Literature DB >> 15766733

Assessing analytical specificity in quantitative analysis using tandem mass spectrometry.

Mark M Kushnir1, Alan L Rockwood, Gordon J Nelson, Bingfang Yue, Francis M Urry.   

Abstract

OBJECTIVES: The necessity of confirmation of compound identity in quantitative analysis is well recognized for methods utilizing single mass spectrometry detection but is not commonly addressed for applications utilizing multiple-stage mass spectrometry (MSn). For MSn detection, no commonly accepted rules for assessment of analytical specificity in quantitative analyses have been established to date.
METHODS: To assure compound identity, we evaluated approaches based on monitoring multiple mass transitions of a target compound followed by comparison of the branching ratios of the mass transitions.
RESULTS: Monitoring multiple mass transitions along with evaluation of the ratio of their relative intensities allows the analyst to distinguish the target analyte from interferences in quantitative analysis. The strategy and the acceptance criteria are compound and method specific and should be established during the method development and validation.
CONCLUSIONS: The certainty of analyte identity is very important in quantitative analysis using MSn detection; methods to verify analyte identity should be used in all critical applications.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15766733     DOI: 10.1016/j.clinbiochem.2004.12.003

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  33 in total

1.  A sensitive and specific LC-MS/MS method for rapid diagnosis of Niemann-Pick C1 disease from human plasma.

Authors:  Xuntian Jiang; Rohini Sidhu; Forbes D Porter; Nicole M Yanjanin; Anneliese O Speak; Danielle Taylor te Vruchte; Frances M Platt; Hideji Fujiwara; David E Scherrer; Jessie Zhang; Dennis J Dietzen; Jean E Schaffer; Daniel S Ory
Journal:  J Lipid Res       Date:  2011-04-24       Impact factor: 5.922

2.  Establishing ion ratio thresholds based on absolute peak area for absolute protein quantification using protein cleavage isotope dilution mass spectrometry.

Authors:  Philip L Loziuk; Ronald R Sederoff; Vincent L Chiang; David C Muddiman
Journal:  Analyst       Date:  2014-11-07       Impact factor: 4.616

3.  Highly Reproducible Automated Proteomics Sample Preparation Workflow for Quantitative Mass Spectrometry.

Authors:  Qin Fu; Michael P Kowalski; Mitra Mastali; Sarah J Parker; Kimia Sobhani; Irene van den Broek; Christie L Hunter; Jennifer E Van Eyk
Journal:  J Proteome Res       Date:  2017-11-10       Impact factor: 4.466

4.  LC-MS/MS in the Clinical Laboratory - Where to From Here?

Authors:  Stefan Kg Grebe; Ravinder J Singh
Journal:  Clin Biochem Rev       Date:  2011-02

5.  Automated detection of inaccurate and imprecise transitions in peptide quantification by multiple reaction monitoring mass spectrometry.

Authors:  Susan E Abbatiello; D R Mani; Hasmik Keshishian; Steven A Carr
Journal:  Clin Chem       Date:  2009-12-18       Impact factor: 8.327

6.  Functional redundancy of the two 5-hydroxylases in monolignol biosynthesis of Populus trichocarpa: LC-MS/MS based protein quantification and metabolic flux analysis.

Authors:  Jack P Wang; Christopher M Shuford; Quanzi Li; Jina Song; Ying-Chung Lin; Ying-Hsuan Sun; Hsi-Chuan Chen; Cranos M Williams; David C Muddiman; Ronald R Sederoff; Vincent L Chiang
Journal:  Planta       Date:  2012-05-25       Impact factor: 4.116

7.  Newborn screening for cerebrotendinous xanthomatosis is the solution for early identification and treatment.

Authors:  Andrea E DeBarber; Limor Kalfon; Ayalla Fedida; Vered Fleisher Sheffer; Shani Ben Haroush; Natalia Chasnyk; Efrat Shuster Biton; Hanna Mandel; Krystal Jeffries; Eric S Shinwell; Tzipora C Falik-Zaccai
Journal:  J Lipid Res       Date:  2018-08-22       Impact factor: 5.922

8.  From lost in translation to paradise found: enabling protein biomarker method transfer by mass spectrometry.

Authors:  Russell P Grant; Andrew N Hoofnagle
Journal:  Clin Chem       Date:  2014-05-08       Impact factor: 8.327

9.  Time-of-flight mass spectrometry analysis of the ferroportin-hepcidin binding domain complex for accurate mass confirmation of bioactive hepcidin 25.

Authors:  David K Crockett; Mark M Kushnir; John D Phillips; Alan L Rockwood
Journal:  Clin Chim Acta       Date:  2009-12-03       Impact factor: 3.786

10.  Validation of a liquid chromatography-tandem mass spectrometry method to enable quantification of 3-iodothyronamine from serum.

Authors:  Andrea E DeBarber; Travis Geraci; Vincent P Colasurdo; Sarah A Hackenmueller; Thomas S Scanlan
Journal:  J Chromatogr A       Date:  2008-09-12       Impact factor: 4.759