| Literature DB >> 33964915 |
Da Young Shin1, Ho Sik Hwang2, Hyun Seung Kim3, Man Soo Kim1, Eun Chul Kim4.
Abstract
BACKGROUND: The aim of this study is to analyze and compare the clinical results of toric intraocular lens (IOL) and monofocal IOL implantation when the target refraction value is -3 diopter (D) in cataract patients with corneal astigmatism > 1.5 diopters (D).Entities:
Keywords: Corneal astigmatism; Distant visual acuity; Myopia; Near visual acuity; Tecnis ZCT toric IOL; Toric intraocular lens implantation
Mesh:
Year: 2021 PMID: 33964915 PMCID: PMC8106217 DOI: 10.1186/s12886-021-01966-8
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Preoperative demographic and biometric data of the subjects (n=60)
| Parameter | Group | ||
|---|---|---|---|
| Monofocal IOL | Toric IOL | ||
| Eyes (n) | 60 | 40 | |
| IOL model | Technis ZCB00 | Technis ZCT toric | |
| Sex (M/F) | 23/37 | 12/28 | |
| Age (years) | 64.4±4.64 | 62.7±11.3 | 0.542 |
| Mean UCVA (logMAR) | 0.65±0.29 | 0.72±0.28 | 0.445 |
| Mean BCVA (logMAR) | 0.40±0.35 | 0.36±0.27 | 0.947 |
| Mean refractive cylinder(D) | 1.84±0.35 | 2.18±1.00 | 0.277 |
| Mean keratometric cylinder(D) | 1.82±0.35 | 1.76±1.07 | 0.102 |
Values are presented as mean±standard deviation unless otherwise indicated
IOL intraocular lens, D diopters
Preoperative and postoperative mean residual refractive cylinder (D) and mean residual keratometric cylinder (D) according to treatment group(monofocal, toric)
| Parameter | Group | ||
|---|---|---|---|
| Monofocal IOL | Toric IOL | ||
| Mean residual refractive cylinder (D) | 1.65±0.77 | 0.80±0.46 | 0.001 |
| Mean residual keratometric cylinder (D) | 1.45±0.64 | 1.50±0.62 | 0.626 |
Values are presented as mean±standard deviation unless otherwise indicated
IOL intraocular lens, D diopters
Fig. 1Pre and postoperative distant UCVA (logMAR) of treatment groups (monofocal, toric). UCVA, uncorrected distance visual acuity; POD, postoperative day
Fig. 2Number of eyes according to postoperative distant UCVA (logMAR) in two treatment groups(monofocal, toric IOL). UCVA, uncorrected distance visual acuity; POD, postoperative day
Postoperative visual acuity(logMAR) according to treatment group(monofocal, toric)
| Parameter | Group | ||
|---|---|---|---|
| Monofocal IOL | Toric IOL | ||
| Mean distance UCVA | 0.55±0.22 | 0.38±0.14 | 0.026 |
| Mean distance BCVA (logMAR) | 0.11±0.14 | 0.08±0.08 | 0.710 |
| Mean near UCVA (logMAR) | 0.48±0.32 | 0.26±0.33 | 0.030 |
Values are presented as mean±standard deviation unless indicated otherwise
IOL intraocular lens, UCVA uncorrected visual acuity, BCVA best-corrected visual acuity
Fig. 3Pre and postoperative mean residual refractive cylinder(D) and mean residual keratometric cylinder(D) of treatment groups (monofocal, toric). POD, postoperative day