Literature DB >> 35617375

Accuracy of rapid point-of-care antigen-based diagnostics for SARS-CoV-2: An updated systematic review and meta-analysis with meta-regression analyzing influencing factors.

Lukas E Brümmer1, Stephan Katzenschlager2, Sean McGrath3, Stephani Schmitz4, Mary Gaeddert1, Christian Erdmann5, Marc Bota6, Maurizio Grilli7, Jan Larmann2, Markus A Weigand2, Nira R Pollock8, Aurélien Macé9, Berra Erkosar9, Sergio Carmona9, Jilian A Sacks9, Stefano Ongarello9, Claudia M Denkinger1,10.   

Abstract

BACKGROUND: Comprehensive information about the accuracy of antigen rapid diagnostic tests (Ag-RDTs) for Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) is essential to guide public health decision makers in choosing the best tests and testing policies. In August 2021, we published a systematic review and meta-analysis about the accuracy of Ag-RDTs. We now update this work and analyze the factors influencing test sensitivity in further detail. METHODS AND
FINDINGS: We registered the review on PROSPERO (registration number: CRD42020225140). We systematically searched preprint and peer-reviewed databases for publications evaluating the accuracy of Ag-RDTs for SARS-CoV-2 until August 31, 2021. Descriptive analyses of all studies were performed, and when more than 4 studies were available, a random-effects meta-analysis was used to estimate pooled sensitivity and specificity with reverse transcription polymerase chain reaction (RT-PCR) testing as a reference. To evaluate factors influencing test sensitivity, we performed 3 different analyses using multivariable mixed-effects meta-regression models. We included 194 studies with 221,878 Ag-RDTs performed. Overall, the pooled estimates of Ag-RDT sensitivity and specificity were 72.0% (95% confidence interval [CI] 69.8 to 74.2) and 98.9% (95% CI 98.6 to 99.1). When manufacturer instructions were followed, sensitivity increased to 76.3% (95% CI 73.7 to 78.7). Sensitivity was markedly better on samples with lower RT-PCR cycle threshold (Ct) values (97.9% [95% CI 96.9 to 98.9] and 90.6% [95% CI 88.3 to 93.0] for Ct-values <20 and <25, compared to 54.4% [95% CI 47.3 to 61.5] and 18.7% [95% CI 13.9 to 23.4] for Ct-values ≥25 and ≥30) and was estimated to increase by 2.9 percentage points (95% CI 1.7 to 4.0) for every unit decrease in mean Ct-value when adjusting for testing procedure and patients' symptom status. Concordantly, we found the mean Ct-value to be lower for true positive (22.2 [95% CI 21.5 to 22.8]) compared to false negative (30.4 [95% CI 29.7 to 31.1]) results. Testing in the first week from symptom onset resulted in substantially higher sensitivity (81.9% [95% CI 77.7 to 85.5]) compared to testing after 1 week (51.8%, 95% CI 41.5 to 61.9). Similarly, sensitivity was higher in symptomatic (76.2% [95% CI 73.3 to 78.9]) compared to asymptomatic (56.8% [95% CI 50.9 to 62.4]) persons. However, both effects were mainly driven by the Ct-value of the sample. With regards to sample type, highest sensitivity was found for nasopharyngeal (NP) and combined NP/oropharyngeal samples (70.8% [95% CI 68.3 to 73.2]), as well as in anterior nasal/mid-turbinate samples (77.3% [95% CI 73.0 to 81.0]). Our analysis was limited by the included studies' heterogeneity in viral load assessment and sample origination.
CONCLUSIONS: Ag-RDTs detect most of the individuals infected with SARS-CoV-2, and almost all (>90%) when high viral loads are present. With viral load, as estimated by Ct-value, being the most influential factor on their sensitivity, they are especially useful to detect persons with high viral load who are most likely to transmit the virus. To further quantify the effects of other factors influencing test sensitivity, standardization of clinical accuracy studies and access to patient level Ct-values and duration of symptoms are needed.

Entities:  

Mesh:

Year:  2022        PMID: 35617375      PMCID: PMC9187092          DOI: 10.1371/journal.pmed.1004011

Source DB:  PubMed          Journal:  PLoS Med        ISSN: 1549-1277            Impact factor:   11.613


Introduction

Antigen rapid diagnostic tests (Ag-RDTs) for Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) have proven to be a cornerstone in fighting the Coronavirus Disease 2019 (COVID-19) pandemic, as they provide results quickly and are easy to use [1]. Nevertheless, the Ag-RDTs’ performance differs widely between manufacturers, the way they are performed, and the patients in which they are used [2,3]. Thus, a comprehensive synthesis of evidence on commercially available Ag-RDTs and the factors influencing their accuracy is vital to guide public health decision makers in choosing the right test for their needs [4]. Starting in October 2020, we conducted a living systematic review (available online at www.diagnosticsglobalhealth.org, updated weekly until August 31, 2021), summarizing the accuracy of commercially available Ag-RDTs reported in scientific literature. To equip public health decision makers with the latest findings, we published the results of our first review as soon as possible in February 2021 (including literature until December 15, 2020) [5]. After peer-review and including studies for 4 further months (until April 30, 2021), we published an updated review. Here, when performed as per the manufacturer’s instructions, pooled estimates of Ag-RDT sensitivity and specificity were 76.3% (95% confidence interval (CI) 73.1% to 79.2%) and 99.1% (95% CI 98.8% to 99.4). The most sensitive test was the SARS-CoV-2 Antigen Test (LumiraDx, United Kingdom; henceforth called LumiraDx) [4]. Since our last update, many additional studies have been published with a substantial increase in studies assessing asymptomatic participants, allowing for further sub-analysis of findings [3,6]. In addition, we and others found Ag-RDT sensitivity to decrease significantly in persons with lower viral load. Viral load is usually estimated through the number of cycles, i.e., the cycle threshold (Ct) value, a reverse transcriptase polymerase chain reaction (RT-PCR) needs to be performed until viral RNA can be detected, with a low Ct-value indicating a high viral load. Furthermore, sensitivity decreased in asymptomatic persons or persons with more than 7 days since symptom onset (DOS > 7) [4]. However, studies including symptomatic patients enroll persons typically within days since onset of symptoms [7], when viral load is highest [8,9]. On the contrary, studies including only asymptomatic persons have a higher chance of including persons at a later stage in the disease and thus with lower viral load. Therefore, the decrease in Ag-RDT sensitivity might only be driven by viral load, irrespective of persons’ symptom status. With the present work, we aim not only to give an updated overview on the accuracy of commercially available Ag-RDTs, but also to further explore the impact of viral load, the presence of symptoms, and testing procedure on the accuracy of Ag-RDTs.

Methods

We developed a study protocol following standard guidelines for systematic reviews [10,11], which is available in the Supporting information (S1 Text). We also completed the PRISMA checklist (S1 PRISMA Checklist) and registered the review on PROSPERO (registration number: CRD42020225140).

Search strategy

We performed a search of the databases PubMed, Web of Science, medRxiv, and bioRxiv. The search terms were developed with an experienced medical librarian (MG), using combinations of subject headings (when applicable) and text-words for the concepts of the search question, and checked against an expert assembled list of relevant papers. The main search terms were “Severe Acute Respiratory Syndrome Corona-virus 2,” “COVID-19,” “Betacoronavirus,” “Coronavirus,” and “Point of Care Testing” with no language restrictions. The full list of search terms is available in S2 Text. Also, 1 author (LEB) manually searched the website of FIND, the global alliance for diagnostics (https://www.finddx.org/sarscov2-eval-antigen/), for additional relevant studies, and search results were checked by a second author (SK). We performed the search biweekly through August 31, 2021. The last manual search of the FIND website was performed on September 10, 2021. In addition to conducting the present review, we updated our website www.diagnosticsglobalhealth.org weekly with the latest search results based on the methods outlined below.

Inclusion criteria

We included studies evaluating the accuracy of commercially available Ag-RDTs to establish a diagnosis of SARS-CoV-2 infection at the point-of-care (POC), against RT-PCR or cell culture as reference standard. We included all study populations irrespective of age, presence of symptoms, or study location. No language restrictions were applied. We considered cohort studies, nested cohort studies, case–control or cross-sectional studies, and randomized studies. We included both peer-reviewed publications and preprints. We excluded studies, in which patients were tested for the purpose of monitoring or ending quarantine. Also, publications with a population size smaller than 10 were excluded (although the size threshold of 10 is arbitrary, such small studies are more likely to give unreliable estimates of sensitivity and specificity). Analytical accuracy studies, where tests are performed on spiked samples with a known quantity of virus, were also excluded.

Index tests

Ag-RDTs for SARS-CoV-2 aim to detect infection by recognizing viral proteins (typically the SARS-CoV-2 nucleoprotein). Most Ag-RDTs dedicated for POC deployment use specific labeled antibodies attached to a nitrocellulose matrix strip (lateral flow assay) to capture and detect the viral antigen. Successful binding of the antigen to the antibodies is either detected visually by the appearance of a line on the matrix strip or through a specific reader instrument for fluorescence detection. Other POC instrument-based tests use chips or cartridges that enable an automated immunoassay testing procedure. Ag-RDTs typically provide results within 10 to 30 minutes [3].

Reference standard

Viral culture detects viable virus that is relevant for transmission but is only available in research settings. Since RT-PCR tests are more widely available and SARS-CoV-2 RNA (as reflected by RT-PCR Ct-value) highly correlates with SARS-CoV-2 antigen quantities [12], we considered RT-PCR an acceptable reference standard for the purposes of this systematic review. Where an international standard for the correlation of the viral load to the Ct-values was used, we also report the viral load [13].

Study selection and data extraction

Two reviewers (LEB and SS, LEB and CE, or LEB and MB) reviewed the titles and abstracts of all publications identified by the search algorithm independently, followed by a full-text review of those eligible, to select the articles for inclusion in the systematic review. Any disputes were solved by discussion or by a third reviewer (CMD). Studies that assessed multiple Ag-RDTs or presented results based on differing parameters (e.g., various sample types) were considered as individual data sets. At first, 4 authors (SK, CE, SS, and MB) extracted 5 randomly selected papers in parallel to align data extraction methods. Afterwards, data extraction and the assessment of methodological quality and independence from test manufacturers (see below) were performed by 1 author per paper (LEB, SK, CE, SS, or MB) and reviewed by a second (LEB, SK, SS, or MB). Any differences were resolved by discussion or by consulting a third author (CMD). The data items extracted can be found in the Supporting information (S1 Table).

Assessment of methodological quality

The quality of the clinical accuracy studies was assessed by applying the QUADAS-2 tool [14]. The tool evaluates 4 domains: study participant selection, index test, reference standard, and flow and timing. For each domain, the risk of bias is analyzed using different signaling questions. Beyond the risk of bias, the tool also evaluates the applicability of each included study to the research question for every domain. We prepared a QUADAS-2 assessment guide specific to the needs of this review, which can be found in the Supporting information (S3 Text).

Assessment of independence from manufacturers

We examined whether a study received financial support from a test manufacturer (including the free provision of Ag-RDTs), whether any study author was affiliated with a test manufacturer, and whether a respective conflict of interest was declared. Studies were judged not to be independent from the test manufacturer if at least 1 of these aspects was present; otherwise, they were considered to be independent.

Statistical analysis and data synthesis

We extracted raw data from the studies and recalculated performance estimates where possible based on the extracted data. Also, some primary studies reported the median Ct-value along with the first and third interquartile range (IQR) and/or minimum and maximum values rather than the sample mean and standard deviation. To incorporate these studies in our analyses, we applied the quantile estimation approach [15] to estimate the mean and standard deviation of the Ct-values. In an effort to use as much of the heterogeneous data as possible, the cutoffs for the Ct-value groups were relaxed by 2 to 3 points within each range. The <20 group included values reported up to ≤20, the <25 group included values reported as ≤24 or <25 or 20 to 25, and the <30 group included values from ≤29 to ≤33 and 25 to 30. The ≥25 group included values reported as ≥25 or 25 to 30, and the ≥30 group included values from ≥30 to ≥35. For the same reason, when categorizing by age, the age group <18 years (children) included samples from persons whose age was reported as <16 or <18 years, whereas the age group ≥18 years (adults) included samples from persons whose age was reported as ≥16 or ≥18 years. Also, for the symptom duration groups, the ≤7 days group included ≤4, ≤5, ≤6, 6 to 7, ≤7, and ≤9 days, and the >7 days group included >5, 6 to 10, 6 to 21, >7, and 8 to 14 days. Relaxing the boundaries for the Ct-value, age, and duration of symptoms subgroup resulted in some overlap within the respective groups. Predominant variants of concern (VoC) for each study were analyzed using the online tool CoVariants [16] with respect to the stated study period. The respective VoCs were extracted according the current WHO listing [17]. The raw data can be found in the Supporting information (S2 Table) and with more details online (https://doi.org/10.11588/data/T3MIB0). If 4 or more data sets were available with at least 20 RT-PCR-positive samples per data set for a predefined analysis, a meta-analysis was performed. We report pooled estimates of sensitivity and specificity for SARS-CoV-2 detection along with 95% CIs using a bivariate model (implemented with the “reitsma” command from the R package “mada,” version 0.5.10). Summary receiver operating characteristic (sROC) curves were created for the 2 Ag-RDTs with the highest sensitivity. In subgroup analyses (below), where papers presented data only on sensitivity, a univariate random effects inverse variance meta-analysis was performed (using the “metagen” command from the R package “meta,” version 5.1–1, and the “rma” command from the R package “metafor,” version 3.0–2). When there were fewer than 4 studies for an index test, only a descriptive analysis was performed, and accuracy ranges are reported. We prepared forest plots for the sensitivity and specificity of each test and visually evaluated the heterogeneity between studies. In addition, heterogeneity was assessed by calculating Cochran’s Q and I2 indices. Because there is no standard method taking into account the correlation between the sensitivity and specificity in bivariate models, we calculated these indices from a pooled diagnostic odds ratio using the “madauni” function from the “mada” package. However, while this was the only approach possible, we do not view it as fully statistical stringent and present the resulting Cochran’s Q and I2 only in the Supporting information (S3 Table). For the univariate models, the heterogeneity measures were obtained from the “metagen” model output directly and are reported in the results section. We predefined subgroups for meta-analysis based on the following characteristics: Ct-value range, testing procedure in accordance with manufacturer’s instructions as detailed in the instructions for use (IFU) (henceforth called IFU-conforming) versus not IFU-conforming, age (<18 versus ≥18 years), sample type, presence or absence of symptoms, symptom duration (≤7 days versus >7 days), viral load, and predominant SARS-CoV-2 variant. We also provide mean Ct-value across true positive (TP) and false negative (FN) test results. For categorization by sample type, we assessed (1) nasopharyngeal (NP) alone or combined with other (e.g., oropharyngeal [OP]); (2) OP alone; (3) anterior nasal (AN) or mid-turbinate (MT); (4) a combination of bronchoalveolar lavage and throat wash (BAL/TW); or (5) saliva. We applied multivariable linear mixed-effect meta-regression models to explore factors that affect diagnostic test sensitivity. Based on our previous analysis [4], we a priori defined an individual’s time since infection and sample type and condition as underlying factors, influencing test sensitivity through an individual’s symptom status (symptomatic versus asymptomatic), the sample’s viral load (estimated by the mean Ct-value as presented in the study for the sub cohort of interest), and the testing procedure (IFU- versus not IFU-conforming). We performed 3 different analyses, each of which obtained unadjusted and adjusted estimates (i.e., an estimate of the association between a factor and test sensitivity, holding the other covariates in the model constant) of the effect of factors on test sensitivity. In the first analysis, we estimated the direct effect of symptom status, viral load, and testing procedure on test sensitivity. For the second and third analysis, we restricted the meta-regression models to data sets of symptomatic persons due to a lack of data. Specifically, the second analysis assessed the effect of time since infection (estimated as the sample mean of symptom duration), viral load, and testing procedure on test sensitivity. The third analysis also assessed the effect of time since infection, viral load, and testing procedure on test sensitivity, but depicted the time since infection as a binary covariate of the symptom duration subgroup (≤7 versus >7 days). Further details on the implementation of the meta-regression models and the underlying casual diagrams are available in the Supporting information (Figs A and B in S4 Text). Data sets with less than 5 RT-PCR positives were excluded. We considered an effect to be statistically significant when the regression coefficient’s 95% CI did not include 0. The analyses were performed using the “metafor” R package, version 3.0–2 [18]. As recommended to investigate publication bias for diagnostic test accuracy meta-analyses, we performed the Deeks’ test for funnel-plot asymmetry [19] (using the “midas” command in Stata, version 15); a p-value < 0.10 for the slope coefficient indicates significant asymmetry.

Sensitivity analysis

Three sensitivity analyses were performed: estimation of sensitivity and specificity excluding case–control studies, estimation of sensitivity and specificity excluding not peer-reviewed studies, and estimation of sensitivity and specificity excluding studies that were potentially influenced through test manufacturers. We compared the results of each sensitivity analysis against the overall results to assess the potential bias introduced by case–control, not peer-reviewed, and manufacturer-influenced studies.

Results

Summary of studies

The systematic search resulted in 31,254 articles. After removing duplicates, 11,462 articles were screened, and 433 papers were considered eligible for full-text review. Of these, 259 were excluded because they did not present primary data or the Ag-RDT was not commercially available. For similar reasons, we also excluded 4 studies from the FIND website. A list of the studies excluded and their reason for exclusion can be found in the Supporting information (S5 Text). This left 174 studies from the systematic search [20-193] as well as further 20 studies from the FIND website [194-213] to be included in the review (Fig 1).
Fig 1

PRISMA flow diagram.

Based on Page and colleagues [214]. Ag-RDT, antigen rapid diagnostic tests; IFU, instructions for use.

PRISMA flow diagram.

Based on Page and colleagues [214]. Ag-RDT, antigen rapid diagnostic tests; IFU, instructions for use. At the end of the data extraction process, 21 studies were still in preprint form [20,21,25,51,54,59,62,69,73,78,88,104,120,125,133,164,171,172,177,178,190]. All studies included were written in English, except for 3 in Spanish [57,66,138], 1 in Turkish [99], and 1 in French [157]. Out of the 194 studies, 26 conducted a case–control study [25,36,38,70,71,76,85,88,92-94,97,98,100,107,112,139,144,148,149,155,160,170,172,186,188], while the remaining 168 were cross-sectional or cohort studies. The reference method was RT-PCR in all except 1 study, which used viral culture [139]. The 194 studies were divided into 333 data sets. Across these, 76 different Ag-RDTs were evaluated (75 lateral flow assays, of which 63 are interpreted visually and 12 required an automated, proprietary reader; 1 assay is an automated immunoassay). The most common reasons for testing were the occurrence of symptoms (98 data sets, 29.4% of data sets) and screening of asymptomatic persons with (3; 0.9%) or without (22; 6.6%) close contact to a SARS-CoV-2 confirmed case. In 142 (42.6%) of the data sets, individuals were tested due to more than 1 of the reasons mentioned and for 68 (20.4%) the reason for testing was unclear. In total, 221,878 Ag-RDTs were performed, with a mean number of samples per data set of 666 (range 15 to 22,994). The age of the individuals tested was specified for only 90,981 samples, of which 84,119 (92.5%) were from adults (age group ≥18) and 6,862 (7.5%) from children (age group <18). Symptomatic persons comprised 74,118 (33.4%) samples, while 97,982 (44.2%) samples originated from asymptomatic persons, and for 49,778 (22.4%) samples, the participant’s symptom status was not stated by the authors. The most common sample type evaluated was NP and mixed NP/OP (117,187 samples, 52.8%), followed by AN/MT (86,354 samples, 38.9%). There was substantially less testing done on the other sample types, with 3,586 (1.6%) tests done from OP samples, 1,256 (0.6%) from saliva, 219 (0.1%) from BAL/TW, and for 13,276 (6.0%) tests the type of sample was not specified in the respective studies. A summary of the tests evaluated in clinical accuracy studies, including study author and sample size, as well as study design aspects that could potentially influence test performance, such as sample type, sample condition, IFU conformity, and symptom status, can be found in the Supporting information (S2 Table). The Standard Q test (SD Biosensor, South Korea; distributed in Europe by Roche, Germany; henceforth called Standard Q) was the most frequently used with 57 (17.1%) data sets and 36,246 (16.3%) tests, while the Panbio test (Abbott Rapid Diagnostics, Germany; henceforth called Panbio) was assessed in 55 (16.5%) data sets with 38,620 (17.4%) tests performed. Detailed results for each clinical accuracy study are available in the Supporting information (S1 Fig).

Methodological quality of studies

The findings on study quality using the QUADAS-2 tool are presented in Fig 2A and 2B. In 294 (88.3%) data sets, a relevant study population was assessed. However, for only 68 (20.4%) of the data sets, the selection of study participants was considered representative of the setting and population chosen (i.e., they avoided inappropriate exclusions or a case–control design and enrollment occurred consecutive or randomly).
Fig 2

(a) Methodological quality of the clinical accuracy studies (risk of bias). (b) Methodological quality of the clinical accuracy studies (applicability).

(a) Methodological quality of the clinical accuracy studies (risk of bias). (b) Methodological quality of the clinical accuracy studies (applicability). The conduct and interpretation of the index tests were considered to have low risk of bias in 176 (52.9%) data sets (e.g., through appropriate blinding of persons interpreting the visual read-out). However, for 155 (46.5%) data sets, sufficient information to clearly judge the risk of bias was not provided. In only 151 (45.3%) data sets, the Ag-RDTs were performed according to IFU, while 138 (41.4%) were not IFU-conforming, potentially impacting the diagnostic accuracy; for 44 (13.2%) data sets, the IFU status was unclear. The most common deviations from the IFU were (1) use of samples that were prediluted in transport media not recommended by the manufacturer (113 data sets, 12 unclear); (2) use of banked samples (103 data sets, 12 unclear); and (3) a sample type that was not recommended for Ag-RDTs (8 data sets, 11 unclear). In 126 (37.8%) data sets, the reference standard was performed before the Ag-RDT, or the operator conducting the reference standard was blinded to the Ag-RDT results, resulting in a low risk of bias. In almost all other data sets (206; 61.9%), this risk could not be assessed, due to missing information and for 1 data set (0.3%) intermediate concern was raised. The applicability of the reference test was judged to be of low concern for all data sets, as viral culture or RT-PCR are considered to adequately define the target condition for the purpose of this study. In 327 (98.2%) data sets, the sample for the index test and reference test were obtained at the same time, while this was unclear in 6 (1.8%). In 227 (68.2%) data sets, the same RT-PCR assay was used as the reference of all included samples, while in 85 (25.5%) data sets, multiple RT-PCR assays were used as the reference. The RT-PCR systems used most frequently were the Cobas SARS-CoV-2 Test (Roche, Germany; used in 79 data sets [23.7%]), the Allplex 2019-nCoV Assay (Seegene, South Korea; used in 61 data sets [18.3%]), and the GeneXpert (Cepheid, United States, CA; used in 34 data sets [10.2%]). For 21 (6.3%) data sets, the RT-PCR used as reference standard was unclear. The RT-PCR system, its limit of detection (if publicly available from the manufacturer) and sample type used in each data set can be found in the Supporting information (S2 Table). Furthermore, for 19 (5.7%) data sets, there was a concern that not all selected study participants were included in the analysis. Finally, 45 (23.2%) of the studies received financial support from the Ag-RDT manufacturer. In 13 of these as well as in 2 others (in total 7.7% of all studies), employment of the authors by the manufacturer of the Ag-RDT studied was indicated. The respective studies are listed in the Supporting information (S6 Text). Overall, a competing interest was found in 47 (24.2%) of the studies. Detailed assessment of each QUADAS domain can be found in the Supporting information (S2 Fig).

Detection of SARS-CoV-2 infection

Overall, 38 data sets were excluded from the meta-analysis, as they included fewer than 20 RT-PCR positive samples. An additional 28 data sets were missing either sensitivity or specificity and were only considered for univariate analyses. The remaining 267 data sets, evaluating 198,584 tests, provided sufficient data for bivariate analysis. The results are presented in Fig 3A–3E. Detailed results for the subgroup analysis are available in the Supporting information (S3–S7 Figs).
Fig 3

(a–f) Pooled sensitivity and specificity by IFU conformity, Ct-value*, sample type, symptom status, duration of symptoms, and age.

*Low Ct-values are the RT-PCR semiquantitative correlate for a high virus concentration, only sensitivity calculated. AN, anterior nasal; CI, confidence interval; IFU, instructions for use; MT, mid-turbinate; N, number of; NP, nasopharyngeal; RT-PCR, reverse transcription polymerase chain reaction.

(a–f) Pooled sensitivity and specificity by IFU conformity, Ct-value*, sample type, symptom status, duration of symptoms, and age.

*Low Ct-values are the RT-PCR semiquantitative correlate for a high virus concentration, only sensitivity calculated. AN, anterior nasal; CI, confidence interval; IFU, instructions for use; MT, mid-turbinate; N, number of; NP, nasopharyngeal; RT-PCR, reverse transcription polymerase chain reaction. Including any test and type of sample, the pooled estimates of sensitivity and specificity were 72.0% (95% CI 69.8 to 74.2) and 98.9% (95% CI 98.6 to 99.1), respectively. When comparing IFU and non-IFU-conform testing, sensitivity markedly differed with 76.3% (95% CI 73.7 to 78.7) compared to 66.7% (95% CI 62.6 to 70.6), respectively. Pooled specificity was similar in both groups: 99.1% (95% CI 98.8 to 99.3) and 98.4% (95% CI 97.8 to 98.8), respectively (Fig 3A).

Subgroup analysis by Ct-value

We use Ct-value as a semiquantitative correlate for the sample’s viral load [12]. As a point of reference, we assume as a median conversion that a Ct-value of 25 corresponds to a viral load of 1.5 * 106 RNA copies per milliliter of transport media, but this varies between the types of RT-PCRs used for measuring viral load [144,215]. In samples with Ct-values <20, a very high estimate of sensitivity was found (97.9% [95% CI 96.9 to 98.9]). The pooled sensitivity for Ct-values <25 was markedly better at 90.6% (95% CI 88.3 to 93.0) compared to the group with Ct ≥ 25 at 54.4% (95% CI 47.3 to 61.5). A similar pattern was observed when the Ct-values were analyzed using cutoffs <30 or ≥30, resulting in an estimated sensitivity of 76.8% (95% CI 73.1 to 80.4) and 18.7% (95% CI 13.9 to 23.4), respectively (Fig 3B). When pooling Ct-value estimates for TP Ag-RDT results (TP; 5,083 samples, 69 data sets) and FN (2,390 samples, 76 data sets) Ag-RDT results, the mean Ct-values were 22.2 (95% CI 21.5 to 22.8) and 30.2 (95% CI 29.6 to 30.9), respectively (S8 Fig). Across both TP and FN samples, mean Ct-value was 26.3 (95% CI 25.5 to 27.1). This demonstrates that RT-PCR positive samples missed by Ag-RDT have a substantially lower viral load (higher Ct-value) compared to those that were detected. Individual forest plots for each data set with mean Ct-values are presented in the Supporting information (S9 Fig).

Subgroup analysis by sample type

Most data sets evaluated NP or combined NP/OP swabs (197 data sets and 104,341 samples) as the sample type for the Ag-RDT. NP or combined NP/OP swabs achieved a pooled sensitivity of 70.8% (95% CI 68.3 to 73.2) and specificity of 98.8% (95% CI 98.6 to 99.1). Data sets that used AN/MT swabs for Ag-RDTs (52 data sets and 84,020 samples) showed a summary estimate for sensitivity of 77.3% (95% CI 73.0 to 81.0) and specificity of 99.1% (95% CI 98.6 to 99.4). However, 2 studies that reported direct head-to-head comparison of NP and AN/MT samples from the same participants using the same Ag-RDT (Standard Q) reported equivalent performance [116,117]. In contrast, saliva swabs (4 data sets, 1,216 samples) showed the lowest pooled sensitivity with only 50.1% (95% CI 7.7 to 92.3) (Fig 3C). In 3 of the data sets utilizing a saliva sample, saliva was collected as whole mouth fluid (sensitivity from 8.1% [95% CI 2.7 to 17.8] to 55.6% [95% CI 35.3 to 74.5]) [24,92,154]. The fourth used a cheek swab for sample collection (sensitivity 100% [95% CI 90.3 to 100]) [55]. Due to only 3 data sets with 3,586 samples, we were not able to estimate pooled sensitivity and specificity for OP samples. Median sensitivity and specificity were 59.4% (range 50.0% to 81.0%) and 99.1% (range 99.0% to 100.0%), respectively. We were also not able to perform a subgroup meta-analysis for BAL/TW due to insufficient data, with only 1 study with 73 samples evaluating the Biocredit Covid-19 Antigen rapid test kit (RapiGEN, South Korea; henceforth called Rapigen), Panbio and Standard Q available and sensitivity ranging between 33.3% and 88.1% [155]. However, the use of BAL/TW sampling would be considered not IFU-conforming.

Subgroup analysis in symptomatic and asymptomatic participants

Within the data sets possible to meta-analyze, 55,186 (43.2%) samples were from symptomatic and 72,457 (56.8%) from asymptomatic persons. The pooled sensitivity for symptomatic persons was markedly higher compared to asymptomatic persons with 76.2% (95% CI 73.3 to 78.9) versus 56.8% (95% CI 50.9 to 62.4). Specificity was above 98.6% for both groups (Fig 3D).

Subgroup analysis comparing symptom duration

Data were analyzed for 9,470 persons from 26 data sets with symptoms less than 7 days, while for persons with symptoms ≥7 days, fewer data were available (620 persons, 13 data sets). The pooled sensitivity estimate for individuals with symptoms <7 days was 81.9% (95% CI 77.7 to 85.5), which is markedly higher than the 51.8% (95% CI 41.5 to 61.9) sensitivity for individuals tested ≥7 days from onset of symptoms (Fig 3D).

Subgroup analysis by virus variant

The 188 data sets with 153,522 samples were conducted in settings where the SARS-CoV-2 wild type was dominant. Here, sensitivity was 72.3% (95% CI 69.7 to 74.7) and specificity was 99.0% (95% CI 98.7 to 99.2). When the alpha variant (26 data sets, 19,512 samples) was the main variant, sensitivity slightly decreased to 67.0% (95% CI 58.5 to 74.5), but with overlapping CIs, and specificity remained similar (99.3% [95% CI 98.7 to 99.6]). In settings where the wild type and the alpha variant were codominant (6 data sets, 8,753 samples), sensitivity and specificity were 72.0% (95% CI 57.9 to 82.8) and 99.6% (95% CI 98.4 to 99.9), respectively. Data were also available for the Beta, Gamma, Delta, Epsilon, Eta, and Kappa variant, but too limited to meta-analyze. Of these, most data were available for the Gamma variant, with sensitivity ranging from 84.6% to 89.9% (3 data sets, 886 samples) [202,209,213]. The main virus variant for each data set is listed in the Supporting information (S2 Table). All studies included in this review were conducted before the occurrence of the Omicron variant.

Subgroup analysis by age

For adults (age group ≥18), it was possible to pool estimates across 62,433 samples, whereas the pediatric group (age group <18) included 5,137 samples. There was only a small difference with overlapping CIs in sensitivity with 74.8% (95% CI 71.5 to 77.8) and 69.8% (95% CI 61.0 to 77.3) for the adult and pediatric group, respectively. For those data sets that reported a median Ct-value per age group, the Ct-value was slightly lower in the adult (median 22.6, Q1 = 20.5, Q3 = 24.6, 48 data sets) compared to the pediatric group (median 23.2, Q1 = 20.3, Q3 = 25.2, 3 data sets). Specificity was similar in both groups with over 99% (Fig 3E).

Meta-regression

The first analysis, assessing all variables that could influence sensitivity (symptom status, testing procedure [IFU-conforming versus not IFU-conforming], and mean Ct-value), included 65 data sets of symptomatic and 18 of asymptomatic persons. The second and third analysis assessed only symptomatic persons with 28 and 50 data sets, respectively. The full list of data sets for each analysis and detailed results are available in the Supporting information (Tables A–D in S4 Text). In the first analysis, we found viral load (as estimated by Ct-value) to be the driving factor of sensitivity. Sensitivity was estimated to increase by 2.9 percentage points (95% CI 1.7 to 4.0) for every unit the mean Ct-value decreased (Table B in S4 Text), after adjusting for symptom status and testing procedure (Fig 4). In addition, sensitivity was estimated to be 20.0 percentage points (95% CI 13.7 to 26.3) higher for samples from symptomatic compared to asymptomatic participants. However, when controlling for testing procedure and mean Ct-value, this difference declined to only 11.1 percentage points (95% CI 4.8 to 17.4). The difference between IFU-conforming versus not IFU-conforming testing procedure was not significant (5.2 percentage points [95% CI ‒2.6 to 13.0] higher for IFU-conforming) after controlling for symptom status and mean Ct-value.
Fig 4

Pooled estimate of sensitivity across mean Ct-values holding symptom status and IFU-status constant at their respective means.

Dotted lines are the corresponding 95% CIs. The size of each point is a function of the weight of the data set in the model, where larger data sets have larger points. CI, confidence interval; Ct, cycle threshold; IFU, instructions for use.

Pooled estimate of sensitivity across mean Ct-values holding symptom status and IFU-status constant at their respective means.

Dotted lines are the corresponding 95% CIs. The size of each point is a function of the weight of the data set in the model, where larger data sets have larger points. CI, confidence interval; Ct, cycle threshold; IFU, instructions for use. When assessing only symptomatic participants, test sensitivity was estimated to decrease by 3.2 percentage points (95% CI ‒1.5 to 7.9) for every 1 day increase in average duration of symptoms (mean duration of symptoms ranged from 2.75 to 6.47 days). However, with the CI including the value 0, this effect was not statistically significant. When controlling for mean Ct-value and testing procedure, the estimated effect of the average duration of symptoms was close to 0 (0.7 percentage points [95% CI ‒5.0 to 6.4], Table C in S4 Text). Concordantly, for samples collected after 7 days of symptom onset sensitivity were estimated to be 22.9 percentage points (95% CI 10.3 to 35.4) lower compared to those collected within 7 days. When controlling for mean Ct-value and testing procedure, the model still showed a decrease in sensitivity for samples collected after 7 days of symptom onset, but again closer to 0 and no longer statistically significant (‒13.8 percentage points [95% CI ‒27.7 to 0.1], Table D in S4 Text).

Analysis of individual tests

Based on 179 data sets with 143,803 tests performed, we were able to perform bivariate meta-analysis of the sensitivity and specificity for 12 different Ag-RDTs (Fig 5). Across these, pooled estimates of sensitivity and specificity on all samples were 71.6% (95% CI 69.0 to 74.1) and 99.0% (95% CI 98.8 to 99.2), which were very similar to the overall pooled estimate across all meta-analyzed data sets (72.0% and 98.9%, above).
Fig 5

Bivariate analysis of 12 Ag-RDTs.

Pooled sensitivity and specificity were calculated based on reported sample size, true positives, true negatives, false positives, and false negatives. Ag-RDT, antigen rapid diagnostic test; CI, confidence interval; N, number of.

Bivariate analysis of 12 Ag-RDTs.

Pooled sensitivity and specificity were calculated based on reported sample size, true positives, true negatives, false positives, and false negatives. Ag-RDT, antigen rapid diagnostic test; CI, confidence interval; N, number of. The highest pooled sensitivity was found for the SARS-CoV-2 Antigen Test (LumiraDx, UK; henceforth called LumiraDx) and the Standard Q nasal test (SD Biosensor, South Korea; distributed in Europe by Roche, Germany; henceforth called Standard Q nasal) with 82.7% (95% CI 73.2 to 89.4) and 81.4% (95% CI 73.8 to 87.2), respectively. However, all tests except the COVID-19 Ag Respi-Strip (Coris BioConcept, Belgium; henceforth called Coris; sensitivity 48.4% [95% CI 36.1 to 61.0]) had CIs that were overlapping. The pooled specificity was above 98% for all of the tests, except for the Standard F test (SD Biosensor, South Korea; henceforth called Standard F) and LumiraDx with specificities of 97.9% (95% CI 96.9 to 98.5) and 96.9% (95% CI 94.4 to 98.3), respectively. Hierarchical summary receiver operating characteristic for LumiraDx and Standard Q nasal are available in the Supporting information (S10 Fig). For 2 Ag-RDTs, we were only able to perform a univariate analysis, due to insufficient data. Sensitivities for the COVID-19 Rapid Antigen Test Cassette (SureScreen, UK; henceforth called SureScreen V) and the Nadal COVID-19 Ag Test (Nal von Minden, Germany; henceforth called Nadal) were similar with 57.7% (95% CI 40.9 to 74.4) and 56.6% (95% CI 26.9 to 86.3), respectively (S11 Fig). Specificity was only possible to calculate for the Nadal, which was lowest throughout the per test analysis with 91.1% (95% CI 80.2 to 100). For the remaining 62 Ag-RDTs, there were insufficient numbers of data sets for a uni-or bivariate meta-analysis. However, performance estimates and factors potentially influencing these are descriptively analyzed in the Supporting information (S4 Table) for each of the 62 tests. For Panbio and Standard Q, it was also possible to pool sensitivity per Ct-value subgroup for each individual test. Panbio and Standard Q reached sensitivities of 97.2% (95% CI 95.3 to 99.2) and 98.1% (95% CI 96.3 to 99.9) for Ct-value <20, 89.8% (95% CI 85.4 to 94.3) and 92.6% (95% CI 88.5 to 96.7) for Ct-value <25 and 73.7% (95% CI 66.0 to 81.3) and 75.7% (95% CI 67.9 to 83.4) for Ct-value <30, respectively. For Ct-value ≥20 sensitivities for Panbio and Standard Q were 89.2% (95% CI 82.1 to 96.3) and 89.0% (95% CI 81.0 to 96.9), 51.2% (95% CI 39.4 to 63.0) and 56.4% (95% CI 45.1 to 67.8) for Ct-value ≥25, and 22.8% (95% CI 12.2 to 33.4) and 20.4% (95% CI 10.5 to 30.3) for Ct-value ≥30, respectively (S4A–S4F Fig). For BinaxNow (Abbott Rapid Diagnostics, Germany), LumiraDx, SD Biosensor, Standard F, Coris, and INNOVA SARS-CoV-2 Antigen Rapid Qualitative Test (Innova Medical Group, United States of America; henceforth called Innova), sufficient data to pool sensitivity was only available for certain Ct-values, which are available in the Supporting information (S4A–S4F Fig) as well. In addition, for 8 tests it was possible to calculate pooled sensitivity and specificity estimates only including data sets that conformed to the IFU. These are also listed in the Supporting information (S5 Table). In total, 31 studies accounting for 106 data sets conducted head-to-head clinical accuracy evaluations of different tests using the same sample(s) from the same participant. These data sets are outlined in the Supporting information (S2 Table). Nine studies performed their head-to-head evaluation as per IFU and on symptomatic individuals. Across 4 studies, the Standard Q nasal (sensitivity 80.5% to 91.2%) and the Standard Q (sensitivity 73.2% to 91.2%) showed a similar range of sensitivity [116,130,216]. One study reported a sensitivity of 60.4% (95% CI 54.9 to 65.6) for the Standard Q and 56.8% (95% CI 51.3 to 62.2) for the Panbio in a mixed study population of symptomatic, asymptomatic, and high-risk contact persons [190]. Another study described a sensitivity of 56.4% (95% CI 44.7 to 67.6) for the Rapigen and 52.6% (95% CI 40.9 to 64) for the SGTi-flex COVID-19 Ag (Sugentech, South Korea) [164]. One study included only very few samples and using a not IFU-conforming sample type (BAL), limiting the ability to draw conclusions from the results [155].

Publication bias

The results of the Deeks’ test for all data sets with complete results (p = 0.24), Standard Q publications (p = 0.39), Panbio publications (p = 0.81), and Lumira ( = 0.61) demonstrate no significant asymmetry in the funnel plots, which suggests no publication bias. All funnel plots are listed in the Supporting information (S12 Fig) We performed 3 sensitivity analyses including 213 data sets for non-case–control studies, 216 data sets including only peer-reviewed studies, and 190 data sets including only data sets without any manufacturer influence. When excluding case–control studies, the sensitivity and specificity remained at 71.9% (95% CI 69.4 to 74.2) and 99.0% (95% CI 98.8 to 99.2), respectively. Similarly, when assessing only peer-reviewed studies, sensitivity and specificity did not change significantly with 71.1% (95% CI 68.5 to 73.6) and 98.9% (95% CI 98.6 to 99.1), respectively. If studies that could have potentially been influenced by test manufacturers were excluded, sensitivity decreased marginally, but with overlapping CIs (sensitivity of 70.3% [95% CI 67.6 to 72.9] and specificity of 99.0% [95% CI 98.7 to 99.2]).

Discussion

After reviewing 194 clinical accuracy studies, we found Ag-RDTs to be 76.3% (95% CI 73.7 to 78.7) sensitive and 99.1% (95% CI 98.8 to 99.3) specific in detecting SARS-CoV-2 compared to RT-PCR when performed according to manufacturers’ instructions. While sensitivity was higher in symptomatic compared to asymptomatic persons, especially when persons were still within the first week of symptom onset, the main driver behind test sensitivity was a sample’s viral load. LumiraDx and Standard Q nasal were the most accurate tests, but heterogeneity in the design of the studies evaluating these tests potentially favored test specific estimates. Using Ct-value as a semiquantitative correlate for viral load, there was a significant correlation between test sensitivity and viral load, with sensitivity increasing by 2.9 percentage points for every unit decrease in mean Ct-value when controlling for symptom status and testing procedure. The pooled Ct-value for TP was on average over 8 points lower than for FN results (Ct-value of 22.2 for TP compared to 30.2 for FN results). Viral load being the deciding factor for test sensitivity confirms prior work [12]. Furthermore, sensitivity was found to be higher when samples were from symptomatic (76.2% sensitivity) compared to asymptomatic participants (56.8% sensitivity). This was confirmed in the regression model, estimating sensitivity to be 20.0 percentage points higher in samples that originated from symptomatic participants. In our previous analysis, we assumed that the increase in sensitivity is not due to the symptom status as such, but results from the fact that in symptomatic study, populations chances are higher to include participants at the beginning of the disease with high viral load [4]. In the present analysis, this assumption shows to be largely true. Controlling for Ct-value, the RT-PCR correlate for viral load, the effect of symptomatic versus asymptomatic participants on test sensitivity strongly decreased to 11.1 percentage points. As others found symptomatic and asymptomatic individuals to have the same viral load when at the same stage of the disease [8], we would have expected the regression coefficient to have decreased even further to 0. This nonzero difference in sensitivity between symptomatic and asymptomatic participants may be due to the lack of access to individual participant Ct-values, which required our analyses to control for the mean Ct-value over all participants in a data set rather than the individual Ct-values. Furthermore, some variability is likely introduced by the testing for the Ag-RDT and the RT-PCR not to occur from the sample. Therefore, some degree of residual confounding is likely present. We also found sensitivity to be higher when participants were tested within 7 days of symptom onset (81.9% sensitivity) compared to >7 days (51.8% sensitivity). Concordantly, our regression model estimated that sensitivity decreases by 3.2 percentage points for every 1-day increase in mean symptom duration. Again, this decrease in sensitivity is driven by viral load as was seen when controlling for Ct-value. Importantly, it is not yet clear how the emergence of new SARS-CoV-2 VoC and the growing vaccination coverage will affect Ag-RDTs sensitivity in the early days after symptom onset. Most of the studies included in this analysis were performed at the time the wild type and Alpha variant were circulating. Test sensitivity was slightly lower for the Alpha variant compared to the wild type (67.0% [95% CI 58.5 to 74.5] versus 72.3% [95% CI 69.7 to 74.7]). However, conclusions on differences in performance between variants are difficult to draw as between study heterogeneity was substantial and, while this does not preclude a difference between groups, CIs were widely overlapping. Furthermore, pooled sensitivity for studies where Alpha and wild type were codominant (72.0% [95% CI 57.9 to 82.8]) were similar to that of the wild type alone. Similar Ag-RDT sensitivity was also found with the Delta variant compared to wild type, and for the Omicron SARS-CoV-2 variant initial data suggests similar clinical performance as well, although analytical performance pointed toward a potentially lower performance [217-220]. Vaccination did not affect viral kinetics in the first week [221] and is unlikely to do so for the Omicron variant [222]. To further inform public health decision makers on the best strategy to apply Ag-RDTs, clinical accuracy studies in settings with high prevalence of the Omicron variant are urgently needed. Looking at specific tests, LumiraDx and Standard Q nasal showed the highest sensitivity, performing above the 80% sensitivity target defined by WHO. However, while the Standard Q nasal was 99.1% (95% CI 98.4 to 99.5) specific, the LumiraDx only had a specificity of 96.9% (95% CI 94.4 to 98.3), which is just below the WHO target of 97%. The reason for the lower specificity is unclear, particularly as independent analytical studies also confirmed the test had no cross-reactivity [106]. Sample to sample variability must be considered, particularly as the sensitivity of the index tests approaches that of the reference test. The 2 most often evaluated tests, namely Panbio (32,370 samples, sensitivity of 71.9%) and Standard Q (35,664 samples, sensitivity of 70.9%), performed slightly below the overall average. Similarly, Panbio and Standard Q were also the most extensively evaluated Ag-RDTs in the prior analysis, and with a sensitivity slightly above average [4]. Nonetheless, this updated analysis indicates that limited added value is to be expected from any further analysis of Ag-RDTs’ overall sensitivity or the sensitivity of the most widely evaluated tests. However, it will be important to continue to reassess tests’ analytical sensitivity for detection of new specific variants (e.g., Omicron). In addition, with a recent WHO guideline on self-performed Ag-RDTs having laid the scientific foundation [223], it would be of interest to further evaluate the accuracy and ease of use of self-performed Ag-RDTs, or specific characteristics of instrument-based Ag-RDTs. Furthermore, sensitivity strongly differed between studies that conducted the Ag-RDTs as per manufacturer’s instructions and those that did not (sensitivity of 66.7% for not IFU-conforming versus 76.3% for IFU-conforming). This was also reflected in our regression model, where test performance decreased when not following manufacturer’s instructions; however, this was not significant (‒5.2 percentage points [95% CI ‒13.0 to 2.6]). In regards to sample types, saliva showed a markedly lower sensitivity of 50.1%, compared to NP or AN/MT samples, confirming what we found in our previous analysis [4]. Especially in light of the current debate on whether saliva or throat swabs might be a more sensitive sample to detect the SARS-CoV-2 Omicron variant than NP or AN/MT samples [224-226], further research is urgently needed to quantify the difference in viral load resulting from different sample types and thus the effect of sample type on test sensitivity. In concordance with the above, many studies reporting an unusually low sensitivity performed the Ag-RDT not as per IFU [30,32,44,70,101,112,137,188] or used saliva samples [24,154,159,227]. However, 2 studies with IFU-conforming testing procedure on NP or AN/MT sample still showed a low sensitivity. This quite likely results from the on average low viral load in 1 study [53] and the asymptomatic study population in the other [179]. On the contrary, compared to the other studies unusual high sensitivity was found in studies where average viral load was high [49,88,148,149] or participants were mainly within the first week of symptom onset [46,58,139]. The main strength of our study lies in its living approach. The ability to update our methodology as the body of evidence grows has enabled an improved analysis. For example, while data were too heterogenous for a meta-regression during the prior analysis, with additional data sets we are now able to analyze the relationship between an Ag-RDT’s sensitivity, the samples’ Ct-value, and the participants’ symptom status in depth. Similarly, we decided to focus on clinical accuracy studies for POC Ag-RDTs in this current review as analytical accuracy studies require a dedicated approach to be comparable. Furthermore, the main results of our latest extractions are publicly available on our website. This has not only equipped public health professionals with an up-to-date overview on the current research landscape [228,229], but also led other researchers and the test manufacturers to check our data, improving the quality of our report through continuous peer-review. Nonetheless, our study is limited in that we use RT-PCR as a reference standard to assess the accuracy of Ag-RDTs, which are generally much more sensitive than Ag-RDTs [230] and might be a less appropriate reference standard than viral culture [139,231,232]. However, viral culture is available in research settings only and its validity as a true proxy of actual transmissibility is not proven; therefore, we find RT-PCR a suitable reference standard for the clinical accuracy studies included in this review. Furthermore, we fully acknowledge that Ct-value is only an estimate of viral load, and that the correlation between Ct-value and viral load varies between RT-PCR assays, potentially affecting the sensitivity and specificity of the evaluated Ag-RDTs [215]; nonetheless, we believe that the analysis of pooled Ct-value data across a very large data set is a useful strategy to understand the overall contribution of viral load to Ag-RDT performance. Moreover, we are aware that the test specific sensitivities and specificities can be influenced by differences in study design. However, we aimed to counterbalance this effect by assessing relevant aspects in study design for each study and analyzing outliers. To enhance comparability in between clinical accuracy studies, future studies should include individuals at a similar stage in the disease, use the same sample types, and adhere to the WHO standard for measuring SARS-CoV-2 viral load [13]. Finally, our study only includes literature up until August 31, 2021. Thus, we were not able to analyze information on Delta or Omicron variants, and look to future research to close this gap in literature.

Conclusions

In summary, Ag-RDTs detect most of the persons infected with SARS-CoV-2 when performed according to the manufacturers’ instructions. While this confirms the results of our previous analysis, the present analysis highlights that the sample’s viral load is the most influential factor underlying test sensitivity. Thus, Ag-RDTs can play a vital role in detecting persons with high viral load and therefore likely to be at highest risk of transmitting the virus. This holds true even in the absence of patient symptoms or differences in the duration of symptoms. To foster further research analyzing specific Ag-RDTs and the factors influencing their sensitivity in more detail, standardization of clinical accuracy studies and access to patient level Ct-value and duration of symptoms are essential.

PRISMA checklist.

(DOCX) Click here for additional data file.

Forest plots of all Ag-RDTs.

Ag-RDT, antigen rapid diagnostic test; CI, confidence interval; FN, false negative; FP, false positive; TN, true negative; TP, true positive. (PDF) Click here for additional data file.

Details of QUADAS assessment.

(PDF) Click here for additional data file.

Forest plots for subgroup analysis by Ct-values.

CI, confidence interval; Ct, cycle threshold. (PDF) Click here for additional data file.

Forest plots for subgroup analysis by Ct-values per test.

CI, confidence interval; Ct, cycle threshold. (PDF) Click here for additional data file.

Forest plots for subgroup analysis by IFU versus non-IFU.

CI, confidence interval; FN, false negative; FP, false positive; IFU, instructions for use; TN, true negative; TP, true positive. (PDF) Click here for additional data file.

Forest plots for subgroup analysis by sample type.

CI, confidence interval; FN, false negative; FP, false positive; TN, true negative; TP, true positive. (PDF) Click here for additional data file.

Forest plots for subgroup analysis by symptomatic versus asymptomatic.

CI, confidence interval. (PDF) Click here for additional data file.

Forest plot for subgroup analysis by mean Ct-values for TP and FN samples.

CI, confidence interval; Ct, cycle threshold; FN, false negative; TP, true positive. (PDF) Click here for additional data file.

Forest plots for subgroup analysis by mean Ct-values for TP and FN samples.

CI, confidence interval; Ct, cycle threshold; FN, false negative; TP, true positive. (PDF) Click here for additional data file.

HSROC curve Standard Q nasal and LumiraDx Ag-RDT.

Ag-RDT, antigen rapid diagnostic test; HSROC, Hierarchical summary receiver-operating characteristic. (PDF) Click here for additional data file.

Forest plot for univariate analysis for Nadal and SureScreen V.

CI, confidence interval. (PDF) Click here for additional data file.

Funnel plots for all, LumiraDx, Panbio, and Standard Q studies.

(PDF) Click here for additional data file.

List of parameters extracted from studies.

(XLSX) Click here for additional data file.

Summary of tests.

(XLSX) Click here for additional data file.

Overall and sensitivity analysis.

(XLSX) Click here for additional data file.

Tests analyzed descriptively not included in meta-analysis.

(XLSX) Click here for additional data file.

Test specific IFU analysis.

(XLSX) Click here for additional data file.

Study protocol submitted to PROSPERO.

(DOCX) Click here for additional data file.

Search strategy.

(DOCX) Click here for additional data file.

QUDAS-2 assessment interpretation guide.

(DOCX) Click here for additional data file.

Details meta-regression.

(DOCX) Click here for additional data file.

List of studies excluded.

(DOCX) Click here for additional data file.

Studies potentially influenced by the test manufacturer.

(DOCX) Click here for additional data file. 10 Feb 2022 Dear Dr Denkinger, Thank you for submitting your manuscript entitled "Accuracy of rapid point-of-care antigen-based diagnostics for SARS-CoV-2: an updated systematic review and meta-analysis with meta regression analyzing influencing factors" for consideration by PLOS Medicine. Your manuscript has now been evaluated by the PLOS Medicine editorial staff and I am writing to let you know that we would like to send your submission out for external assessment. However, we first need you to complete your submission by providing the metadata that is required for full assessment. To this end, please login to Editorial Manager where you will find the paper in the 'Submissions Needing Revisions' folder on your homepage. Please click 'Revise Submission' from the Action Links and complete all additional questions in the submission questionnaire. Please re-submit your manuscript within two working days, i.e. by Feb 14 2022 11:59PM. Login to Editorial Manager here: https://www.editorialmanager.com/pmedicine Once your full submission is complete, your paper will undergo a series of checks in preparation for full assessment. Feel free to email us at plosmedicine@plos.org if you have any queries relating to your submission. Kind regards, Richard Turner, PhD Senior Editor, PLOS Medicine plosmedicine@plos.org 8 Mar 2022 Dear Dr. Denkinger, Thank you very much for submitting your manuscript "Accuracy of rapid point-of-care antigen-based diagnostics for SARS-CoV-2: an updated systematic review and meta-analysis with meta regression analyzing influencing factors" (PMEDICINE-D-22-00454R1) for consideration at PLOS Medicine. Your paper was discussed with an academic editor with relevant expertise and sent to independent reviewers, including a statistical reviewer. The reviews are appended at the bottom of this email and any accompanying reviewer attachments can be seen via the link below: [LINK] In light of these reviews, we will not be able to accept the manuscript for publication in the journal in its current form, but we would like to invite you to submit a revised version that addresses the reviewers' and editors' comments fully. You will appreciate that we cannot make a decision about publication until we have seen the revised manuscript and your response, and we expect to seek re-review by one or more of the reviewers. In revising the manuscript for further consideration, your revisions should address the specific points made by each reviewer and the editors. Please also check the guidelines for revised papers at http://journals.plos.org/plosmedicine/s/revising-your-manuscript for any that apply to your paper. In your rebuttal letter you should indicate your response to the reviewers' and editors' comments, the changes you have made in the manuscript, and include either an excerpt of the revised text or the location (eg: page and line number) where each change can be found. Please submit a clean version of the paper as the main article file; a version with changes marked should be uploaded as a marked up manuscript. In addition, we request that you upload any figures associated with your paper as individual TIF or EPS files with 300dpi resolution at resubmission; please read our figure guidelines for more information on our requirements: http://journals.plos.org/plosmedicine/s/figures. While revising your submission, please upload your figure files to the PACE digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email us at PLOSMedicine@plos.org. We hope to receive your revised manuscript by Mar 29 2022 11:59PM. Please email us (plosmedicine@plos.org) if you have any questions or concerns. ***Please note while forming your response, if your article is accepted, you may have the opportunity to make the peer review history publicly available. The record will include editor decision letters (with reviews) and your responses to reviewer comments. If eligible, we will contact you to opt in or out.*** We ask every co-author listed on the manuscript to fill in a contributing author statement, making sure to declare all competing interests. If any of the co-authors have not filled in the statement, we will remind them to do so when the paper is revised. If all statements are not completed in a timely fashion this could hold up the re-review process. If new competing interests are declared later in the revision process, this may also hold up the submission. Should there be a problem getting one of your co-authors to fill in a statement we will be in contact. YOU MUST NOT ADD OR REMOVE AUTHORS UNLESS YOU HAVE ALERTED THE EDITOR HANDLING THE MANUSCRIPT TO THE CHANGE AND THEY SPECIFICALLY HAVE AGREED TO IT. You can see our competing interests policy here: http://journals.plos.org/plosmedicine/s/competing-interests. Please use the following link to submit the revised manuscript: https://www.editorialmanager.com/pmedicine/ Your article can be found in the "Submissions Needing Revision" folder. To enhance the reproducibility of your results, we recommend that you deposit your laboratory protocols in protocols.io, where a protocol can be assigned its own identifier (DOI) such that it can be cited independently in the future. Additionally, PLOS ONE offers an option to publish peer-reviewed clinical study protocols. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols Please ensure that the paper adheres to the PLOS Data Availability Policy (see http://journals.plos.org/plosmedicine/s/data-availability), which requires that all data underlying the study's findings be provided in a repository or as Supporting Information. For data residing with a third party, authors are required to provide instructions with contact information for obtaining the data. PLOS journals do not allow statements supported by "data not shown" or "unpublished results." For such statements, authors must provide supporting data or cite public sources that include it. Please let me know if you have any questions, and we look forward to receiving your revised manuscript. Sincerely, Richard Turner, PhD Senior editor, PLOS Medicine rturner@plos.org ----------------------------------------------------------- Requests from the editors: Please update the search to the most recent date feasible. Our academic editor commented: "the variant specific analyses one of the reviewers mentioned may be plausible with alpha, delta, and omicron estimates if up to date data are available (would be interesting to see whether there is effect modification depending on what was the predominant variant at the time of testing)". Please add a new final sentence to the "Methods and findings" subsection of your abstract, which should begin "Study limitations include ..." or similar and should quote 2-3 of the study's main limitations. In the abstract and throughout the paper, please quote p values alongside 95% CI, where available. After the abstract, please add a new and accessible "Author summary" section in non-identical prose. You may find it helpful to consult one or two recently-published research papers in PLOS Medicine to get a sense of the preferred style. Please add some additional detail in the Methods section (main text) regarding the methods for extracting studies and how disagreement between researchers was resolved, for example. At line 162, we note that heterogeneity was estimated "visually". Can statistical analyses be reported? Please restructure the start of the Discussion section (main text), as the first paragraph should summarize the study's main findings. Throughout the text, please remove spaces from within reference call-outs (e.g., "... in which they are used [2,3]."). In the reference list, please convert all italic and boldface text into plain text. Please list 6 authors names followed by "et al.", where appropriate. Noting reference 181, please use the journal name abbreviation "PLoS ONE". Noting reference 222 and others, please ensure that all citations have full access details. Comments from the reviewers: *** Reviewer #1: The authors describe an update/publication of their living systematic review of antigen tests, with particular emphasis on summarizing/clarifying results on subsets of patient populations. This work is important, comprehensive, and well written, and the technical details of the meta-analysis are appropriate (I caution that while I have a fairly strong grasp of the mathematics and techniques described, I would not consider myself a meta-analysis expert). The sensitivity analysis to exclude papers that might have been un/intentionally "bought" by manufacturers is a nice touch. In my opinion it is thus worthy of publication. I recommend acceptance with only minor revisions (below). Major points -------------- The only substantive comment regards viral load. As the authors state, and as is well known, viral load ranges widely over the course of infection and among infected individuals. Unfortunately, and to some consternation, even after two-plus years, viral loads are still not reported clinically, as they are for numerous other infections (for example, HIV, HCV), for various reasons, or in studies, where the quantitative measure used is the Ct value. This is problematic, as Ct value ranges can differ from one reference test to another, sometimes markedly. To their credit, the authors of this study are well aware of this issue, discuss it ably, and and use viral loads in the cases where they are available (I feel their discussion in 554-555 is perhaps a bit more defensive-sounding than it needs be: Ct value is essentially the only available quantitation at present, so however imperfect, it is completely reasonable to do; I would be surprised if there were pushback to that). However: --it would be of great utility to name the specific RT-qPCR assay/platform used in each assay, where available, in a supplemental document. (Apologies if this was done and I missed it; I looked ca. L283-286.) I suspect this would explain some of the outliers in Fig. 4. Ideally, the authors would add platform as a regression variable, but I respect that this might be too much to ask, so would not require redoing the regression with platform as a condition of publication. The effect of differences in limits of detection of these platforms on sensitivity/specificity/concordance between tests should be mentioned (illustrated well in Fig. 2 of https://pubmed.ncbi.nlm.nih.gov/34076471/ , which should be cited). --it would be useful (va. L308-319) to give the reader even a rough sense of what viral loads these Ct value cutoffs most likely correspond to, on average, if possible. If not possible, a sentence in this section declaring that to be the case, explaining that Ct scales vary (their reference 3 is a suitable citation). My hope is the field will at some point move to viral loads, and it would wonderful to "future-proof" this excellent review, simply by adding such a Ct-to-viral-load conversion anchor (even a rough one, for example, plus-or-minus two or three orders of magnitude). Hopefully these (and the minor comments below) will be straightforward points for the authors to address, further strengthening this exceptional study ahead of publication. Minor comments ------------------ Given that it is known what the predominant strain was in each geography (for example https://github.com/hodcroftlab/covariants), is it not possible to list the likely strain for each study? That would offer the opportunity to regress by strain in the future. L99-110, PRISMA checklist: I missed whether the search was done independently by two individuals. This may be a requirement of MOOSE but not PRISMA, or other journals, but not PRISMA/PLoS Medicine. If it was done, please mention. If it was not, that is also fine. (I am *not* suggesting it be done, if it wasn't, as a condition for publication.) L228-230: Were the non-English-language papers included? I assume yes, but please clarify. L151-155: I did not quite follow whether this "relaxing" allowed/resulted in overlap, or just aligned bin edges. Perhaps the authors could clarify. L302-303: the 95% CIs overlap. Would the authors comment on the probability that this increase is real? I ask especially since the 95% CIs do not overlap for the specificities, yet the authors do not claim that the specificity was lower when using the IFU (which would not be unexpected, given a lower sensitivity). The comparator adjective should be used similarly: I would suggest either these are both comparable, or the first is comparable (since the 95% CIs overlap) and the second is lower. I would suggest they are both comparable. L328-331: It is interesting that saliva swabs showed the lowest sensitivity, since saliva has been shown to be comparable to NP as a specimen type (https://pubmed.ncbi.nlm.nih.gov/34406838/). The 95%CI range for saliva in Fig. 3 raises the same question (5% to 95%!). Was there/was it possible to detect a difference between the swabbed saliva and the whole-mouth saliva samples? Fig. 5: Given the narrow range of specificities, might the x-axis be usefully contracted to 0.85 or 0.9-1.0, so that differences might be visualized? L551-552: Would also cite https://www.medrxiv.org/content/10.1101/2021.12.22.21268274v1 *** Reviewer #2: Brümmer and colleagues present a systematic review and meta-analysis of the diagnostic accuracy of rapid antigen tests for SARS-CoV-2 compared to a gold standard RT-PCR. Across 194 studies that could be meta-analyzed, the overall sensitivity was 72% and specificity was 98.9%. The study builds upon a previous living systematic review by adding additional studies, though overall findings remain near identical. The added value of this study is in its meta-regression and stratified analyses, which help clarify the impact of various factors of Ag-RDT sensitivity/specificity. The methods used are rigorous and search thorough. I do have some suggestions for the authors to better describe the performance of Ag-RDTs: * I believe the IFU evaluation is valuable, but think there is likely more that can be done, if the data is available. The value of Ag-RDTs can only be fully realized if they can be performed without aid or observation of health workers. The authors specifically mention on lines 523-524 that future research could evaluate "Ag-RDTs that are self-performed or those that are instrument-based." It would be interesting to know if any of the 194 studies were self-performed and/or self-read, and if further stratified analyses could be done to evaluate if sensitivity is impacted (as compared to health care professional performed and read). This would better clarify the role and impact of Ag-RDTs as a public health tool. * In the same vein, performance of Ag-RDT may be impacted by circulating variants. Were data available to evaluate this (e.g., for wild type, alpha, delta) within the existing cohorts in the review (understanding this study was done pre-omicron)? From the discussion section, it is clear that most studies were done during wild-type or alpha dominance. Proxies could be used such as dominating variant circulating in the study setting during the study period (which I believe is defensible as study mean age and Ct were used in meta-regression), and would permit comparison. * With respect to specific brands of Ag-RDT being the best performers, I was left wondering if studies that used these Ag-RDT also contained participants with characteristics that increased sensitivity (e.g., IFU-conforming, symptomatic patients, early in symptoms, lower Ct), which would limit our ability to conclude which test may have highest sensitivity. I think more could be done to contextualize these findings, particularly in the discussion. Forgive me if I missed this information. * What were the specimens collected for RT-PCR for the reference standard across studies? I was unable to find this information, however I may have missed it in the dense supplement. There are differences in sensitivity between specimen types for RT-PCR, so if there were different specimens used for the reference standard across studies this should be clear and evaluated. * I believe there may be typos with respect to number of included studies and data sets throughout the manuscript (e.g., Fig 1 states 194 studies included, line 216 says 174). Please review. *** Reviewer #3: Alex McConnachie, Statistical Review Brummer et al present the latest update from the living review of rapid antigen tests for SARS-Cov-2. The analysis now includes meta regression, looking at factors associated with diagnostic performance. This review considers the statistical element of the paper. Generally, the statistical methods are good, with the results presented and interpreted appropriately. I have a few comments, which are fairly minor. As a non-clinical reader, I would have appreciated an explanation of Ct values, in the sense that a low Ct value indicates a higher viral load, earlier in the paper, or even in the abstract. The statistical methods section of the paper is quite good, but does not mention the use of summary ROC curves, when perhaps it should. Also, line 181 mentions multivariate regression of factors associated with sensitivity; "multivariable" is a more accurate term. There is mention of the assessment of publication bias using Deeks' test for funnel plot asymmetry. However, I could not find any funnel plots in the paper or the supplement, nor any results of these tests. Why are these not reported? Also, assessment of publication bias is usually not recommended when the number of studies is small; did the authors have specific criteria in terms of the number of studies, when it came to the assessment of publication bias? Most research studies require a justification for the sample size of the study. For meta analyses, this is generally not an issue. However, in this case, since this is a living review, the authors have been able to see the data accumulating over time, so perhaps some sort of sample size justification is warranted. Why do the analysis now? Were there pre-defined criteria that triggered this analysis? 3 sensitivity analyses are reported. What is the rationale for doing these as sensitivity analyses (i.e. repeating the analysis after excluding a group of studies) rather than as subgroup analyses (i.e. estimating the difference in sensitivity and specificity estimates obtained from studies of different types)? In Figures 3 and 5, it is noticeable that the pooled specificity estimates are all very high. So high, in fact, that it is very difficult to see what is going on, as all the points and confidence intervals are bunched up to the right of the figure. I can understand that there is a need to keep the axes the same, for comparability, but would it help to truncate the x-axis in these figures, e.g. with a lower range of 90%? A lower limit can be used in the supplement, but for the main body of the paper, a higher value perhaps makes sense. *** Reviewer #4: Estimated Authors, I've read the present study with great interest. Following a living review of POC antigen-based test for SARS-CoV-2 (updated until August 2021), Authors were able to summarise available evidence on sensitivity and specificity of such instruments. Because of the reduced turnaround time, as well as the option to perform such tests in settings other than hospitals and medical laboratories, POC antigen-based test represent a valuable asset in in our global efforts against SARS-CoV-2 pandemic. Unfortunately, as previously stressed by several studies, the actual sensitivity of these POC tests may be quite inappropriate to fulfil the target of promptly and accurately identify incident cases of SARS-CoV-2 infections. With a sensitivity of around 70%, a large share of cases may fail to be diagnosed, particularly in cases with a low to moderate replication of the pathogen, and in early stages of infections. Authors, through an appropriate study design and an accurate and diligent application of a proper methodology were able to provide a valuable piece of information for all professionals interested in this specific topic. Briefly, I've neither requests or recommendations for improving this paper, whose acceptance "as it is" I strongly recommend. *** Any attachments provided with reviews can be seen via the following link: [LINK] 30 Mar 2022 Submitted filename: Point-to-point response.docx Click here for additional data file. 22 Apr 2022 Dear Dr. Denkinger, Thank you very much for re-submitting your manuscript "Accuracy of rapid point-of-care antigen-based diagnostics for SARS-CoV-2: an updated systematic review and meta-analysis with meta regression analyzing influencing factors" (PMEDICINE-D-22-00454R2) for consideration at PLOS Medicine. I have discussed the paper with our academic editor and it was also seen again by three reviewers. I am pleased to tell you that, once the remaining editorial and production issues are fully dealt with, we expect to be able to accept the paper for publication in the journal. The remaining issues that need to be addressed are listed at the end of this email. Any accompanying reviewer attachments can be seen via the link below. Please take these into account before resubmitting your manuscript: [LINK] ***Please note while forming your response, if your article is accepted, you may have the opportunity to make the peer review history publicly available. The record will include editor decision letters (with reviews) and your responses to reviewer comments. If eligible, we will contact you to opt in or out.*** In revising the manuscript for further consideration here, please ensure you address the specific points made by each reviewer and the editors. In your rebuttal letter you should indicate your response to the reviewers' and editors' comments and the changes you have made in the manuscript. Please submit a clean version of the paper as the main article file. A version with changes marked must also be uploaded as a marked up manuscript file. Please also check the guidelines for revised papers at http://journals.plos.org/plosmedicine/s/revising-your-manuscript for any that apply to your paper. If you haven't already, we ask that you provide a short, non-technical Author Summary of your research to make findings accessible to a wide audience that includes both scientists and non-scientists. The Author Summary should immediately follow the Abstract in your revised manuscript. This text is subject to editorial change and should be distinct from the scientific abstract. We hope to receive your revised manuscript within 1 week. Please email us (plosmedicine@plos.org) if you have any questions or concerns. We ask every co-author listed on the manuscript to fill in a contributing author statement. If any of the co-authors have not filled in the statement, we will remind them to do so when the paper is revised. If all statements are not completed in a timely fashion this could hold up the re-review process. Should there be a problem getting one of your co-authors to fill in a statement we will be in contact. YOU MUST NOT ADD OR REMOVE AUTHORS UNLESS YOU HAVE ALERTED THE EDITOR HANDLING THE MANUSCRIPT TO THE CHANGE AND THEY SPECIFICALLY HAVE AGREED TO IT. Please ensure that the paper adheres to the PLOS Data Availability Policy (see http://journals.plos.org/plosmedicine/s/data-availability), which requires that all data underlying the study's findings be provided in a repository or as Supporting Information. For data residing with a third party, authors are required to provide instructions with contact information for obtaining the data. PLOS journals do not allow statements supported by "data not shown" or "unpublished results." For such statements, authors must provide supporting data or cite public sources that include it. To enhance the reproducibility of your results, we recommend that you deposit your laboratory protocols in protocols.io, where a protocol can be assigned its own identifier (DOI) such that it can be cited independently in the future. Additionally, PLOS ONE offers an option to publish peer-reviewed clinical study protocols. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. Please note, when your manuscript is accepted, an uncorrected proof of your manuscript will be published online ahead of the final version, unless you've already opted out via the online submission form. If, for any reason, you do not want an earlier version of your manuscript published online or are unsure if you have already indicated as such, please let the journal staff know immediately at plosmedicine@plos.org. Please let me know if you have any questions, and we look forward to receiving the revised manuscript. Sincerely, Richard Turner, PhD Senior Editor, PLOS Medicine rturner@plos.org ------------------------------------------------------------ Requests from Editors: Our academic editor requests that you state in the paper, as a limitation, perhaps, that the absence of literature post-August 2021 means that information on testing for the Delta and Omicron variants, for example, is lacking. At line 81, please make that "... proved to be" or similar. At line 649, we suggest "... enabled an improved analysis". Please add an institutional author name to references 1 & 3, and any others. Please spell out the group author name in reference 10 and any other relevant references. Comments from Reviewers: *** Reviewer #1: Overall the authors have addressed all the points brought up in my review, further strengthening this major contribution to the literature. The variant discussion in particular was excellent. I have now just the following minor points; if the authors make these, I recommend acceptance. In the interest of time, I would not need to see a revision again before publication (I trust the authors/editors). L385: Would suggest the following change (additions in ALL CAPS): "Can be assumed to be" --> "AS A POINT OF REFERENCE, we assume AS A MEDIAN CONVERSION that Ct value of 25 corresponds to a viral load of 1.5 * 10^6 RNA copies per milliliter OF TRANSPORT MEDIA, but [rest of sentence and refs are fine]". Gives a better idea of why the assumption was made. L222-223: please cite Covariants.org as follows, per the author's recommendation (https://covariants.org/faq#how-should-i-cite-or-acknowledge-this-work">https://covariants.org/faq#how-should-i-cite-or-acknowledge-this-work): Emma B. Hodcroft. 2021. "CoVariants: SARS-CoV-2 Mutations and Variants of Interest." https://covariants.org/. This will help the author rationalize to funding agencies the value of this work, keeping it available to the scientific community going forward. L662 (or possibly L175-177): It is worth noting explicitly that RT-PCR tests are generally much more sensitive than antigen tests, which fits nicely with the better correlation at lower Ct values (most COVID-19-experienced readers will know this, but safer to not assume that future/all readers will). Please cite https://academic.oup.com/cid/article/73/9/e3042/6127024 as to the importance of LoD. Table S2: Would add a column for the limit of detection of the comparator and a column for the limit of detection of the Ag-RDT (I went through all the supplements; apologies if I missed this). These are also usefully mentioned where appropriate in L360-362. *** Reviewer #2: I'd like to thank the authors for thoroughly revising their manuscript based on the editor and reviewer comments. My only remaining comment is for the authors to potentially re-consider how they speak about overlapping confidence intervals - which are not a perfect method to state two measures are not significantly different (see: https://www.tandfonline.com/doi/abs/10.1198/000313001317097960 and https://cscu.cornell.edu/wp-content/uploads/73_ci.pdf). The authors can maintain their messaging even with these statements on overlapping CIs removed. Currently, there is an implication that differences in sensitivity do not exist between tests since confidence intervals overlap - though I think the authors intend to say there is heterogeneity. *** Reviewer #3: Alex McConnachie, Statistical Review I thank the authors for their consideration of my previous comments. I am happy with the responses, and have no further comments to make. *** Any attachments provided with reviews can be seen via the following link: [LINK] 30 Apr 2022 Submitted filename: Point-by-point response.docx Click here for additional data file. 4 May 2022 Dear Dr Denkinger, On behalf of my colleagues and the Academic Editor, Dr Suthar, I am pleased to inform you that we have agreed to publish your manuscript "Accuracy of rapid point-of-care antigen-based diagnostics for SARS-CoV-2: an updated systematic review and meta-analysis with meta regression analyzing influencing factors" (PMEDICINE-D-22-00454R3) in PLOS Medicine. Before your manuscript can be formally accepted you will need to complete some formatting changes, which you will receive in a follow up email. Please be aware that it may take several days for you to receive this email; during this time no action is required by you. Once you have received these formatting requests, please note that your manuscript will not be scheduled for publication until you have made the required changes. In the meantime, please log into Editorial Manager at http://www.editorialmanager.com/pmedicine/, click the "Update My Information" link at the top of the page, and update your user information to ensure an efficient production process. PRESS We frequently collaborate with press offices. If your institution or institutions have a press office, please notify them about your upcoming paper at this point, to enable them to help maximise its impact. If the press office is planning to promote your findings, we would be grateful if they could coordinate with medicinepress@plos.org. If you have not yet opted out of the early version process, we ask that you notify us immediately of any press plans so that we may do so on your behalf. We also ask that you take this opportunity to read our Embargo Policy regarding the discussion, promotion and media coverage of work that is yet to be published by PLOS. As your manuscript is not yet published, it is bound by the conditions of our Embargo Policy. Please be aware that this policy is in place both to ensure that any press coverage of your article is fully substantiated and to provide a direct link between such coverage and the published work. For full details of our Embargo Policy, please visit http://www.plos.org/about/media-inquiries/embargo-policy/. To enhance the reproducibility of your results, we recommend that you deposit your laboratory protocols in protocols.io, where a protocol can be assigned its own identifier (DOI) such that it can be cited independently in the future. Additionally, PLOS ONE offers an option to publish peer-reviewed clinical study protocols. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols Thank you again for submitting to PLOS Medicine. We look forward to publishing your paper. Sincerely, Richard Turner, PhD Senior Editor, PLOS Medicine rturner@plos.org
  185 in total

1.  Comparison of four commercial, automated antigen tests to detect SARS-CoV-2 variants of concern.

Authors:  Andreas Osterman; Maximilian Iglhaut; Andreas Lehner; Patricia Späth; Marcel Stern; Hanna Autenrieth; Maximilian Muenchhoff; Alexander Graf; Stefan Krebs; Helmut Blum; Armin Baiker; Natascha Grzimek-Koschewa; Ulrike Protzer; Lars Kaderali; Hanna-Mari Baldauf; Oliver T Keppler
Journal:  Med Microbiol Immunol       Date:  2021-08-20       Impact factor: 3.402

2.  Diagnostic Accuracy of the Panbio Severe Acute Respiratory Syndrome Coronavirus 2 Antigen Rapid Test Compared with Reverse-Transcriptase Polymerase Chain Reaction Testing of Nasopharyngeal Samples in the Pediatric Population.

Authors:  Serena Villaverde; Sara Domínguez-Rodríguez; Gema Sabrido; Conchita Pérez-Jorge; Marta Plata; María Pilar Romero; Carlos Daniel Grasa; Ana Belén Jiménez; Elena Heras; Antonio Broncano; María Del Mar Núñez; Marta Illán; Paloma Merino; Beatriz Soto; David Molina-Arana; Amanda Bermejo; Pablo Mendoza; Manuel Gijón; Begoña Pérez-Moneo; Cinta Moraleda; Alfredo Tagarro
Journal:  J Pediatr       Date:  2021-01-21       Impact factor: 4.406

3.  Clinical application of a rapid antigen test for the detection of SARS-CoV-2 infection in symptomatic and asymptomatic patients evaluated in the emergency department: A preliminary report.

Authors:  Gianni Turcato; Arian Zaboli; Norbert Pfeifer; Laura Ciccariello; Serena Sibilio; Giovanna Tezza; Dietmar Ausserhofer
Journal:  J Infect       Date:  2020-12-19       Impact factor: 6.072

4.  Comparative diagnostic performance of different rapid antigen detection tests for COVID-19 in the real-world hospital setting.

Authors:  Bianca Bruzzone; Vanessa De Pace; Patrizia Caligiuri; Valentina Ricucci; Giulia Guarona; Beatrice M Pennati; Simona Boccotti; Andrea Orsi; Alexander Domnich; Giorgio Da Rin; Giancarlo Icardi
Journal:  Int J Infect Dis       Date:  2021-04-27       Impact factor: 3.623

5.  Implementing SARS-CoV-2 Rapid Antigen Testing in the Emergency Ward of a Swiss University Hospital: The INCREASE Study.

Authors:  Giorgia Caruana; Antony Croxatto; Eleftheria Kampouri; Antonios Kritikos; Onya Opota; Maryline Foerster; René Brouillet; Laurence Senn; Reto Lienhard; Adrian Egli; Giuseppe Pantaleo; Pierre-Nicolas Carron; Gilbert Greub
Journal:  Microorganisms       Date:  2021-04-10

6.  SARS-CoV-2 rapid antigen test: Fast-safe or dangerous? An analysis in the emergency department of an university hospital.

Authors:  Carola Holzner; Dirk Pabst; Olympia Evdoxia Anastasiou; Ulf Dittmer; Randi Katrin Manegold; Joachim Risse; David Fistera; Clemens Kill; Maximilian Falk
Journal:  J Med Virol       Date:  2021-05-09       Impact factor: 20.693

7.  Comparison of SARS-CoV-2 Reverse Transcriptase Polymerase Chain Reaction and BinaxNOW Rapid Antigen Tests at a Community Site During an Omicron Surge : A Cross-Sectional Study.

Authors:  John Schrom; Carina Marquez; Genay Pilarowski; Chung-Yu Wang; Anthea Mitchell; Robert Puccinelli; Doug Black; Susana Rojas; Salustiano Ribeiro; Valerie Tulier-Laiwa; Jacqueline Martinez; Joselin Payan; Susy Rojas; Diane Jones; Daniel Martinez; Robert Nakamura; Gabriel Chamie; Vivek Jain; Maya Petersen; Joe DeRisi; Diane Havlir
Journal:  Ann Intern Med       Date:  2022-03-15       Impact factor: 25.391

8.  Accuracy of a Novel SARS-CoV-2 Antigen-Detecting Rapid Diagnostic Test from Standardized Self-Collected Anterior Nasal Swabs.

Authors:  Bilgin Osmanodja; Klemens Budde; Daniel Zickler; Marcel G Naik; Jörg Hofmann; Maximilian Gertler; Claudia Hülso; Heike Rössig; Philipp Horn; Joachim Seybold; Stephanie Lunow; Melanie Bothmann; Astrid Barrera-Pesek; Manuel Mayrdorfer
Journal:  J Clin Med       Date:  2021-05-13       Impact factor: 4.241

9.  Time scale performance of rapid antigen testing for SARS-CoV-2: Evaluation of 10 rapid antigen assays.

Authors:  Zakarya Abusrewil; Inas M Alhudiri; Hamza H Kaal; Salah Eldin El Meshri; Fawzi O Ebrahim; Tarek Dalyoum; Abdussamee A Efrefer; Khaled Ibrahim; Mohammed B Elfghi; Suleiman Abusrewil; Adam Elzagheid
Journal:  J Med Virol       Date:  2021-07-28       Impact factor: 20.693

10.  A Real-World Comparison of SARS-CoV-2 Rapid Antigen vs. Polymerase Chain Reaction Testing in Florida.

Authors:  Lao-Tzu Allan-Blitz; Jeffrey D Klausner
Journal:  J Clin Microbiol       Date:  2021-08-04       Impact factor: 5.948

View more
  2 in total

1.  Combined Diagnosis of SARS-CoV-2: Rapid Antigen Detection as an Adjunct to Nucleic Acid Detection.

Authors:  Xuewen Li; Yiting Wang; Junqi Pan; Jiancheng Xu; Qi Zhou