Comparison of a new Scheimpflug imaging combined with partial coherence interferometry biometer and a low-coherence reflectometry biometer.

Ramón Ruiz-Mesa1, Antonio Abengózar-Vela2, María Ruiz-Santos2.   

Abstract

PURPOSE: To evaluate the repeatability of a new biometer using Scheimpflug technology combined with partial coherence interferometry (PCI) (Pentacam AXL) and its agreement with a device based on optical low-coherence reflectometry (OLCR), the Allegro Biograph.
SETTING: Oftalvist Centro Integral Ocular Jerez, Jerez de la Frontera, Spain.
DESIGN: Evaluation of a diagnostic test.
METHODS: The mean keratometry (K), central corneal thickness (CCT), anterior chamber depth (ACD), and axial length (AL) were measured with the 2 devices 3 times by the same examiner in 2 groups (patients with cataract and patients without cataract). The repeatability was determined using the within-subject standard deviation, test-retest repeatability, coefficient of variation, and intraclass correlation coefficient. The correlation was evaluated with the Pearson coefficient and interchangeability with the Bland-Altman plot.
RESULTS: Eighty eyes (40 eyes in each group) of 80 patients were analyzed. Significant differences were found between the Scheimpflug-PCI device and the OLCR device for mean K in the normal group (P < .001) and for CCT in the normal group (P < .05) and the cataract group (P < .001). There were no differences between devices in ACD and AL in either group. The repeatability between devices was similar. Although a significant correlation between devices was found for all measurements (all P < .001), wide limits of agreement were found in both groups for all biometric parameters.
CONCLUSIONS: The Scheimpflug-PCI and OLCR devices showed excellent intravisit repeatability and high correlation for mean K, CCT, ACD, and AL in healthy and cataractous eyes. No differences were found in AL; however, the 2 devices might not be interchangeable.
Copyright © 2017 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2017        PMID: 29223229     DOI: 10.1016/j.jcrs.2017.08.016

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  6 in total

1.  Comparison of Anterior Segment Parameters and Axial Length Using Two Scheimpflug Devices with Integrated Optical Biometers.

Authors:  Chantaka Supiyaphun; Sasivimol Rattanasiri; Passara Jongkhajornpong
Journal:  Clin Ophthalmol       Date:  2020-10-23

2.  The repeatability and agreement of biometric measurements by dual Scheimpflug device with integrated optical biometer.

Authors:  Hassan Hashemi; Sara Sardari; Abbasali Yekta; Mehdi Khabazkhoob
Journal:  Sci Rep       Date:  2022-05-11       Impact factor: 4.996

3.  Repeatability of total Keratometry and standard Keratometry by the IOLMaster 700 and comparison to total corneal astigmatism by Scheimpflug imaging.

Authors:  Giacomo Savini; Leonardo Taroni; Domenico Schiano-Lomoriello; Kenneth J Hoffer
Journal:  Eye (Lond)       Date:  2020-11-02       Impact factor: 4.456

4.  Evaluation and comparison of a novel Scheimpflug-based optical biometer with standard partial coherence interferometry for biometry and intraocular lens power calculation.

Authors:  Ziyang Wang; Wenli Yang; Dongjun Li; Wei Chen; Qi Zhao; Yifeng Li; Rui Cui; Lin Shen; Junfang Xian
Journal:  Exp Ther Med       Date:  2021-02-05       Impact factor: 2.447

5.  Comparison of a New Optical Biometer That Combines Scheimpflug Imaging With Partial Coherence Interferometry With That of an Optical Biometer Based on Swept-Source Optical Coherence Tomography and Placido-Disk Topography.

Authors:  Shihao Chen; Qiaoyue Zhang; Giacomo Savini; Shuangzhe Zhang; Xiaomin Huang; Jinjin Yu; Yirang Wang; Rui Ning; Jinhai Huang; Ruixue Tu
Journal:  Front Med (Lausanne)       Date:  2022-02-10

6.  Development and Verification of an Adjustment Factor for Determining the Axial Length Using Optical Biometry in Silicone Oil-Filled Eyes.

Authors:  Gerd U Auffarth; Tadas Naujokaitis; Louise Blöck; Anna Daghbashyan; Jan Meis; Victor A Augustin; Ramin Khoramnia; Timur M Yildirim
Journal:  Diagnostics (Basel)       Date:  2022-01-11
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.