Helena Čičak1, Vanja Radišić Biljak1, Ana-Maria Šimundić1,2.
Abstract
Introduction: The aim of this study was to perform a comprehensive verification of a 6-part differential haematology analyser Siemens Advia 2120i (Erlangen, Germany), prior to its routine implementation. Materials and methods: Our verification protocol included: precision (within- and between-run), estimated bias (%) as measure of trueness, which was calculated from observed and manufacturers' declared value, analytical measuring interval (AMI), carryover, confirmation of a limit of blank (LoB), determination of a limit of detection (LoD) and limit of quantitation (LoQ). The K2 ethylenediaminetetraacetic acid (EDTA) patients' leftover samples were used for verification of analyser Advia 2021i. Acceptance criteria were based on manufacturer technical specifications (Siemens), 2016 state-of-the-art criteria (Vis and Huisman), and EFLM Biological Variation Database.Entities:
Keywords: blood cell count; haematology; haematology analyser; verification
Mesh:
Year: 2022 PMID: 35799991 PMCID: PMC9195601 DOI: 10.11613/BM.2022.020710
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.515
Estimated within- and between-run precision for Siemens Advia 2120i haematology analyser
|
|
|
|
|
| ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| L | N | H | L | N | H | Siemens | SOTA | L | N | H | SOTA for values | EuBIVAS | EuBIVAS | |
| WBC, (x109/L) | 4.7 (0.53) | 2.0 | 2.4 | 2.4 | 1.6 | 1.8 | 2.7 | 2.5 | 2.3 | 2.4 | 2.5 | 6.0/2.5/1.5* | 8.1 | 5.4 |
| RBC, (x1012/L) | 0.8 | 0.6 | 0.7 | 0.7 | 0.7 | 0.9 | 1.2 | 1.1 | 0.8 | 0.8 | 1.0 | 1.1 | 2.0 | 1.3 |
| Haemoglobin, (g/L) | 0.8 | 0.6 | 0.9 | 1.1 | 0.7 | 0.7 | 0.9 | 0.9 | 1.7 | 1.3 | 0.4 | 1.0 | 2.0 | 1.4 |
| Haematocrit, (L/L) | / | / | / | 0.8 | 0.8 | 1.0 | / | 1.2 | 1.2 | 1.3 | 1.2 | 1.4 | 2.1 | 1.4 |
| MCV, (fL) | 0.3 | 0.3 | 0.3 | 0.5 | 0.3 | 0.3 | 0.8 | 0.6 | 0.7 | 0.8 | 0.6 | 0.8 | 0.6 | 0.4 |
| Platelets, (x109/L) | 7.6 | 2.5 | 2.3 | 4.8 | 2.4 | 1.6 | 2.9 | 4.5 | 4.4 | 2.5 | 2.2 | 4.5/3.0/5.0* | 4.2 | 2.8 |
| MPV, (fL) | / | / | / | 4.6 | 1.2 | 1.0 | / | 2.5 | 4.7 | 1.5 | 1.2 | 2.5 | 1.7 | 1.1 |
| Lighter gray - meets at least one of the criteria. Darker gray – did not meet any criteria. *The SOTA criteria define limits of acceptable imprecision for three cut-off values for WBC and platelets. Ranges of values WBC based on SOTA are: L: < 1.0 x109/L, N: 1-10 x109/L and H: > 10 x109/L. Ranges of values platelets based on SOTA are: very low: 10-20 x109/L, low: ~ 50 x109/L and normal: > 50 x109/L. L– low values. N – normal values. H – high values. WBC – White Blood Cell count. RBC – Red Blood Cell count. MCV – Mean Corpuscular Volume. MPV – Mean Platelet Volume. SOTA – State-of-the-Art acceptance criteria. EuBIVAS DES – EFLM Biological Variation Database, Desirable specification. EFLM – European Federation of Clinical Chemistry and Laboratory Medicine. | ||||||||||||||
Estimated accuracy on control samples for haematology analyser Siemens Advia 2120i
|
|
| |||||
|---|---|---|---|---|---|---|
| L | N | H | SOTA | EuBIVAS | EuBIVAS | |
| WBC, (x109/L) | 1.1 | 1.5 | 1.4 | 4.4 | 7.4 | 4.9 |
| RBC, (x1012/L) | -1.7 | 0.5 | -0.4 | 3.2 | 2.6 | 1.8 |
| Haemoglobin, (g/L) | 2.1 | 1.0 | 1.7 | 1.3 | 2.4 | 1.6 |
| Haematocrit, (L/L) | -0.8 | 1.8 | 1.8 | 1.8 | 2.3 | 1.5 |
| MCV, (fL) | 0.9 | 1.4 | 2.2 | 2 | 1.4 | 0.9 |
| Platelets, (x109/L) | -4.9 | -2.7 | -4.5 | 6.4 | 7.6 | 5.0 |
| MPV, (fL) | 0.1 | -0.7 | -1.6 | / | 2.8 | 1.9 |
| Lighter gray – meets at least one of the criteria. Darker gray – did not meet any criteria. L – low values. N – normal values. H – high values. WBC – White Blood Cell count. RBC – Red Blood Cell count. MCV – Mean Corpuscular Volume. MPV – Mean Platelet Volume. SOTA – State-of-the-Art acceptance criteria. EuBIVAS DES – EFLM Biological Variation Database, Desirable specification. EFLM – European Federation of Clinical Chemistry and Laboratory Medicine. | ||||||
Estimated precision, bias and total error for Siemens Advia 2120i haematology analyser
|
|
|
|
|
| ||
|---|---|---|---|---|---|---|
|
|
|
| ||||
| WBC, (x109/L) | 2.4 (7.13) | 1.5 | 5.5 | 6.9 | 13.8 | 20.7 |
| RBC, (x1012/L) | 0.8 (4.27) | 0.5 | 1.8 | 1.9 | 3.9 | 5.8 |
| Haemoglobin, (g/L) | 1.3 (111) | 1.0 | 3.5 | 1.9 | 3.8 | 5.8 |
| Haematocrit, (L/L) | 1.3 (0.33) | 1.9 | 3.9 | 1.9 | 3.9 | 5.8 |
| MCV, (fL) | 0.8 (78) | 1.4 | 2.7 | 0.8 | 1.6 | 2.4 |
| Platelets, (x109/L) | 2.5 (214) | -2.7 | 6.8 | 4.8 | 9.7 | 14.5 |
| MPV, (fL) | 1.5 (7.2) | -0.7 | 3.2 | 1.9 | 3.8 | 5.6 |
| Gray – did not meet any criteria. WBC – White Blood Cell count. RBC – Red Blood Cell count. MCV – Mean Corpuscular Volume. MPV – Mean Platelet Volume. SOTA – State-of-the-Art acceptance criteria. EuBIVAS OPT – EFLM Biological Variation Database, Optimal specification. EuBIVAS DES – EFLM Biological Variation Database, Desirable specification. EuBIVAS MIN – EFLM Biological Variation Database, Minimum specification. EFLM – European Federation of Clinical Chemistry and Laboratory Medicine | ||||||