Determination of the effective dose delivered by image guided radiotherapy in head & neck and breast treatments.

Mireille Conrad1, Grégory Bolard2, Marie Nowak3, Berardino De Bari4, Wendy Jeanneret-Sozzi4, Jean Bourhis4, Jean-François Germond3, François Bochud3, Raphaël Moeckli5.   

Abstract

PURPOSE: Image guided radiotherapy (IGRT) improves patient positioning for treatment delivery at the cost of an additional dose. This work aimed to calculate the effective dose (as an indicator of dose) for head & neck (H&N) and breast IGRT treatments by implementing dose calculation models to determine the dose distributions.
METHODS: The kV dose-models were created for the IGRT systems of Elekta Synergy (XVI) and Varian Clinac (OBI) linear accelerators within Philips Pinnacle TPS. Profiles and depth dose curves were measured in water. The models were validated in a CIRS thorax phantom. The IGRT dose distributions for five H&N and five breast patients were calculated. The effective dose was determined from the dose distributions following ICRP 103 recommendations. Moreover, time-saving approximations were studied in order to propose an alternative way of segmenting the tissues for a clinical implementation of the method. RESULTS AND
CONCLUSION: The effective dose specifically associated with IGRT varied from 1 to 10mSv depending on the protocol. The kV dose-model allowed us to calculate the dose distributions from IGRT for different configurations and patients, and to determine effective dose for IGRT protocols. The clinical implementation of the method was found to reduce time and to introduce a small enough increase of uncertainty in the results to be clinically usable.
Copyright © 2018. Published by Elsevier GmbH.

Entities:  

Keywords:  Breast; Effective dose; Head and Neck; IGRT

Mesh:

Year:  2018        PMID: 29426589     DOI: 10.1016/j.zemedi.2018.01.001

Source DB:  PubMed          Journal:  Z Med Phys        ISSN: 0939-3889            Impact factor:   4.820


  2 in total

1.  Tables for effective dose assessment from diagnostic radiology (period 1946-1995) in epidemiologic studies.

Authors:  Merle Friederike Meiboom; Wolfgang Hoffmann; Kerstin Weitmann; Heiner von Boetticher
Journal:  PLoS One       Date:  2021-04-01       Impact factor: 3.240

2.  Evaluation of patient organ doses from kilovoltage cone-beam CT imaging in radiation therapy.

Authors:  Alper Özseven; Bahar Dirican
Journal:  Rep Pract Oncol Radiother       Date:  2021-04-14
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.