Daiane Nogueira1, Carlos Eduardo Da Silveira Bueno1, Augusto Shoji Kato1, Alexandre Sigrist De Martin1, Rina Andréa Pelegrine1, Ana Grasiela Da Silva Limoeiro2, Daniel Guimarães Pedro Rocha3, Carlos Eduardo Fontana4.
Abstract
CONTEXT: It is essential to know if the immersion in sodium hypochlorite (NaOCl) influences the cyclic fatigue resistance of nickel-titaniumm (NiTi) files. AIM: This study was aimed to evaluate the resistance to the cyclic fatigue of three NiTi systems, as well as the influence of 5.25% NaOCl on their resistance.Entities:
Keywords: Corrosion; cyclic fatigue; rotary nickel-titanium instruments; sodium hypochlorite
Year: 2021 PMID: 34083908 PMCID: PMC8095701 DOI: 10.4103/JCD.JCD_117_19
Source DB: PubMed Journal: J Conserv Dent ISSN: 0972-0707
Mean of cycles (in absolute numbers±standard deviation) until fracture in the different groups, submitted to different treatments.
| SEQUENCE | HYFLEX | PROTAPER NEXT | |
|---|---|---|---|
| Control | 372.4±113.6Ca | 3169.0±155.3Aa | 622.8±54.4Ba |
| NaOCl - 1 minute | 370.0±147.1Ca | 2692.5±331.5Ab | 596.7±139.9Ba |
| NaOCl - 5 minutes | 368.4±168.9Ca | 2214.5±191.3Ac | 563.1±94.7Ba |
Legend: Mean and standard deviation off the number of cycles until the instrument fracture, in different experimental times. Equal capital letters indicate statistical similarity between groups, while lower case letters indicate intragroup relations.
Mean of the size of the fragments (in millimeters±standard deviation) after the fracture in the different groups, submitted to different treatments.
| SEQUENCE | HYFLEX | PROTAPER NEXT | |
|---|---|---|---|
| Control | 9,45±0,54Aa | 8,02±0,67Ba | 6,49±0,81Ca |
| NaOCl - 1 minute | 7,18±0,197Ab | 7,81±0,58Aa | 5,91±0,66Ba |
| NaOCl - 5 minutes | 7,78±0,61Ab | 8,33±0,58Aa | 5,54±0,85Ba |
Legend: Mean and standard deviation off the number of cycles until the instrument fracture, in different experimental times. Equal capital letters indicate statistical similarity between groups, while lower case letters indicate intragroup relations.
Figure 1(a, c and e): Cross section of the Sequence, Hyflex EDM and ProTaper Next instruments, respectively, by ×200. (b, d, and f): Instruments Sequence, Hyflex EDM and ProTa -per Next, respectively, at ×3000 increase, displaying the “dimples,” “cracks”