| Literature DB >> 34795762 |
Shiva Shojaeian1, Niloofar Mortezapour1, Fatemeh Soltaninejad1, Nazanin Zargar1, Babak Zandi1, Yazdan Shantiaee1, Amin Bidaki2.
Abstract
AIMS: A glide path is created prior to root canal instrumentation by nickel-titanium (NiTi) rotary files to increase the efficiency and safety of cleaning and shaping. This study aimed to assess root canal transportation in use of different glide path files in curved canals.Entities:
Year: 2021 PMID: 34795762 PMCID: PMC8594984 DOI: 10.1155/2021/4898684
Source DB: PubMed Journal: Int J Dent ISSN: 1687-8728
Properties of the files.
| Name | Size | Speed (rpm) | Torque (N/cm) | Alloy material |
|---|---|---|---|---|
| Neolix | #15, 3% | 250 | 1.5 | Treated NiTi |
| EdgeGlidePath | #16, progressive | 350 | 2 | Heat-treated NiTi |
| One-G | #14, 3% | 350 | 1.2 | Conventional NiTi |
Figure 1The pre- and postoperative images at 5 mm from the apical foramen: (a) One-G, (b) EdgeGlidePath, and (c) Neolix.
Mean canal transportation and centering ratio of One-G, EdgeGlidePath, and Neolix files at 1, 3, 5, and 7 mm from the apex.
| File | Parameter | 1 mm (mean ± SD) | 3 mm (mean ± SD) | 5 mm (mean ± SD) | 7 mm (mean ± SD) |
|---|---|---|---|---|---|
| One-G | Transportation | 0.014 ± 0.066a | 0.007 ± 0.047a | 0.029 ± 0.069a | 0.017 ± 0.044a |
| Centering ratio | 0.461 ± 0.287a | 0.614 ± 0.294a | 0.443 ± 0.212a | 0.673 ± 0.174a | |
|
| |||||
| EdgeGlidePath | Transportation | 0.002 ± 0.032a | 0.011 ± 0.06a | 0.07 ± 0.197a | 0.019 ± 0.055a |
| Centering ratio | 0.459 ± 0.352a | 0.493 ± 0.229a | 0.643 ± 0.305a | 0.639 ± 0.349a | |
|
| |||||
| Neolix | Transportation | 0.019 ± 0.052a | 0.004 ± 0.051a | 0.03 ± 0.086a | 0.03 ± 0.068a |
| Centering ratio | 0.537 ± 0.269a | 0.487 ± 0.283a | 0.381 ± 0.265a | 0.434 ± 0.281a | |
Different superscripted letters indicate a significant difference between the groups (P < 0.05).