Kyungsoo Bae1,2, Kyung Nyeo Jeon1,2, Moon Jung Hwang3, Yunsub Jung3, Joonsung Lee3.
Abstract
(1) Background: Highly flexible adaptive image receive (AIR) coil has become available for clinical use. The present study aimed to evaluate the performance of AIR anterior array coil in lung MR imaging using a zero echo time (ZTE) sequence compared with conventional anterior array (CAA) coil. (2)Entities:
Keywords: adaptive image receive coil; lung; magnetic resonance imaging; ultrashort T2; zero-echo-time imaging
Year: 2022 PMID: 35054316 PMCID: PMC8774338 DOI: 10.3390/diagnostics12010148
Source DB: PubMed Journal: Diagnostics (Basel) ISSN: 2075-4418
Qualitative scoring system of ZTE lung MR images.
|
|
| 1, indistinguishable segmental vessels |
| 2, blurred visualization of segmental vessels |
| 3, clear visualization of segmental vessels |
| 4, visualization of subsegmental vessels |
| 5, visualization of sub-subsegmental vessels |
|
|
| 1, indistinguishable lobar bronchus |
| 2, visualization of lobar bronchus |
| 3, visualization of segmental bronchus |
| 4, visualization of subsegmental bronchus |
| 5, visualization of sub-subsegmental bronchus |
|
|
| 1, no visualization |
| 2, partial visualization of one fissure |
| 3, partial visualization of two fissures |
| 4, visualization of near whole course of one fissure |
| 5, visualization of near whole course of two fissures |
|
|
| 1, unacceptable noise/artifacts |
| 2, above-average noise/artifacts |
| 3, average and acceptable noise/artifacts |
| 4, less-than average noise/artifacts |
| 5, minimum or nothing |
|
|
| 1, unacceptable |
| 2, suboptimal |
| 3, satisfactory |
| 4, good |
| 5, excellent |
Comparison of signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and image sharpness between ZTE-CAA and ZTE-AIR.
| ZTE-CAA | ZTE-AIR | Effect Size | ||
|---|---|---|---|---|
|
| 7.47 ± 2.81 | 11.79 ± 5.75 | <0.001 | 0.368 |
|
| 1.17 ± 1.25 | 2.10 ± 1.84 | <0.001 | 0.147 |
|
| 0.49 ± 0.37 * | 0.44 ± 0.32 * | <0.001 | 0.378 |
* Expressed as blur metric values, with a lower value indicating a sharper image. A p-value < 0.05 was considered statistically significant. ZTE-CAA, zero echo time lung MR imaging using conventional anterior array coil; ZTE-AIR, zero echo time lung MR imaging using adaptive image receive coil.
Subjective image quality of ZTE-CAA and ZTE-AIR by two independent readers (n = 66).
| Reader 1 | Reader 2 | |||||||
|---|---|---|---|---|---|---|---|---|
| ZTE-CAA | ZTE-AIR | Effect Size | ZTE-CAA | ZTE-AIR | Effect Size | |||
|
| 4.74 ± 0.56 | 4.89 ± 0.31 | 0.003 | 0.063 | 4.73 ± 0.54 | 4.83 ± 0.41 | 0.007 | 0.053 |
|
| 3.59 ± 0.68 | 3.98 ± 0.64 | <0.001 | 0.183 | 3.52 ± 0.64 | 4.82 ± 0.70 | <0.001 | 0.156 |
|
| 2.45 ± 1.13 | 2.79 ± 1.00 | <0.001 | 0.130 | 2.53 ± 1.03 | 2.77 ± 0.87 | <0.001 | 0.137 |
|
| 3.36 ± 1.11 | 3.89 ± 0.86 | <0.001 | 0.153 | 3.14 ± 1.12 | 3.50 ± 1.04 | <0.001 | 0.175 |
|
| 3.5 ± 1.04 | 3.89 ± 0.88 | <0.001 | 0.128 | 3.29 ± 1.12 | 3.83 ± 0.92 | <0.001 | 0.125 |
Data are presented as mean and standard deviation. A p-value < 0.05 was considered statistically significant. ZTE-CAA, zero echo time lung MR imaging using conventional anterior array coil; ZTE-AIR, zero echo time lung MR imaging using adaptive image receive coil.
Comparison of diagnostic performance for nodules and emphysema on a zone-by-zone basis between ZTE-CAA and ZTE-AIR.
| Reader 1 | Reader 2 | |||||
|---|---|---|---|---|---|---|
| ZTE-CAA | ZTE-AIR | ZTE-CAA | ZTE-AIR | |||
| Nodules | ||||||
|
| 87.0 (47/54) | 88.9 (48/54) | 1.00 | 85.2 (46/54) | 88.9 (48/54) | 0.625 |
|
| 98.1 (206/210) | 98.6 (207/210) | 1.00 | 98.1 (206/210) | 98.6 (207/210) | 1.00 |
|
| 95.8 (253/264) | 96.6 (255/264) | 0.625 | 95.5 (252/264) | 96.6 (255/264) | 0.375 |
| Emphysema/cysts | ||||||
|
| 78.3 (36/46) | 91.3 (42/46) | 0.031 | 73.9 (34/46) | 89.1 (41/46) | 0.039 |
|
| 98.6 (215/218) | 98.6 (215/218) | 1.00 | 98.6 (215/218) | 99.5 (217/218) | 0.500 |
|
| 95.1 (251/264) | 97.3 (257/264) | 0.031 | 94.3 (249/264) | 97.7 (258/264) | 0.012 |
Numbers are shown in percentages. Raw data are shown in parentheses. A p-value < 0.05 was considered statistically significant. ZTE-CAA, zero echo time lung MR imaging using conventional anterior array coil; ZTE-AIR, zero echo time lung MR imaging using adaptive image receive coil.
Comparison of sensitivity on nodule-wise evaluation between ZTE-CAA and ZTE-AIR.
| Reader 1 | Reader 2 | |||||
|---|---|---|---|---|---|---|
| ZTE-CAA | ZTE-AIR | ZTE-CAA | ZTE-AIR | |||
|
| 80.0 (64/80) | 88.8 (71/80) | 0.065 | 80.0 (64/80) | 87.5 (70/80) | 0.070 |
|
| 93.1 (27/29) | 93.1 (27/29) | 1.000 | 96.8 (28/29) | 93.1 (27/29) | 1.000 |
|
| 72.5 (37/51) | 86.3 (44/51) | 0.039 | 70.6 (36/51) | 84.3 (43/51) | 0.016 |
Numbers are shown in percentages. Raw data are shown in parentheses. A p-value < 0.05 was considered statistically significant. ZTE-CAA, zero echo time lung MR imaging using conventional anterior array coil; ZTE-AIR, zero echo time lung MR imaging using adaptive image receive coil.
Figure 1A 61-year-old male with lung cancer in the right upper lobe. A 24 mm-sized primary nodule and a tiny satellite nodule in the right upper lobe in ZTE-CAA (a) and ZTE-AIR (b) are shown. Chest CT image (c) is presented as reference. ZTE-CAA, zero echo time lung MR image using conventional anterior array coil; ZTE-AIR, zero echo time lung MR image using adaptive image receive coil.
Figure 2A 55-year-old male with lung cyst in the right upper lobe. Note signal difference between lung parenchyma and cyst (short arrows) and visualization of peripheral bronchi (thin arrows) in ZTE-CAA (a) and ZTE-AIR (b). Chest CT image (c) is presented as reference. ZTE-CAA, zero echo time lung MR image using conventional anterior array coil; ZTE-AIR, zero echo time lung MR image using adaptive image receive coil.
Figure 3An 83-year-old male with lung cancer. A lobulated mass in the left upper lobe is well-defined in both ZTE-CAA (a) and ZTE-AIR (b). However, the presence of emphysema is more clearly defined in ZTE-AIR. CT image is presented for comparison (c). ZTE-CAA, zero echo time lung MR image using conventional anterior array coil; ZTE-AIR, zero echo time lung MR image using adaptive image receive coil.