Dose measurements in a thorax phantom at 3DCRT breast radiation therapy.

Elsa Bifano Pimenta1, Luciana Batista Nogueira2, Tarcísio Passos Ribeiro de Campos1.   

Abstract

BACKGROUND: The anthropomorphic and anthropometric phantom developed by the research group NRI (Núcleo de Radiações Ionizantes) can reproduce the effects of the interactions of radiation occurring in the human body. The whole internal radiation transport phenomena can be depicted by film dosimeters in breast RT. Our goal was to provide a dosimetric comparison of a radiation therapy (RT) plan in a 4MV 3D-conformal RT (4MV-3DCR T) and experimental data measured in a breast phantom.
MATERIALS AND METHODS: The RT modality was two parallel opposing fields for the left breast with a prescribed dose of 2.0 Gy in 25 fractions. The therapy planning system (TPS) was performed on CA T3D software. The dose readings at points of interest (POI) pre-established in TPS were recorded. An anthropometric thorax-phantom with removal breast was used. EBT2 radiochromic films were inserted into the ipisilateral breast, contralateral breast, lungs, heart and skin. The irradiation was carried out on 4/80 Varian linear accelerator at 4MV.
RESULTS: The mean dose at the OAR's presented statistically significant differences (p < 0.001) of 34.24%, 37.96% and 63.47% for ipsilateral lung, contralateral lung, and heart, respectively. The films placed at the skin-surface interface in the ipsilateral breast also showed statistically significant differences (p < 0.001) of 16.43%, -10.16%, -14.79% and 15.67% in the four quadrants, respectively. In contrast, the PTV dosimeters, representative of the left breast volume, encompassed by the electronic equilibrium, presented a non-significant difference with TPS, p = 0.20 and p = 0.90.
CONCLUSION: There was a non-significant difference of doses in PTV with electronic equilibrium; although no match is achieved outside electronic equilibrium.
© 2021 Greater Poland Cancer Centre.

Entities:  

Keywords:  breast cancer; radiochromic film; thorax phantom

Year:  2021        PMID: 34211775      PMCID: PMC8241292          DOI: 10.5603/RPOR.a2021.0037

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  20 in total

1.  Comparison of inhomogeneity correction algorithms in small photon fields.

Authors:  Andrew O Jones; Indra J Das
Journal:  Med Phys       Date:  2005-03       Impact factor: 4.071

2.  Accurate skin dose measurements using radiochromic film in clinical applications.

Authors:  S Devic; J Seuntjens; W Abdel-Rahman; M Evans; M Olivares; E B Podgorsak; Té Vuong; Christopher G Soares
Journal:  Med Phys       Date:  2006-04       Impact factor: 4.071

3.  Comparison of dose calculation algorithms for treatment planning in external photon beam therapy for clinical situations.

Authors:  Tommy Knöös; Elinore Wieslander; Luca Cozzi; Carsten Brink; Antonella Fogliata; Dirk Albers; Håkan Nyström; Søren Lassen
Journal:  Phys Med Biol       Date:  2006-10-24       Impact factor: 3.609

4.  GafChromic EBT film dosimetry with flatbed CCD scanner: a novel background correction method and full dose uncertainty analysis.

Authors:  Sigrun Saur; Jomar Frengen
Journal:  Med Phys       Date:  2008-07       Impact factor: 4.071

5.  Variations of lung density and geometry on inhomogeneity correction algorithms: a Monte Carlo dosimetric evaluation.

Authors:  James C L Chow; Michael K K Leung; Jake Van Dyk
Journal:  Med Phys       Date:  2009-08       Impact factor: 4.071

6.  Accurate dosimetry with GafChromic EBT film of a 6 MV photon beam in water: what level is achievable?

Authors:  L J van Battum; D Hoffmans; H Piersma; S Heukelom
Journal:  Med Phys       Date:  2008-02       Impact factor: 4.071

7.  Evaluation of a Gafchromic EBT2 film dosimetry system for radiotherapy quality assurance.

Authors:  T Aland; T Kairn; J Kenny
Journal:  Australas Phys Eng Sci Med       Date:  2011-04-05       Impact factor: 1.430

8.  Comparison of phantom materials for use in quality assurance of microbeam radiation therapy.

Authors:  Matthew Cameron; Iwan Cornelius; Dean Cutajar; Jeremy Davis; Anatoly Rosenfeld; Michael Lerch; Susanna Guatelli
Journal:  J Synchrotron Radiat       Date:  2017-05-18       Impact factor: 2.616

9.  Forcing lateral electron disequilibrium to spare lung tissue: a novel technique for stereotactic body radiation therapy of lung cancer.

Authors:  Brandon Disher; George Hajdok; Stewart Gaede; Matthew Mulligan; Jerry J Battista
Journal:  Phys Med Biol       Date:  2013-09-09       Impact factor: 3.609

10.  Electronic equilibrium and primary dose in collimated photon beams.

Authors:  P S Nizin
Journal:  Med Phys       Date:  1993 Nov-Dec       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.