Evaluation of total corneal power measurements with a new optical biometer.

Mehdi Shajari1, Ruven Sonntag, Michaela Ramsauer, Thomas Kreutzer, Efstathios Vounotrypidis, Thomas Kohnen, Siegfried Priglinger, Wolfgang J Mayer.   

Abstract

PURPOSE: To evaluate the repeatability and validity of total corneal power measurements (total keratometry [TK]) obtained with a recently introduced optical biometer (IOLMaster 700, Carl Zeiss Meditec AG) to a Scheimpflug device (Pentacam, Oculus).
SETTING: Department of Ophthalmology, Ludwig-Maximilians-University, Munich, Germany.
DESIGN: Prospective randomized controlled trial.
METHODS: The inclusion criteria were a corneal astigmatism of 3.00 diopters (D) or less, no previous ocular surgery, no known corneal irregularities, and no known dry eyes. All eyes were measured 3 times using the optical biometer and the Scheimpflug device. The results were statistically compared using Bland-Altman, within-subject SD, and astigmatism vector analysis.
RESULTS: Ninety-three eyes of 93 subjects were included. Overall, the repeatability of all eyes measured was 0.42 for standard K, 0.40 for TK, 0.45 for total corneal refractive power (TCRP), 0.43 for true net power (TNP), and 0.39 for simulated K. Bland-Altman analysis showed no significant difference between the optical biometer's TK compared with standard K and the Scheimpflug device's simulated K, TNP, and TCRP, the P value exceeding .05 in each case. A difference of >0.50 diopters between astigmatism measurements between TK and TCRP was found in 10 cases compared with standard K and TNP with 0 and 2 cases.
CONCLUSIONS: Repeatability of both devices was high, although measurements of meridian showed a great variability, suggesting that numerous measurements are needed to enhance accuracy. TCRP measurements between recently introduced TK and TCRP cannot be used interchangeably. Future studies are necessary to evaluate which measurement will result in a better outcome when respective measurements are used for toric intraocular lens calculations.

Entities:  

Mesh:

Year:  2020        PMID: 32358259     DOI: 10.1097/j.jcrs.0000000000000136

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  13 in total

1.  Comparison of IOL-Master 700 and IOL-Master 500 biometers in ocular biological parameters of adolescents.

Authors:  Qiang Shi; Guang-Yan Wang; Yu-Hong Cheng; Cheng Pei
Journal:  Int J Ophthalmol       Date:  2021-07-18       Impact factor: 1.779

2.  Repeatability of OCT-Based versus Scheimpflug- and Reflection-Based Keratometry in Patients with Hyperosmolar and Normal Tear Film.

Authors:  Bjørn Gjerdrum; Kjell Gunnar Gundersen; Per Olof Lundmark; Bente Monica Aakre
Journal:  Clin Ophthalmol       Date:  2020-11-18

3.  Application of total keratometry in ten intraocular lens power calculation formulas in highly myopic eyes.

Authors:  Ling Wei; Kaiwen Cheng; Wenwen He; Xiangjia Zhu; Yi Lu
Journal:  Eye Vis (Lond)       Date:  2022-06-09

4.  Predicted vs measured posterior corneal astigmatism for toric intraocular lens calculations.

Authors:  H John Shammas; Fei Yu; Maya C Shammas; Renu Jivrajka; Christine Hakimeh
Journal:  J Cataract Refract Surg       Date:  2021-09-21       Impact factor: 3.528

5.  Repeatability of total Keratometry and standard Keratometry by the IOLMaster 700 and comparison to total corneal astigmatism by Scheimpflug imaging.

Authors:  Giacomo Savini; Leonardo Taroni; Domenico Schiano-Lomoriello; Kenneth J Hoffer
Journal:  Eye (Lond)       Date:  2020-11-02       Impact factor: 4.456

6.  Refractive Precision of Ray Tracing IOL Calculations Based on OCT Data versus Traditional IOL Calculation Formulas Based on Reflectometry in Patients with a History of Laser Vision Correction for Myopia.

Authors:  Bjørn Gjerdrum; Kjell Gunnar Gundersen; Per Olof Lundmark; Bente Monica Aakre
Journal:  Clin Ophthalmol       Date:  2021-02-26

7.  Evaluation of preoperative corneal astigmatism using swept-source optical biometry in Chinese cataract surgery candidates with high myopia: a prospective, comparative observational study.

Authors:  Yehui Tan; Liangping Liu; Jianbing Li; Yingyan Qin; Ao Sun; Mingxing Wu
Journal:  Ann Transl Med       Date:  2021-04

8.  Acquisition Time for Swept-Source Optical Biometry Plus Corneal Power Measurement During Cataract Evaluation.

Authors:  Ramón Ruiz-Mesa; Maria Ruiz-Santos; Julia Blanch-Ruiz; Ana Jiménez-Nieto
Journal:  Clin Ophthalmol       Date:  2022-03-05

9.  Agreement in anterior segment measurements between swept-source and Scheimpflug-based optical biometries in keratoconic eyes: a pilot study.

Authors:  Evangelia Chalkiadaki; Panos S Gartaganis; Thomas Ntravalias; Ioannis Giannakis; Evangelos Manousakis; Efthymios Karmiris
Journal:  Ther Adv Ophthalmol       Date:  2022-03-31

10.  Accuracy of OCT-derived net corneal astigmatism measurement.

Authors:  Clara Llorens-Quintana; Elias Pavlatos; Omkar Thaware; Seema Gupta; Daniel Gradin; Denzil Romfh; Yan Li; David Huang
Journal:  J Cataract Refract Surg       Date:  2022-03-01       Impact factor: 3.528

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.