Literature DB >> 33270987

Evaluation and comparison of the new Mindray BC-6200 hematology analyzer with ADVIA 2120i.

Katarzyna Kulik1, Iwona Kwiecień1, Beata Chełstowska2, Elżbieta Rutkowska1, Piotr Rzepecki3.   

Abstract

BACKGROUND: The Mindray BC-6200 is a new automatic hematology analyzer that quantifies the parameters of blood morphology and leukocyte differential in five populations (5-Diff). The aim of the study was to evaluate the BC-6200 and compare it with the Siemens ADVIA 2120i analyzer.
MATERIALS AND METHODS: The comparison between BC-6200 and ADVIA 2120i analyzers was performed using 390 whole blood samples collected on K3 EDTA. For the BC-6200, the carryover effect, precision, and linearity were evaluated. 138 samples were used to assess the sensitivity and flag ability, suggesting the presence of abnormal cells such as blasts, immature granulocytes, or atypical lymphocytes. Flagging results were compared with microscopic evaluation of blood smears.
RESULTS: The BC-6200 analyzer showed a high correlation (r ≥ .97) with ADVIA 2120i for most of the compared parameters except RDW (r = .8350), MPV (r = .7634), Mon# (r = .8366), Baso# (r = .9205), and NRBC (r = .3768). The BC-6200 had better correlation with microscopic evaluation for NRBC (r = .8902) compared with ADVIA 2120i (r = .5677). The BC-6200 has shown high efficiency for flagging blasts (80.4%), immature granulocytes (80.5%), and atypical lymphocytes (69.0%).
CONCLUSION: The new Mindray BC-6200 hematology analyzer provides high measurements precision and good correlation with ADVIA 2120i for most of the morphology and 5-diff parameters.
© 2020 John Wiley & Sons Ltd.

Entities:  

Keywords:  ADVIA2120i; Mindray BC-6200; automated hematology analyzer; blood cell count; performance evaluation

Year:  2020        PMID: 33270987     DOI: 10.1111/ijlh.13418

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  1 in total

1.  Verification of a 6-part differential haematology analyser Siemens Advia 2120i.

Authors:  Helena Čičak; Vanja Radišić Biljak; Ana-Maria Šimundić
Journal:  Biochem Med (Zagreb)       Date:  2022-06-15       Impact factor: 2.515

  1 in total