Ilaria Baccani1, Fabio Morecchiato1, Chiara Chilleri2, Chiara Cervini2, Emanuele Gori3, Daniela Matarrese4, Andrea Bassetti5, Manuela Bonizzoli6, Jessica Mencarini7, Alberto Antonelli8, Gian Maria Rossolini9.
Abstract
Three assays for SARS-CoV-2 antigen detection in nasopharyngeal swabs (Lumipulse® G SARS-CoV-2 Ag [LPG], STANDARDTM F COVID-19 Ag FIA [STF] and AFIAS COVID-19 Ag [AFC] were evaluated. Compared to RT-PCR, LPG, AFC and STF showed a variable sensitivity (87.9%, 37.5%, and 35.7%, respectively) and an overall high specificity (> 95%).Entities:
Keywords: COVID-19; SARS-CoV-2; antigen; chemiluminescent assay; lateral flow immunofluorescence assay
Year: 2021 PMID: 34174523 PMCID: PMC8137375 DOI: 10.1016/j.diagmicrobio.2021.115434
Source DB: PubMed Journal: Diagn Microbiol Infect Dis ISSN: 0732-8893 Impact factor: 2.803
Performance of Lumipulse® G SARS-CoV-2 Ag, AFIAS COVID 19 Ag and STANDARDTM F COVID-19 Ag FIA on UTM swabs.
| Total sample tested | Sensitivity (%) | 95% CI-Sensitivity | Specificity (%) | 95% CI-Specificity | PPV (%) | 95% CI-PPV | NPV (%) | 95% CI-NPV | |
|---|---|---|---|---|---|---|---|---|---|
| Lumipulse® G SARS-CoV-2 Ag | 201 | 29/33 (87.9) | 73.7-95.8 | 161/168 (95.8) | 92-98.1 | 80.5 | 65.6-90.9 | 97 | 93.7-98.8 |
| AFIAS COVID 19 Ag | 81 | 9/24 (37.5) | 20.4-57.4 | 57/57 (100) | 95.7-100 | 100 | 76.2-100 | 79.2 | 68.8-87.3 |
| STANDARDTM F COVID-19 Ag FIA | 93 | 10/28 (35.7) | 20.1-54.2 | 65/65 (100) | 96.2-100 | 100 | 78.3-100 | 78.3 | 68.8-86.1 |
12 invalid samples were excluded from the analysis
Fig. 1Number of positive isolates detected by Lumipulse® G SARS-CoV-2 Ag, AFIAS COVID 19 Ag and STANDARDTM F COVID-19 Ag FIA correlated to the N gene Ct value by RT-PCR. Samples with Ct≥ 40 have been considered as negative.