Thomas D Mann1,2, Kundan S Thind1,3,4, Nicolas P Ploquin1,2,3.
Abstract
BACKGROUND ANDEntities:
Keywords: Automation; Heuristics; Organs at risk; Particle accelerators; Radiosurgery; Retrospective studies
Year: 2022 PMID: 35243038 PMCID: PMC8885579 DOI: 10.1016/j.phro.2022.02.009
Source DB: PubMed Journal: Phys Imaging Radiat Oncol ISSN: 2405-6316
Fig. 1Visual example of the “Island Blocking” Problem. This occurs when multiple PTV structures align with the MLC leaves resulting in a gap of MLC coverage in the center.
Patient Plan Characteristics.
| Metric | Average | Median | Min | Max |
|---|---|---|---|---|
| Prescription Dose (Gy) | – | 21 | 18 | 30* |
| Number of PTVs | – | 4 | 2 | 13 |
| PTV Volume [cm3] | 1.58 | 0.38 | 0.05 | 13.09 |
| Total PTV Volume [cm3] | 7.06 | 5.94 | 0.38 | 25.57 |
| Off-axis Distance [cm] | 3.50 | 3.67 | 0 | 7.66 |
*max prescription dose was delivered in five fractions.
Generic optimization objectives template used for VMAT photon optimization.
| Structure | Objective Type | Volume [%] | Dose [cGy] | Weighting |
|---|---|---|---|---|
| PTV | Upper | 0 | 1.4 × PD* | 150 |
| PTV | Lower | 100 | PD* | 150 |
| GTV | Lower | 100 | 1.1 × PD* | 100 |
| Ring | Upper | 0 | 0.5 × PD* | 150 |
| OAR | Upper | 0 | 150 | 50 |
| NTO | Upper | Auto | Auto | 150 |
*PD is prescription dose to the 80% isodose line.
Fig. 2SOAR planning user interface a) Patient, course, and plan selection b) PTV Target and OAR selection c) Optimization results data grid, results for each isocenter are loaded separately d) Isocenter number selection and optimization start button e) Load results button, isocenter selection, and number of beams selection f) Build plan button with option to perform VMAT optimization.
Fig. 3Boxplots of OAR dose metrics comparing manual planning (MNL) in red to SOAR planning (SOAR) in grey. The central line shows the median, edges of the box denote the 25th and 75th percentiles, and whiskers are the min and max points not including outliers. Outliers shown as black circles. a) Normal brain volume receiving 10 and 12 Gy. b) Max doses to the eyes and lenses. c) Max doses to the optic apparatus. d) Max dose to the brainstem and mean dose to the cochlea.
Dose metric comparison results for clinical and SOAR plans.
| Structure | Dose Metric | N | Manual Plans | SOAR Plans | ||||
|---|---|---|---|---|---|---|---|---|
| Median | Min | Max | Median | Min | Max | |||
| Healthy Brain | V12Gy[cm3] | 25 | 13.79 | 1.62 | 128.62 | 14.17 | 1.57 | 116.69 |
| Healthy Brain | V10Gy[cm3] | 25 | 22.94 | 2.54 | 208.26 | 23.00 | 2.50 | 184.30 |
| Eye_L | D0.1cc[Gy] | 25 | 1.06 | 0.19 | 3.12 | 0.87 | 0.19 | 3.81 |
| Eye_R | D0.1cc[Gy] | 25 | 1.11 | 0.30 | 4.94 | 1.13 | 0.09 | 3.38 |
| Lens_L | D0.1cc[Gy] | 24 | 0.43 | 0.00 | 1.50 | 0.40 | 0.00 | 1.51 |
| Lens_R | D0.1cc[Gy] | 24 | 0.38 | 0.00 | 1.93 | 0.39 | 0.00 | 1.45 |
| Optic Nerve L | D0.035cc[Gy] | 25 | 1.18 | 0.08 | 4.46 | 0.89 | 0.23 | 3.86 |
| Optic Nerve R | D0.035cc[Gy] | 25 | 1.64 | 0.08 | 5.89 | 1.25 | 0.14 | 3.73 |
| Cochlea_L | Mean[Gy] | 20 | 0.97 | 0.07 | 2.30 | 0.94 | 0.03 | 3.02 |
| Cochlea_R | Mean[Gy] | 20 | 1.28 | 0.20 | 2.56 | 0.83 | 0.13 | 2.41 |
| Brainstem | D0.03cc[Gy] | 25 | 2.69 | 1.43 | 10.72 | 2.84 | 1.41 | 11.56 |
| Chiasm | D0.035cc[Gy] | 25 | 1.52 | 0.10 | 4.66 | 1.51 | 0.40 | 4.28 |
| PTV | D2% [%] | 112 | 138.5 | 117.0 | 159.1 | 113.9 | 156.4 | |
| PTV | CI | 98 | 0.71 | 0.38 | 0.90 | 0.39 | 0.90 | |
| PTV | GI | 74 | 1.85 | 1.41 | 3.13 | 1.82 | 1.39 | 3.45 |
*Statistically significant results are highlighted in bold.