Literature DB >> 26864081

Structural analysis of HyFlex EDM instruments.

F Iacono1, C Pirani1, L Generali2, G Bolelli3, P Sassatelli3, L Lusvarghi3, M G Gandolfi1, L Giorgini4, C Prati1.   

Abstract

AIM: To compare the phase transformation behaviour, the microstructure, the nano-hardness and the surface chemistry of electro-discharge machined HyFlex EDM instruments with conventionally manufactured HyFlex CM.
METHODOLOGY: New and laboratory used HyFlex EDM were examined by X-ray diffraction (XRD) and differential scanning calorimetry (DSC). Nano-hardness and modulus of elasticity were also investigated using a maximum load of 20 mN with a minimum of 40 significant indentations for each sample. Raman spectroscopy and field emission-scanning electron microscope (FE-SEM) were used to assess the surface chemistry of HyFlex EDM. HyFlex CM were subjected to the same investigations and used as a comparison. Nano-indentation data were statistically analysed using the Student's t-test.
RESULTS: XRD analysis on HyFlex EDM revealed the presence of martensite and rhombohedral R-phase, while a mixture of martensite and austenite structure was identified in HyFlex CM. DSC analysis also disclosed higher austenite finish (Af) temperatures for electro-discharge machining (EDM) instruments. Significant differences in nano-hardness and modulus of elasticity were found between EDM and CM files (P < 0.05). FE-SEM and EDS analyses confirmed that both new EDM and CM files were covered by an oxide layer. Micro-Raman spectroscopy assessed the presence of rutile-TiO2 .
CONCLUSIONS: HyFlex EDM revealed peculiar structural properties, such as increased phase transformation temperatures and hardness. Present results corroborated previous findings and shed light on the enhanced mechanical behaviour of these instruments.
© 2016 International Endodontic Journal. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  CM wire; NiTi endodontic instruments; differential scanning calorimetry; electro-discharge machining; metallurgical properties; microstructure

Mesh:

Substances:

Year:  2016        PMID: 26864081     DOI: 10.1111/iej.12620

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  21 in total

1.  Effect of multiple autoclave cycles on the surface roughness of HyFlex CM and HyFlex EDM files: an atomic force microscopy study.

Authors:  K Yılmaz; G Uslu; T Özyürek
Journal:  Clin Oral Investig       Date:  2018-02-13       Impact factor: 3.573

2.  Effect of autoclave sterilization on cyclic fatigue and torsional fracture resistance of NiTi rotary instruments.

Authors:  Wooyoung Kim; Soram Oh; Gil-Joo Ryu; Tae-Hwan Kim; Sung-Jae Kim; Dong-Hyung Kim; Bin-Na Lee; Kee-Yeon Kum; Seok Woo Chang; Ji-Hyun Jang
Journal:  Odontology       Date:  2019-09-11       Impact factor: 2.634

Review 3.  Endodontic Rotary Files, What Should an Endodontist Know?

Authors:  Ana-Belén Dablanca-Blanco; Pablo Castelo-Baz; Ramón Miguéns-Vila; Pablo Álvarez-Novoa; Benjamín Martín-Biedma
Journal:  Medicina (Kaunas)       Date:  2022-05-27       Impact factor: 2.948

4.  Comparison of the cyclic fatigue resistance of One Curve, F6 Skytaper, Protaper Next, and Hyflex CM endodontic files.

Authors:  Charlotte Gouédard; Laurent Pino; Reza Arbab-Chirani; Shabnam Arbab-Chirani; Valérie Chevalier
Journal:  Restor Dent Endod       Date:  2022-03-04

5.  Comparative evaluation of apically extruded debris using three different thermomechanically heat treated file systems with two different motions: An in vitro study.

Authors:  Thangi Sowjanya; Krishna Prasad Parvathaneni; Tbvg Raju; N Mahendra Varma; Gowtham Dev Dondapati; Srivalli Podili
Journal:  J Conserv Dent       Date:  2022-06-13

6.  Evaluation of the Fractured Surface of Five Endodontic Rotary Instruments: A Metallurgical Study.

Authors:  Mohsen Aminsobhani; Mohamad Saleh Khalatbari; Naghmeh Meraji; Abdollah Ghorbanzadeh; Ehsan Sadri
Journal:  Iran Endod J       Date:  2016

7.  Comparison of Shaping Ability of 10 Rotary and Reciprocating Systems: an In Vitro Study with AutoCad.

Authors:  Jorge Rubio; José Ignacio Zarzosa; Antonio Pallarés
Journal:  Acta Stomatol Croat       Date:  2017-09

8.  Ex-Vivo Comparison of Torsional Stress on Nickel-Titanium Instruments Activated by Continuous Rotation or Adaptive Motion.

Authors:  Joo Yeong Lee; Sang Won Kwak; Jung-Hong Ha; Hyeon-Cheol Kim
Journal:  Materials (Basel)       Date:  2020-04-17       Impact factor: 3.623

Review 9.  Nickel-Titanium Rotary File Systems: What's New?

Authors:  Sadia Tabassum; Kamil Zafar; Fahad Umer
Journal:  Eur Endod J       Date:  2019-10-18

10.  Micro-computed tomographic evaluation of dentinal defects after root canal preparation with hyflex edm and vortex blue rotary systems.

Authors:  Jyothi Mandava; Rajiv-Kumar Yelisela; Sampath-Kumar Arikatla; Ravi-Chandra Ravi
Journal:  J Clin Exp Dent       Date:  2018-09-01