| Literature DB >> 35470272 |
Abstract
Liquid chromatography-tandem mass spectrometry (LC-MS/MS) is increasingly utilized in clinical laboratories because it has advantages in terms of specificity and sensitivity over other analytical technologies. These advantages come with additional responsibilities and challenges given that many assays and platforms are not provided to laboratories as a single kit or device. The skills, staff, and assays used in LC-MS/MS are internally developed by the laboratory, with relatively few exceptions. Hence, a laboratory that deploys LC-MS/MS assays must be conscientious of the practices and procedures adopted to overcome the challenges associated with the technology. This review discusses the post-development landscape of LC-MS/MS assays, including validation, quality assurance, operations, and troubleshooting. The content knowledge of LC-MS/MS users is quite broad and deep and spans multiple scientific fields, including biology, clinical chemistry, chromatography, engineering, and MS. However, there are no formal academic programs or specific literature to train laboratory staff on the fundamentals of LC-MS/MS beyond the reports on method development. Therefore, depending on their experience level, some readers may be familiar with aspects of the laboratory practices described herein, while others may be not. This review endeavors to assemble aspects of LC-MS/MS operations in the clinical laboratory to provide a framework for the thoughtful development and execution of LC-MS/MS applications.Entities:
Keywords: Calibration; Liquid chromatography-tandem mass spectrometry; Operations; Quality control; Troubleshooting; Validation; Verification
Mesh:
Year: 2022 PMID: 35470272 PMCID: PMC9057814 DOI: 10.3343/alm.2022.42.5.531
Source DB: PubMed Journal: Ann Lab Med ISSN: 2234-3806 Impact factor: 4.941
Example features to be evaluated prior to sample extraction or LC-MS/MS analysis
| Component | Required information | |
|---|---|---|
| Assay reagent | Calibration expiry | Expiration date |
| QC expiry | Expiration date | |
| Critical reagent expiry | Expiration date | |
| Critical reagent sufficient volume | Number of samples and volume used per sample | |
| Sample | Sample volume | Sufficient volume and insufficient quantity SOP |
| Sample temperature | Outside of SOP requirements (e.g., sample received at room temperature instead of frozen) | |
| Sample interferences | Gross hemolysis, lipemia, icteria, etc. | |
| Sample stability | Date of draw and stability SOP | |
| Sample tube type | Colored cap or subaliquot identifier for type | |
| Sample pertinent information | History of sample, if provided. Monitoring of certain patients may indicate gross elevations in the compound(s). Predilution may prevent the need to address carryover or subsequent re-assay on dilution | |
| Sample previous analysis | Freeze–thaw cycles and pertinent prior results (e.g., dilution required) | |
| Sample abnormal observations | Bacterial contamination, inappropriate color, etc. | |
| System | Mobile phase(s) expiry | Expiration date |
| Mobile phase volume | Sufficient volume | |
| Mobile phase abnormal observations | Odd material in bottle (dust, bacteria), filter stones above liquid level | |
| Column check | Column manufacturer, dimensions, and stationary phase | |
| MS gas pressures/volume | Pressure within range and sufficient volume of gas available | |
| MS base pressure | System pressure within range | |
| Autosampler wash expiry | Expiration date | |
| Autosampler wash volume | Sufficient volume | |
| System suitability review | See |
Abbreviations: LC-MS/MS, liquid chromatography-tandem mass spectrometry; QC, quality control; SOP, standard operating procedure.
Description of available data components derived from triplicate analysis of SST injections
| Data component | Acute determination | Longitudinal determination | Possible trouble shooting target(s) | |
|---|---|---|---|---|
| IS | Peak width above baseline | Gross deviation | Increasing trend | Column, LC pumps, tubing, mobile phases |
| Peak width at 50% height | Gross deviation | Increasing trend | Column, LC pumps, tubing, mobile phases | |
| Peak height | Gross deviation | Increasing trend | Autosampler, MS source, MS optics | |
| Peak area reproducibility | Poor precision | Increasing trend | Autosampler, MS source | |
| Retention time | Gross deviation | Increasing or decreasing trend | Column, LC pumps, tubing, SST solvent | |
| Analyte | Peak width above baseline | Gross deviation | Increasing trend | Column, LC pumps, tubing, mobile phases |
| Peak width at 50% height | Gross deviation | Increasing trend | Column, LC pumps, tubing, mobile phases | |
| Peak height | Gross deviation | Increasing trend | Autosampler, MS, mobile phases | |
| Retention time | Gross deviation | Increasing or decreasing trend | LC pumps, tubing, SST solvent | |
| Peak area | Confidence in appropriate data analysis for batch (enough response) | MS source, MS optics, mobile phases, autosampler | ||
| LC system | Pressure trace | Min/max outside of normal operating range | Initial (equilibrated conditions) increasing or decreasing trend | Column, LC, tubing |
| Other | Known interferences with resolution calculated | Minimum resolution achieved | Change in resolution over time | Column, mobile phases, LC |
Abbreviations: LC, liquid chromatography; MS, mass spectrometry; SST, system suitability test; IS, internal standard.
Example details to be reviewed or captured during the test phase of LC-MS/MS assays
| Component | Required information | |
|---|---|---|
| Batch | Sample list checked against samples to be assayed | Sample identity |
| Batch order confirmed | Batch layout | |
| Sample preparation | Sample preparation materials available | Method SOP |
| Test samples at equilibrium and homogeneous | Time and mixing check | |
| Calibrator and quality control lots confirmed | Lot numbers and record | |
| Pertinent times recorded (time-defined steps in the SOP) | Start and stop time for appropriate components | |
| Pipette(s) utilized | Pipette serial number and volume | |
| Reagent lots | Lots recorded, documented confirmed purity/activity for critical components | |
| Steps documented | Preparation checklist | |
| Abnormal results/preparative deviations | Note to preparation checklist | |
| Analysis | System equilibrated | Pre-run details |
| Pre-run injections | Method SOP | |
| Plate/vial location | Preparation checklist |
Abbreviations: LC-MS/MS, liquid chromatography-tandem mass spectrometry; SOP, standard operating procedure.
Examples of data metrics and method of review for LC-MS/MS assays
| Post-analytical data component | Check |
|---|---|
| IS recovery–sample | Plot or percentages compared to knowns (calls, QCs) |
| IS recovery–batch | Plot or linear regression |
| Retention times | Plot or linear regression |
| Transition ratio | Compared to expected ratio from knowns (calibrators/QCs) |
| Calibration curve–fit | High calibrator with lower % accuracy in a linear fit indicates quadratic best-fit. May be due to extraction, source saturation, or detector blinding |
| Calibration curve–outliers | Gross outliers (accuracy outside of 85–115/80%–120%) |
| Calibration curve–regression equation between batches | Large changes in slope or intercept may indicate need for troubleshooting |
| QCs–acute accuracy | QCs within 15% of expected concentration or tighter criteria based on assay requirements |
| QCs–precision (Westgard) | QCs reviewed based on Westgard rules |
| Carryover–high standard to blank | Establishes expectation of carryover within the batch |
| Critical values | Concentrations with a defined threshold are immediately reported to the physician |
| Blank contamination | Contribution of IS to analyte |
| Double blank contamination | Contribution of procedure to IS response |
| Carryover–samples | Samples greater than carryover limit reviewed/reinjected |
*QC review by Westgard rules for multianalyte panels may be difficult to apply. See section “Operation of LC-MS/MS assays.”
Abbreviations: LC-MS/MS, liquid chromatography-tandem mass spectrometry; IS, internal standard; QC, quality control.
Examples of reasons for sample re-analysis
| Observation | Re-analysis of patient samples | ||
|---|---|---|---|
|
| |||
| Incidence (sample or batch) | Action | Notes | |
| Sample | Re-extraction on dilution | See “Dilution” section for re-injection options | |
| Carryover | Sample | Blanks prior to re-analysis to confirm system cleanliness | Beware of well-to-well contamination with certain auto- samplers or exceptionally high concentration samples |
| Carryover | Batch | Equipment maintenance | Root cause may be associated with extraction equipment |
| QC failure | Batch | Root-cause analysis to determine origin | Use of ISs generally precludes meaningful QC value change on re-injection |
| Chromatographic degradation/failure | Batch | Root-cause analysis to determine origin | Ensure stability and volume of extracts prior to re-injection |
| Low response | Batch | Root-cause analysis to determine origin | Any re-injections after repairs must be within the post- extraction stability limits |
| Failed calibration curve | Batch | If instrument response is not the root cause, re-preparation of batch is most likely outcome | Multiple calibration points must not be rejected to achieve acceptable curve or QC accuracy |
| Persistent interference | Batch | Re-injection on reflex method | Evaluate frequency to determine need for exclusively utilizing the reflex method |
| Transition ratio failure | Sample | Re-extraction on dilution or reflex method | |
| IS recovery | Sample | Re-extraction on dilution or reflex method | Very high analyte concentrations may inhibit ionization of the IS |
| IS recovery | Batch | Root-cause determines re-extraction (error in sample preparation) or re-injection (error in instrumentation) | Any re-injections after repairs must be within the post- extraction stability limits |
| Retention time shift | Sample | If re-injection is insufficient, dilution on re- extraction or reflex method | Pressure trace associated with sample may indicate acute liquid-flow issue |
| Retention time shift | Batch | LC troubleshooting to determine root-cause | |
Abbreviations: LC, liquid chromatography; QC, quality control; IS, internal standard.
Example analyte-to-IS ratio of a sample diluted by under-aliquoting of the sample as opposed to volumetric dilution of the sample by addition of solvent
| Sample volume (µL) | Analyte concentration, sample (nM) | IS concentration (nM) | IS volume (µL) | Final volume (µL) | Analyte concentration, final mixture (nM) | IS concentration, final mixture (nM) | Analyte-to-IS ratio |
|---|---|---|---|---|---|---|---|
| 100 | 5,000 | 100 | 500 | 600 | 833.33 | 83.33 | 10 |
| 10 | 5,000 | 100 | 500 | 510 | 98.04 | 98.04 | 1 |
Note that the analyte-to-IS ratio is preserved despite reducing the volume of the sample aliquot.
Abbreviation: IS, internal standard.
Examples of LC-MS/MS assay components that require verification prior to assay execution with suggested verification materials and data components for possible review
| Reagent(s) being changed/replaced | Material(s) used for verification | Data for acceptance |
|---|---|---|
| LC column | SST | Retention time, response, relative resolution (if known, isobars are included in SST solution) |
| Mobile phase replacement | SST | Retention time, response, relative resolution (if known, isobars are included in SST solution), response of analyte |
| LC hardware change (e.g., replacement tubing or pump seal) | SST | Retention time, response, relative resolution (if known, isobars are included in SST solution) |
| Autosampler wash solution | High calibrator followed by blank | Response of analyte in blank following high calibrator (or other carryover check) |
| Autosampler hardware change (e.g., new needle or rotor seal) | High calibrator followed by blank and SST | Response of analyte in blank following high calibrator (or other carryover check) Retention time and response of SST |
| LC preventive maintenance | SST | Retention time, response, relative resolution (if known, isobars are included in SST solution) |
| MS component replacement (e.g., electrode) | SST | Response |
| MS preventive maintenance | SST | Response |
| New IS solution | Extracted blank with IS | Response of analyte in extracted blank Expected response of IS (more important for SPE/LLE analyses) |
| New calibrators | Calibrators, QCs, and patient samples. Previous PT’s or certified reference materials if available and within stability range | Agreement between old and new calibrators. |
| New QCs | Replicates of QCs | Accuracy/trueness of results (if necessary) |
| CV (%) of QCs and established range | ||
| New assay material (e.g., new vendor of plates or pipette tips) | Calibrators, QCs, blanks, and patient samples | Response, accuracy, imprecision, interfering signals |
| Critical assay reagents (e.g., hydrolysis enzyme or precipitation solvent) | Calibrators, QCs, blanks, and patient samples | Response, accuracy, imprecision, interfering signals |
| Critical assay material (e.g., new lot of SPE media) | Calibrators, QCs, blanks, and patient samples | Response, accuracy, imprecision, interfering signals |
Abbreviations: LC, liquid chromatography; SST, system suitability test; MS, mass spectrometer; PT, proficiency test; SPE, solid phase extraction; LLE, liquid-liquid extraction.
Data recorded from five injections of PCP and its IS, PCP-D5
| Precursor ( | Product ( | Mean peak area | CV (%) of peak area | % Difference from accurate mass area |
|
| ||||
| 244.0 | 91 | 6,228,332 | 1.2 | –9 |
| 244.2 | 91 | 6,863,210 | 1.3 | NA |
| 244.4 | 91 | 6,165,287 | 1.0 | –10 |
|
| ||||
| Precursor ( | Product ( | Mean peak area ratio | CV (%) of peak area ratio | % Difference from accurate mass area ratio |
|
| ||||
| 244.0 | 91 | 62 | 1.4 | –9.0 |
| 244.2 | 91 | 68 | 1.5 | NA |
| 244.4 | 91 | 61 | 1.0 | –10.0 |
The peak areas (top) demonstrate a reduction in signal due to incorrect m/z acquisition. Normalization by the IS (bottom) is indicative of the quantitative effect.
Abbreviations: PCP, phencyclidine; IS, internal standard; NA, not applicable.
Fig. 1(A) Assay of 11-nor-9-carboxy-Δ9-tetrahydrocannabinol using the original method, showing the analyte and IS with a retention time of 1.379 min and a near-co-eluting isobaric interferent right before the peak of interest (retention time of approximately 1.27 min) observed in all three analyte transitions. (B) The same patient sample assayed using the modified chromatographic separation (see Table 9) from the original extract (re-injection only).
Abbreviation: IS, internal standard.
Chromatographic program for the original validated gradient for 11-nor-9-carboxy-Δ9-tetrahydrocannabinol and the modified gradient to provide resolution of the delta-8 isomer generated as a reflex test for samples with inadequate resolution
| Originally validated gradient | Modified reflex gradient | |||||||
|---|---|---|---|---|---|---|---|---|
|
|
| |||||||
| Time (min) | Flow rate (mL/min) | %A | %B | Time (min) | Flow rate (mL/min) | %A | %B | |
| 0 | 0.5 | 100 | 0 | 0 | 0.55 | 100 | 0 | |
| 0.05 | 0.5 | 41 | 59 | 0.05 | 0.55 | 48 | 52 | |
| 1 | 0.5 | 41 | 59 | 1.45 | 0.55 | 48 | 52 | |
| 1.9 | 0.5 | 38 | 62 | 3 | 0.55 | 45 | 55 | |
| 1.95 | 0.5 | 0 | 100 | 3.15 | 0.55 | 0 | 100 | |
| 2.15 | 0.7 | 0 | 100 | 3.3 | 0.7 | 0 | 100 | |
| 2.23 | 0.7 | 100 | 0 | 3.5 | 0.7 | 0 | 100 | |
| 2.3 | 0.7 | 100 | 0 | 3.55 | 0.7 | 100 | 0 | |
| 3.65 | 0.7 | 100 | 0 | |||||
Fig. 2Visualization of the gradient program for the original validated method (solid line) and the validated reflex method to afford isobaric separation in samples with failed chromatographic/transition ratio analysis in the analysis of 11-nor-9-carboxy-Δ9-tetrahydrocannabinol.
Abbreviations: LC, liquid chromatography.