Sharib Gaffar1,2, Anthony C McCanta2,3, Pierangelo Renella2,3,4.
Abstract
Entities:
Keywords: Arrhythmia; Cardiac magnetic resonance imaging; Cardiomyopathy; Congenital heart disease; ICD; Image quality; Subcutaneous implantable cardioverter-defibrillator
Year: 2022 PMID: 35860767 PMCID: PMC9289066 DOI: 10.1016/j.hrcr.2022.04.014
Source DB: PubMed Journal: HeartRhythm Case Rep ISSN: 2214-0271
Figure 1Anteroposterior and lateral chest radiographs. A: The conductor coil of the subcutaneous implantable cardioverter-defibrillator (SICD) is subcutaneously tunneled from the anterior chest wall to the xiphoid process. B: The conductor is then subcutaneously tunneled to the left axilla, where the generator is placed. C: This patient did not achieve appropriate SICD sensing and defibrillation on initial device placement because the coil and generator did not vector across the majority of ventricular myocardium.
Patient demographics, clinical results, and image quality score of each cardiac magnetic resonance imaging study
| Patient | BSA (m2) | Days between SICD implant & CMR | Indications for CMR | Clinical question answered? (Y/N) | RV qualitative (size + function) | RV quantitative | RV IQS | LV qualitative (size + function) | LV quantitative | LV IQS | Sample image |
|---|---|---|---|---|---|---|---|---|---|---|---|
| 1A | 1.54 | 382 | CMR 1: Evaluate ARVD | No | Yes | No | 2 | No | No | 1 | |
| 1B | 1.56 | 811 | CMR 2: Evaluate ARVD | Yes | Yes | 3 | Yes | No | 2 | ||
| 2 | 2.04 | 442 | Hypertrophic cardiomyopathy | Yes | Yes | Yes | 3 | Yes | Yes | 3 | |
| 3 | 1.99 | 52 | Evaluate ventricular function in setting of recurrent myocarditis | No; could not evaluate for LV enlargement | Yes | Yes | 3 | Yes | No | 2 | |
| 4 | 2.31 | 547 | Evaluate ARVD & subendocardial ischemia | No; RV free wall visualized but no reproducible volume could be obtained | Yes | Yes | 3 | No | No | 2 | |
| Mean | 1.89 | 446.8 | 2.8 | 2 | |||||||
| Median | 1.99 | 442 | 3 | 2 |
ARVC = arrhythmogenic right ventricular cardiomyopathy; BSA = body surface area; CMR = cardiac magnetic resonance imaging; IQS = image quality score; LV = left ventricle; RV = right ventricle; SICD = subcutaneous implantable cardioverter-defibrillator.
IQS: 1 = nondiagnostic (poor visualization of ventricular chambers); 2 = poor quality (chamber walls seen but poorly defined); 3 = moderate quality (chambers clearly identified but with suboptimal definition of the endocardium and trabeculae); 4 = good quality (chambers clearly identified with good definition of the endocardium and trabeculae).
Figure 2Referred sample cardiac magnetic resonance images for each patient from Table 1.
Figure 3Image quality score (IQS) shown alongside corresponding cardiac magnetic resonance images, including steady-state free precession sequences and gradient echo sequences. LV = left ventricle; RV = right ventricle.