Myocardial T1 mapping for patients with implanted cardiac devices using wideband inversion recovery spoiled gradient echo readout.

Jiaxin Shao1, Shams Rashid1, Pierangelo Renella2,3, Kim-Lien Nguyen4,5, Peng Hu1,6.   

Abstract

PURPOSE: To develop and validate a technique for myocardial T1 mapping in patients with implantable cardioverter defibrillators (ICDs).
METHODS: A MOLLI-based pulse sequence, named Wideband-FLASH-MOLLI, was developed by incorporating a fast low angle shot (FLASH) readout and a wideband inversion pulse. The performance of Wideband-FLASH-MOLLI was evaluated using phantom studies and validated in eight healthy volunteers and ten patients with ICDs at 1.5 Tesla. Comparisons were made between Wideband-FLASH-MOLLI, FLASH-MOLLI, and bSSFP-MOLLI sequences.
RESULTS: In phantom studies, the maximum T1 estimation errors using Wideband-FLASH-MOLLI with and without an ICD were less than 3% for T1 range from 212 to 1673 ms. In all healthy volunteers, there was no significant native myocardial T1 estimation difference using Wideband-FLASH-MOLLI before and after the external attachment of an ICD to the body coil (1178 ± 27 ms versus 1174 ± 33 ms; P = 0.41). Due to the presence of an ICD, the magnitude images acquired using bSSFP-MOLLI and FLASH-MOLLI showed severe artifacts within the myocardium. In contrast, no or negligible device-induced artifacts were noted within the myocardial regions of the healthy volunteers or the patients with ICDs when using Wideband-FLASH-MOLLI.
CONCLUSION: This study demonstrates the feasibility of using Wideband-FLASH-MOLLI to mitigate image artifacts and to produce accurate myocardial T1 maps in patients with ICDs. Magn Reson Med 77:1495-1504, 2017.
© 2016 International Society for Magnetic Resonance in Medicine. © 2016 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  ICD; MOLLI; T1 mapping; myocardial fibrosis; wideband

Mesh:

Year:  2016        PMID: 27018872     DOI: 10.1002/mrm.26223

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   3.737


  14 in total

1.  Feasibility of Cardiac Magnetic Resonance Wideband Protocol in Patients With Implantable Cardioverter Defibrillators and Its Utility for Defining Scar.

Authors:  Amita Singh; Keigo Kawaji; Neha Goyal; Noreen T Nazir; Andrew Beaser; Virginia O'Keefe-Baker; Karima Addetia; Roderick Tung; Peng Hu; Victor Mor-Avi; Amit R Patel
Journal:  Am J Cardiol       Date:  2019-01-31       Impact factor: 2.778

2.  Simultaneous multislice imaging for native myocardial T1 mapping: Improved spatial coverage in a single breath-hold.

Authors:  Sebastian Weingärtner; Steen Moeller; Sebastian Schmitter; Edward Auerbach; Peter Kellman; Chetan Shenoy; Mehmet Akçakaya
Journal:  Magn Reson Med       Date:  2017-06-05       Impact factor: 4.668

Review 3.  Towards accurate and precise T 1 and extracellular volume mapping in the myocardium: a guide to current pitfalls and their solutions.

Authors:  Donnie Cameron; Vassilios S Vassiliou; David M Higgins; Peter D Gatehouse
Journal:  MAGMA       Date:  2017-06-12       Impact factor: 2.310

4.  Accuracy, precision, and reproducibility of myocardial T1 mapping: A comparison of four T1 estimation algorithms for modified look-locker inversion recovery (MOLLI).

Authors:  Jiaxin Shao; Dapeng Liu; Kyunghyun Sung; Kim-Lien Nguyen; Peng Hu
Journal:  Magn Reson Med       Date:  2016-12-04       Impact factor: 4.668

Review 5.  Role of Cardiac MRI Imaging of Focal and Diffuse Inflammation and Fibrosis in Cardiomyopathy Patients Who Have Pacemakers/ICD Devices.

Authors:  Ananna Zaman; Samantha Zhao; Jordana Kron; Antonio Abbate; Anna Tomdio; W Gregory Hundley; Jennifer H Jordan
Journal:  Curr Cardiol Rep       Date:  2022-08-19       Impact factor: 3.955

6.  Wideband myocardial perfusion pulse sequence for imaging patients with a cardiac implantable electronic device.

Authors:  KyungPyo Hong; Jeremy D Collins; Bradley P Knight; James C Carr; Daniel C Lee; Daniel Kim
Journal:  Magn Reson Med       Date:  2018-09-09       Impact factor: 4.668

7.  Wideband LGE MRI permits unobstructed viewing of myocardial scarring in a patient with an MR-conditional subcutaneous implantable cardioverter-defibrillator.

Authors:  Amir Ali Rahsepar; Jeremy D Collins; Bradley P Knight; KyungPyo Hong; James C Carr; Daniel Kim
Journal:  Clin Imaging       Date:  2018-05-04       Impact factor: 1.605

8.  Accelerated Wideband Myocardial Perfusion Pulse Sequence with Compressed Sensing Reconstruction for Myocardial Blood Flow Quantification in Patients with a Cardiac Implantable Electronic Device.

Authors:  KyungPyo Hong; Jeremy D Collins; Benjamin H Freed; Lexiaozi Fan; Andrew E Arai; Li-Yueh Hsu; Daniel C Lee; Daniel Kim
Journal:  Radiol Cardiothorac Imaging       Date:  2020-04-16

Review 9.  Cardiac magnetic resonance fingerprinting: Trends in technical development and potential clinical applications.

Authors:  Brendan L Eck; Scott D Flamm; Deborah H Kwon; W H Wilson Tang; Claudia Prieto Vasquez; Nicole Seiberlich
Journal:  Prog Nucl Magn Reson Spectrosc       Date:  2020-11-06       Impact factor: 9.795

10.  Evaluation of image quality of wideband single-shot late gadolinium-enhancement MRI in patients with a cardiac implantable electronic device.

Authors:  Sarah M Schwartz; Ashitha Pathrose; Ali M Serhal; Ann B Ragin; Jessica Charron; Bradley P Knight; Rod S Passman; Ryan J Avery; Daniel Kim
Journal:  J Cardiovasc Electrophysiol       Date:  2020-11-13
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.