Marc Lenfant1,2, Pierre-Olivier Comby1,2, Kevin Guillen2,3, Felix Galissot1, Karim Haioun4, Anthony Thay4, Olivier Chevallier2,3, Frédéric Ricolfi1, Romaric Loffroy2,3.
Abstract
OBJECTIVE: To compare the image quality of computed tomography angiography of the supra-aortic arteries (CTSA) at different tube voltages in low doses settings with deep learning-based image reconstruction (DLR) vs. hybrid iterative reconstruction (H-IR).Entities:
Keywords: CT angiography; deep-learning reconstruction; emergency; image quality; low-dose CT scan
Year: 2022 PMID: 35626442 PMCID: PMC9142122 DOI: 10.3390/diagnostics12051287
Source DB: PubMed Journal: Diagnostics (Basel) ISSN: 2075-4418
CT acquisition parameters in each peak kilovoltage (kVp) group.
| Parameters | 120 kVp | 100 kVp | 80 kVp |
|---|---|---|---|
| Acquisition mode | Helical | Helical | Helical |
| Tube voltage (kVp) | 120 | 100 | 80 |
| Tube current range (mA) | 200–700 | 200–700 | 200–700 |
| Collimation (mm) | 0.5 × 80 | 0.5 × 80 | 0.5 × 80 |
| Rotation time (s) | 0.35 | 0.35 | 0.35 |
| Field of view (mm) | 320 | 320 | 320 |
| Slice thickness (mm) | 1 | 1 | 1 |
| Interval (mm) | 0.8 | 0.8 | 0.8 |
| Pitch | 0.8 | 0.8 | 0.8 |
| Noise index | 10 | 10 | 10 |
mA, milliampere; mm, millimeter; s, second.
Five-point Likert-scale used for qualitative image evaluation.
| Grading | Confidence | Artifacts | Sharpness | Noise |
|---|---|---|---|---|
| 1 | Non-diagnostic | Extensive | Very blurry | Very high/Very coarse graininess |
| 2 | Low confidence | Significant | Blurry | High/Coarse graininess |
| 3 | Average | Average | Average | Average |
| 4 | Good confidence | Few | Sharp | Low/Fine graininess |
| 5 | Highest confidence | Very few | Very Sharp | Very low/Very fine graininess |
Figure 1Flow chart of the study. kVp, kilovoltage peak.
Comparison of radiation dose to the patient with the three kVp values.
| Parameters | 120 kVp | 100 kVp | 80 kVp | 100 kVp vs. 120 kVp | 80 kVp vs. 100 kVp | 80 kVp vs. 120 kVp | |||
|---|---|---|---|---|---|---|---|---|---|
| % Change | % Change | % Change | |||||||
| Age | 67 ± 17.7 | 65.7 ± 19 | 63.4 ± 18.8 |
| 0.96 |
| 0.64 |
| 0.67 |
| Female/male | 14/13 | 18/12 | 21/8 |
| 0.54 |
| 0.32 |
| 0.12 |
| BMI, kg/m2 | 25.6 ± 4.6 | 25 ± 4.1 | 25.6 ± 3.3 |
| 0.55 |
| 0.48 |
| 0.99 |
| Right/left arm injection | 14/13 | 14/16 | 14/15 |
| 0.7 |
| 0.9 |
| 0.8 |
| CTDIvol, mGy | 4.8 ± 1.1 | 4 ± 1.1 | 2.3 ± 0.2 | −17% | 0.01 | −42% | <0.01 | −51% | <0.01 |
| DLP, mGy.cm | 197.1 ± 51.2 | 150.7 ± 37.8 | 90.4 ± 9.4 | −24% | <0.01 | −40% | <0.01 | −54% | <0.01 |
| ED, mSv | 1.5 ± 0.4 | 1.1 ± 0.3 | 0.68 ± 0.1 | −24% | <0.01 | −40% | <0.01 | −54% | <0.01 |
Figure 2Arterial vascular attenuation profile for the anterior cerebral vascularization with H-IR (a) and DLR (b), with 80 kVp in blue, 100 kVp in yellow and 120 kVp in orange. Mean values are displayed in bold lines. Vessel contrast for each patient are in the corresponding light colors. AO: Aortic arch; CCA: common carotid artery; ICAc: cervical internal carotid artery; ICAp: intrapetrous internal carotid artery; ICAt: terminal internal carotid artery; MCA: middle cerebral artery.
Comparison of the two reconstruction techniques at regional levels with constant kVp value.
| Parameters, | 120 kVp, 27 | 100 kVp, 30 | 80 kVp, 29 | All kVp, 86 | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| H-IR | DLR | H-IR | DLR | H-IR | DLR | H-IR | DLR | |||||
| Close to the aorta | ||||||||||||
| Image signal (HU) | 382.8 ± 115.8 | 351.1 ± 102.4 | <0.01 | 474.3 ± 127.7 | 453.6 ± 115.4 | <0.01 | 580.9 ± 124 | 595.8 ± 122.7 | <0.01 | 475.3 ± 144.3 | 465.8 ± 150.4 | <0.01 |
| Image noise (HU) | 18.9 ± 4.7 | 12.2 ± 1.9 | <0.01 | 19.4 ± 6.1 | 12 ± 2.7 | <0.01 | 23.5 ± 6.1 | 15.6 ± 3.1 | <0.01 | 20.8 ± 5.9 | 13.4 ± 3.1 | <0.01 |
| SNR | 21.9 ± 9.9 | 29.5 ± 10.2 | <0.01 | 28.1 ± 15.8 | 39.9 ± 15.3 | <0.01 | 27 ± 11.5 | 39.7 ± 12.4 | <0.01 | 25.1 ± 12.5 | 36 ± 13.3 | <0.01 |
| CNR | 20.5 ± 8.9 | 27.5 ± 10.6 | <0.01 | 25.3 ± 8.0 | 38.3 ± 14 | <0.01 | 26.5 ± 10.1 | 39.5 ± 11.2 | <0.01 | 24 ± 10.4 | 34.9 ± 12.9 | <0.01 |
| Close to bones | ||||||||||||
| Image signal (HU) | 384.2 ± 95.2 | 371.3 ± 86.4 | <0.01 | 490.2 ± 117.4 | 486 ± 117.1 | 0.2 | 612.1 ± 114 | 647.8 ± 121.4 | <0.01 | 491.3 ± 142.2 | 501.1 ± 157.2 | <0.01 |
| Image noise (HU) | 16.9 ± 4.8 | 11.7 ± 2.7 | <0.01 | 17 ± 4.6 | 11.9 ± 3.3 | <0.01 | 22.7 ± 5.5 | 15.8 ± 3.2 | <0.01 | 19.1 ± 5.7 | 13.3 ± 3.6 | <0.01 |
| SNR | 24.2 ± 8.4 | 33.4 ± 11.3 | <0.01 | 30.7 ± 11.2 | 43.1 ± 15.3 | <0.01 | 28 ± 6.8 | 42.7 ± 12.6 | <0.01 | 27.2 ± 6.8 | 39.2 ± 13.7 | <0.01 |
| CNR | 22 ± 7.8 | 30.9 ± 11.4 | <0.01 | 27.6 ± 6.3 | 40.6 ± 11.7 | <0.01 | 27.8 ± 6.2 | 44.8 ± 11.7 | <0.01 | 40.1 ± 6.2 | 57.2 ± 29.4 | <0.01 |
| Intra-dural arteries | ||||||||||||
| Image signal (HU) | 366.7 ± 79.3 | 345.1 ± 79.1 | <0.01 | 456.4 ± 106.9 | 450.1 ± 107.4 | <0.01 | 584.7 ± 107.3 | 618.7 ± 116 | <0.01 | 470.9 ± 133.2 | 472.9 ± 152.6 | 0.3 |
| Image noise (HU) | 18 ± 5.5 | 14.2 ± 4.7 | <0.01 | 18.3 ± 5 | 13.9 ± 3.5 | <0.01 | 26.3 ± 7.3 | 20.6 ± 5.5 | <0.01 | 20.9 ± 7.2 | 16.3 ± 5.6 | <0.01 |
| SNR | 22.1 ± 8.5 | 26.2 ± 8.9 | <0.01 | 26.5 ± 8.8 | 34.5 ± 12.7 | <0.01 | 23.8 ± 7.3 | 32 ± 10.1 | <0.01 | 24.1 ± 8.3 | 30.9 ± 11.2 | <0.01 |
| CNR | 21.7 ± 6.8 | 27.1 ± 10.3 | <0.01 | 26.0 ± 6.0 | 35.9 ± 9.0 | <0.01 | 24.4 ± 5.6 | 36.0 ± 9.2 | <0.01 | 29.1 ± 11.6 | 38.8 ± 13.6 | <0.01 |
| All vascular segments | ||||||||||||
| Image signal (HU) | 382.3 ± 98.6 | 358 ± 90.2 | <0.01 | 468 ± 118.5 | 456.6 ± 113.5 | <0.01 | 597.6 ± 114.9 | 625.9 ± 118.5 | <0.01 | 484.9 ± 141.7 | 482.8 ± 154.4 | 0.16 |
| Image noise (HU) | 17.5 ± 5.2 | 12.6 ± 3.6 | <0.01 | 18.1 ± 5.4 | 12.6 ± 3.2 | <0.01 | 24.4 ± 6.7 | 17.7 ± 4.9 | <0.01 | 20 ± 6.6 | 14.3 ± 4.6 | <0.01 |
| SNR | 23.8 ± 9.7 | 30.3 ± 11.1 | <0.01 | 28.4 ± 12.5 | 38.7 ± 14.7 | <0.01 | 26.4 ± 9.1 | 37.9 ± 12.3 | <0.01 | 26.3 ± 10.7 | 35.8 ± 13.4 | <0.01 |
| CNR | 22.1 ± 7.5 | 28.9 ± 9.5 | <0.01 | 26.9 ± 9.5 | 38.1 ± 13.3 | <0.01 | 26.4 ± 7.7 | 39.6 ± 10.4 | <0.01 | 25.2 ± 8.6 | 35.7 ± 12.2 | <0.01 |
Quantitative image analysis at regional levels of low-kVp protocol with DLR vs. higher-kVp protocol with DLR.
| Parameters | 120 kVp + DLR | 100 kVp + DLR | 80 kVp + DLR | 80 kVp + DLR vs. | |||
|---|---|---|---|---|---|---|---|
| 120 kVp + DLR | 100 kVp + DLR | ||||||
| % Change | % Change | ||||||
| Close to the aorta | |||||||
| Image signal (HU) | 351.1 ± 102.4 | 453.6 ± 115.4 | 595.8 ± 122.7 | +70% | <0.01 | +31% | <0.01 |
| Image noise (HU) | 12.2 ± 1.9 | 12 ± 2.7 | 15.6 ± 3.1 | +28% | <0.01 | +30% | <0.01 |
| SNR | 29.5 ± 10.2 | 39.9 ± 15.3 | 39.7 ± 12.4 | +35% | <0.01 | +0% | 0.9 |
| CNR | 27.5 ± 10.6 | 38.3 ± 14 | 39.5 ± 11.2 | +44% | <0.01 | +3% | 0.62 |
| Close to bones | |||||||
| Image signal (HU) | 371.3 ± 86.4 | 486 ± 117.1 | 647.8 ± 121.4 | +74% | <0.01 | +33% | <0.01 |
| Image noise (HU) | 11.7 ± 2.7 | 11.9 ± 3.3 | 15.8 ± 3.2 | +35% | <0.01 | +33% | <0.01 |
| SNR | 33.4 ± 11.3 | 43.1 ± 15.3 | 42.7 ± 12.6 | +28% | <0.01 | −1% | 0.68 |
| CNR | 30.9 ± 11.4 | 40.6 ± 11.7 | 44.8 ± 11.7 | +45% | <0.01 | +10% | 0.08 |
| Intra-dural arteries | |||||||
| Image signal (HU) | 345.1 ± 79.1 | 450.1 ± 107.4 | 618.7 ± 116 | +79% | <0.01 | +37% | <0.01 |
| Image noise (HU) | 14.2 ± 4.7 | 13.9 ± 3.5 | 20.6 ± 5.5 | +45% | <0.01 | +48% | <0.01 |
| SNR | 26.2 ± 8.9 | 34.5 ± 12.7 | 32 ± 10.1 | +22% | <0.01 | −9% | 0.07 |
| CNR | 27.1 ± 10.3 | 35.9 ± 9.0 | 36.0 ± 9.2 | +33% | <0.01 | +0% | 0.95 |
| All vascular segments | |||||||
| Image signal (HU) | 358 ± 90.2 | 456.6 ± 113.5 | 625.9 ± 118.5 | +75% | <0.01 | +37% | <0.01 |
| Image noise (HU) | 12.6 ± 3.6 | 12.6 ± 3.2 | 17.7 ± 4.9 | +40% | <0.01 | +40% | <0.01 |
| SNR | 30.3 ± 11.1 | 38.7 ± 14.7 | 37.9 ± 12.3 | +25% | <0.01 | −2% | 0.4 |
| CNR | 28.9 ± 9.5 | 38.1 ± 13.3 | 39.6 ± 10.4 | +37% | <0.01 | +4% | 0.76 |
AO: Aortic arch; CCA: common carotid artery; ICAc: cervical internal carotid artery; ICAp: intrapetrous internal carotid artery; ICAt: terminal internal carotid artery; MCA: middle cerebral artery; V1, V2, V3, and V4: vertebral artery, four portions; BA, basilar artery.
Quantitative image analysis at regional levels of low-kVp protocol with DLR vs. higher-kVp protocol with H-IR.
| Parameters | 120 kVp + H-IR | 100 kVp + H-IR | 80 kVp + DLR | 80 kVp + DLR vs. | |||
|---|---|---|---|---|---|---|---|
| 120 kVp + H-IR | 100 kVp + H-IR | ||||||
| % Change | % Change | ||||||
| Close to the aorta | |||||||
| Image signal (HU) | 382.8 ± 115.8 | 474.3 ± 127.7 | 595.8 ± 122.7 | +56% | <0.01 | +26% | <0.01 |
| Image noise (HU) | 18.9 ± 4.7 | 19.4 ± 6.1 | 15.6 ± 3.1 | −17% | <0.01 | −20% | <0.01 |
| SNR | 21.9 ± 9.9 | 28.1 ± 15.8 | 39.7 ± 12.4 | +81% | <0.01 | +41% | <0.01 |
| CNR | 20.5 ± 8.9 | 25.3 ± 8.0 | 39.5 ± 11.2 | +92% | <0.01 | +56% | <0.01 |
| Close to bones | |||||||
| Image signal (HU) | 384.2 ± 95.2 | 490.2 ± 117.4 | 647.8 ± 121.4 | +69% | <0.01 | +32% | <0.01 |
| Image noise (HU) | 16.9 ± 4.8 | 17 ± 4.6 | 15.8 ± 3.2 | −7% | 0.1 | −7% | 0.05 |
| SNR | 24.2 ± 8.4 | 30.7 ± 11.2 | 42.7 ± 12.6 | +76% | <0.01 | +39% | <0.01 |
| CNR | 22 ± 7.8 | 27.6 ± 6.3 | 44.8 ± 11.7 | +104% | <0.01 | +62% | <0.01 |
| Intra-dural arteries | |||||||
| Image signal (HU) | 366.7 ± 79.3 | 456.4 ± 106.9 | 618.7 ± 116 | +69% | <0.01 | +36% | <0.01 |
| Image noise (HU) | 18 ± 5.5 | 18.3 ± 5 | 20.6 ± 5.5 | +14% | <0.01 | +13% | <0.01 |
| SNR | 22.1 ± 8.5 | 26.5 ± 8.8 | 32 ± 10.1 | +45% | <0.01 | +21% | <0.01 |
| CNR | 21.7 ± 6.8 | 26.0 ± 6.0 | 36.0 ± 9.2 | +65% | <0.01 | +38% | <0.01 |
| All vascular segments | |||||||
| Image signal (HU) | 382.3 ± 98.6 | 468 ± 118.5 | 625.9 ± 118.5 | +64% | <0.01 | +34% | <0.01 |
| Image noise (HU) | 17.5 ± 5.2 | 18.1 ± 5.4 | 17.7 ± 4.9 | +1% | 0.7 | −2% | 0.3 |
| SNR | 23.8 ± 9.7 | 28.4 ± 12.5 | 37.9 ± 12.3 | +59% | <0.01 | +33% | <0.01 |
| CNR | 22.1 ± 7.5 | 26.9 ± 9.5 | 39.6 ± 10.4 | +79% | <0.01 | +47% | <0.01 |
CCA: common carotid artery; ICAc: cervical internal carotid artery; ICAp: intrapetrous internal carotid artery; ICAt: terminal internal carotid artery; MCA: middle cerebral artery; V1, V2, V3, and V4: vertebral artery, four portions; BA, basilar artery.
Qualitative image evaluation including mean scores of confidence, artifacts, sharpness, and noise with AIDR 3D (H-IR) and AiCE (DLR). Regional levels: close to the aorta (AO, CCA, V1), close to bones (ICAp, V2, and V3), and intracranial arteries (ICAt, MCA, V4, BA). CCA: common carotid artery; ICAc: cervical internal carotid artery; ICAp: intrapetrous in-ternal carotid artery; ICAt: terminal internal carotid artery; MCA: middle cerebral artery; V1, V2, V3, and V4: vertebral artery, four portions; BA, basilar artery.
| Parameters | 120 kVp | 100 kVp | 80 kVp | All kVp | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| H-IR | DLR | H-IR | DLR | H-IR | DLR | H-IR | DLR | |||||
| Close to the aorta | ||||||||||||
| Confidence | 4.3 ± 0.7 | 4.8 ± 0.4 | <0.01 | 4.2 ± 0.6 | 4.7 ± 0.5 | <0.01 | 3.6 ± 0.8 | 4.4 ± 0.6 | <0.01 | 4 ± 0.8 | 4.6 ± 0.5 | <0.01 |
| Artifacts | 3.9 ± 1.1 | 4.1 ± 1 | <0.01 | 3.9 ± 1.1 | 4.1 ± 1 | <0.01 | 2.5 ± 1.2 | 2.9 ± 1.1 | <0.01 | 3.4 ± 1.3 | 3.7 ± 1.2 | <0.01 |
| Sharpness | 3.6 ± 0.6 | 4.8 ± 0.4 | <0.01 | 3.8 ± 0.5 | 4.8 ± 0.4 | <0.01 | 3 ± 0.7 | 4.5 ± 0.5 | <0.01 | 3.4 ± 0.7 | 4.7 ± 0.5 | <0.01 |
| Noise | 3.2 ± 0.6 | 4.5 ± 0.5 | <0.01 | 3.3 ± 0.6 | 4.8 ± 0.4 | <0.01 | 3 ± 0.5 | 4.9 ± 0.3 | <0.01 | 3.1 ± 0.6 | 4.7 ± 0.4 | <0.01 |
| Close to bones | ||||||||||||
| Confidence | 4.5 ± 0.6 | 4.9 ± 0.3 | <0.01 | 4.3 ± 0.7 | 4.9 ± 0.3 | <0.01 | 4.2 ± 0.7 | 4.9 ± 0.3 | <0.01 | 4.3 ± 0.7 | 4.9 ± 0.3 | <0.01 |
| Artifacts | 4.5 ± 0.6 | 4.9 ± 0.4 | <0.01 | 4.4 ± 0.7 | 4.9 ± 0.3 | <0.01 | 4.3 ± 0.7 | 4.9 ± 0.3 | <0.01 | 4.4 ± 0.7 | 4.9 ± 0.3 | <0.01 |
| Sharpness | 3.6 ± 0.5 | 4.7 ± 0.5 | <0.01 | 3.7 ± 0.5 | 4.8 ± 0.4 | <0.01 | 3.5 ± 0.5 | 4.9 ± 0.3 | <0.01 | 3.6 ± 0.5 | 4.8 ± 0.4 | <0.01 |
| Noise | 3.4 ± 0.6 | 4.8 ± 0.4 | <0.01 | 3.6 ± 0.5 | 4.8 ± 0.4 | <0.01 | 3.4 ± 0.5 | 4.9 ± 0.3 | <0.01 | 3.5 ± 0.5 | 4.8 ± 0.4 | <0.01 |
| Intra-dural arteries | ||||||||||||
| Confidence | 4.6 ± 0.5 | 5 ± 0.2 | <0.01 | 4.4 ± 0.6 | 4.9 ± 0.3 | <0.01 | 4.3 ± 0.7 | 4.9 ± 0.4 | <0.01 | 4.4 ± 0.6 | 4.9 ± 0.3 | <0.01 |
| Artifacts | 5 ± 0.2 | 5 ± 0.2 | - | 4.9 ± 0.4 | 5 ± 0.2 | 0.3 | 4.8 ± 0.4 | 5 ± 0.2 | 0.04 | 4.9 ± 0.3 | 5 ± 0.2 | 0.02 |
| Sharpness | 3.7 ± 0.5 | 4.7 ± 0.5 | <0.01 | 3.8 ± 0.4 | 4.8 ± 0.4 | <0.01 | 3.4 ± 0.6 | 4.8 ± 0.5 | <0.01 | 3.6 ± 0.5 | 4.8 ± 0.5 | <0.01 |
| Noise | 3.4 ± 0.6 | 4.9 ± 0.4 | <0.01 | 3.6 ± 0.5 | 4.9 ± 0.3 | <0.01 | 3.4 ± 0.5 | 5 ± 0.2 | <0.01 | 3.4 ± 0.5 | 4.9 ± 0.3 | <0.01 |
| All vascular segments | ||||||||||||
| Confidence | 4.5 ± 0.6 | 4.9 ± 0.3 | <0.01 | 4.4 ± 0.7 | 4.9 ± 0.3 | <0.01 | 4 ± 0.6 | 4.9 ± 0.3 | <0.01 | 4.3 ± 0.7 | 4.9 ± 0.3 | <0.01 |
| Artifacts | 4.3 ± 0.8 | 4.6 ± 0.5 | <0.01 | 4.1 ± 0.8 | 4.4 ± 0.7 | 0.03 | 3.5 ± 0.8 | 4.1 ± 0.8 | <0.01 | 4 ± 0.8 | 4.4 ± 0.7 | <0.01 |
| Sharpness | 3.7 ± 0.5 | 4.7 ± 0.5 | <0.01 | 3.8 ± 0.4 | 4.8 ± 0.4 | <0.01 | 3.4 ± 0.6 | 4.9 ± 0.3 | <0.01 | 3.6 ± 0.5 | 4.8 ± 0.4 | <0.01 |
| Noise | 3.4 ± 0.6 | 4.8 ± 0.4 | <0.01 | 3.6 ± 0.5 | 4.8 ± 0.4 | <0.01 | 3.3 ± 0.5 | 5 ± 0.2 | <0.01 | 3.4 ± 0.5 | 4.8 ± 0.4 | <0.01 |
Figure 3(a) Curved planar reconstructed CT image, (b) Axial MIP 20mm of patients from each group for both DLR (AiCE) and H-IR (AIDR 3D) reconstructions with a constant window setting (W:1000 L:300). 120 kVp patient was a 68-year-old woman (BMI: 25.7), CTDI 4.7 mGy, DLP 196.8 mGy·cm.; 100 kVp patient was an 86-year-old woman (BMI: 25.2), CTDI 4.7 mGy, DLP 176.6 mGy·cm; 80 kVp patient was a 79-year-old woman (BMI: 23.8), CTDI 2.5 mGy, DLP 88.3 mGy·cm. 80 kVp image provided increased contrast with both DLR and H-IR. Note that DLR maintain a noticeably low noise level for the 3 groups.
Figure 4Illustration of the differences between DLR and H-IR for the vascular regions close to the aortic arch. A 10 mm, maximum intensity projection, coronal CT reconstruction (top row) and 1-mm axial CT reconstruction (bottom row). Left: AIDR 3D, hybrid iterative reconstruction. Middle: AiCE, deep learning-based reconstruction. Right: Color-mapped subtraction [AiCE-AIDR 3D] layered on the DLR image to visually highlight the differences between the two reconstruction techniques.
Figure 5Schematic representation of the low-tube voltage approach combined with deep learning reconstruction to increase the signal-to-noise ratio vs. standard-kVp and iterative reconstruction protocol.