel Matej; Scott D Metzler
Journal:  Phys Med Biol       Date:  2015-12-22       Impact factor: 3.609

Review 4.  Small animal PET: aspects of performance assessment.

Authors:  Simone Weber; Andreas Bauer
Journal:  Eur J Nucl Med Mol Imaging       Date:  2004-10-02       Impact factor: 9.236

5.  Instrumentation quality assurance and performance.

Authors:  Kenneth J Nichols; Stephen L Bacharach; Steven R Bergmann; S James Cullom; Edward P Ficaro; James R Galt; Gary V Heller; Jonathan Links; Josef Machac
Journal:  J Nucl Cardiol       Date:  2006-11       Impact factor: 5.952

6.  Data acquisition and analysis: the strength of methodology in nuclear medicine and molecular imaging.

Authors:  Peter Knoll; Horst Köhn; Siroos Mirzaei
Journal:  Eur J Nucl Med Mol Imaging       Date:  2007-06       Impact factor: 9.236

7.  Performance of the new generation of whole-body PET/CT scanners: Discovery STE and Discovery VCT.

Authors:  M Teräs; T Tolvanen; J J Johansson; J J Williams; J Knuuti
Journal:  Eur J Nucl Med Mol Imaging       Date:  2007-07-28       Impact factor: 9.236

8.  Benefit of time-of-flight in PET: experimental and clinical results.

Authors:  Joel S Karp; Suleman Surti; Margaret E Daube-Witherspoon; Gerd Muehllehner
Journal:  J Nucl Med       Date:  2008-02-20       Impact factor: 10.057

9.  Advancement in PET quantification using 3D-OP-OSEM point spread function reconstruction with the HRRT.

Authors:  Andrea Varrone; Nils Sjöholm; Lars Eriksson; Balazs Gulyás; Christer Halldin; Lars Farde
Journal:  Eur J Nucl Med Mol Imaging       Date:  2009-05-13       Impact factor: 9.236

10.  Performance evaluation of the new whole-body PET/CT scanner: Discovery ST.

Authors:  Valentino Bettinardi; Massimo Danna; Annarita Savi; Michela Lecchi; Isabella Castiglioni; Maria Carla Gilardi; Helmut Bammer; Giovanni Lucignani; Ferruccio Fazio
Journal:  Eur J Nucl Med Mol Imaging       Date:  2004-02-10       Impact factor: 9.236

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.