Validation of methods performance for routine biochemistry analytes at Cobas 6000 analyzer series module c501.

Vesna Supak Smolcic1, Lidija Bilic-Zulle, Elizabeta Fisic.   

Abstract

INTRODUCTION: Cobas 6000 (Roche, Germany) is biochemistry analyzer for spectrophotometric, immunoturbidimetric and ion-selective determination of biochemical analytes. Hereby we present analytical validation with emphasis on method performance judgment for routine operation.
MATERIALS AND METHODS: Validation was made for 30 analytes (metabolites, enzymes, trace elements, specific proteins and electrolytes). Research included determination of within-run (N = 20) and between-run imprecision (N = 30), inaccuracy (N = 30) and method comparison with routine analyzer (Beckman Coulter AU640) (N = 50). For validation of complete analytical process we calculated total error (TE). Results were judged according to quality specification criteria given by European Working Group.
RESULTS: Within-run imprecision CVs were all below 5% except for cholesterol, triglycerides, IgA and IgM. Between-run CVs for all analytes were below 10%. Analytes that did not meet the required specifications for imprecision were: total protein, albumin, calcium, sodium, chloride, immunoglobulins and HDL cholesterol. Analytes that did not fulfill requirements for inaccuracy were: total protein, calcium, sodium and chloride. Analytes that deviated from quality specifications for total error were: total protein, albumin, calcium, sodium, chloride and IgM. Passing-Bablok regression analysis provided linear equation and 95% confidence interval for intercept and slope. Complete accordance with routine analyzer Beckman Coulter AU640 showed small number of analytes. Other analytes showed small proportional and/or small constant difference and therefore need to be adjusted for routine operation.
CONCLUSIONS: Regarding low CV values, tested analyzer has satisfactory accuracy and precision and is extremely stable. Except for analytes that are coherent on both analyzers, some analytes require adjustments of slope and intercept for complete accordance.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22135859     DOI: 10.11613/bm.2011.028

Source DB:  PubMed          Journal:  Biochem Med (Zagreb)        ISSN: 1330-0962            Impact factor:   2.313


  9 in total

1.  Validation of Method Performance of pH, PCO2, PO2, Na(+), K(+) of Cobas b121 ABG Analyser.

Authors:  Sunil Kumar Nanda; Lopamudra Ray; Asha Dinakaran
Journal:  J Clin Diagn Res       Date:  2014-06-20

2.  Low birthweight is associated with an increased risk of LADA and type 2 diabetes: results from a Swedish case-control study.

Authors:  Rebecka Hjort; Lars Alfredsson; Per-Ola Carlsson; Leif Groop; Mats Martinell; Petter Storm; Tiinamaija Tuomi; Sofia Carlsson
Journal:  Diabetologia       Date:  2015-07-25       Impact factor: 10.122

3.  Performance evaluation of the high sensitive troponin I assay on the Atellica IM analyser.

Authors:  Antonio Buño Soto; Katell Peoc'h; Tommaso Fasano; Jorge Diaz-Garzon; Bernardino González de la Presa; Valerie Chicha-Cattoir; Simone Canovi; Maria Sanz de Pedro; Nayra Rico; Tiphaine Robert; Efrem Bonelli; Pilar Fernández Calle; Aurea Mira; Guillaume Lefevre; Luigi Vecchia; Jose Luis Bedini
Journal:  Biochem Med (Zagreb)       Date:  2022-06-15       Impact factor: 2.515

4.  Verification of a 6-part differential haematology analyser Siemens Advia 2120i.

Authors:  Helena Čičak; Vanja Radišić Biljak; Ana-Maria Šimundić
Journal:  Biochem Med (Zagreb)       Date:  2022-06-15       Impact factor: 2.515

5.  Application of sigma metrics for the assessment of quality control in clinical chemistry laboratory in Ghana: A pilot study.

Authors:  Justice Afrifa; Seth A Gyekye; William K B A Owiredu; Richard K D Ephraim; Samuel Essien-Baidoo; Samuel Amoah; David L Simpong; Aaron R Arthur
Journal:  Niger Med J       Date:  2015 Jan-Feb

6.  Fatty fish consumption and risk of latent autoimmune diabetes in adults.

Authors:  J E Löfvenborg; T Andersson; P-O Carlsson; M Dorkhan; L Groop; M Martinell; T Tuomi; A Wolk; S Carlsson
Journal:  Nutr Diabetes       Date:  2014-10-20       Impact factor: 5.097

7.  Alcohol and the risk for latent autoimmune diabetes in adults: results based on Swedish ESTRID study.

Authors:  Bahareh Rasouli; Tomas Andersson; Per-Ola Carlsson; Mozhgan Dorkhan; Valdemar Grill; Leif Groop; Mats Martinell; Tiinamaja Tuomi; Sofia Carlsson
Journal:  Eur J Endocrinol       Date:  2014-08-12       Impact factor: 6.664

8.  The effect of obstructive sleep apnea on peripheral blood amino acid and biogenic amine metabolome at multiple time points overnight.

Authors:  Ott Kiens; Egon Taalberg; Viktoria Ivanova; Ketlin Veeväli; Triin Laurits; Ragne Tamm; Aigar Ottas; Kalle Kilk; Ursel Soomets; Alan Altraja
Journal:  Sci Rep       Date:  2021-05-24       Impact factor: 4.379

9.  The importance of having a flexible scope ISO 15189 accreditation and quality specifications based on biological variation--the case of validation of the biochemistry analyzer Dimension Vista.

Authors:  Pilar Fernandez-Calle; Sandra Pelaz; Paloma Oliver; Maria Jose Alcaide; Ruben Gomez-Rioja; Antonio Buno; Jose Manuel Iturzaeta
Journal:  Biochem Med (Zagreb)       Date:  2013       Impact factor: 2.313

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.