| Literature DB >> 36267671 |
Hana R Ubaed1, Diyar Kh Bakr1.
Abstract
Objective: Cyclic fatigue occurred in the curved canal when the instrument freely rotated leading to repeated compressive and tensile stresses. This study aims to evaluate the cyclic fatigue resistance (CFR) of new and used 2Shape and AF F-One rotary instrument systems by using an artificial stainless-steel canal.Entities:
Year: 2022 PMID: 36267671 PMCID: PMC9578904 DOI: 10.1155/2022/1716008
Source DB: PubMed Journal: Appl Bionics Biomech ISSN: 1176-2322 Impact factor: 1.664
Figure 1Dimensions of the artificial canal performed by (AutoCAD).
Figure 2Dental handpiece fixation on the dental surveyor.
Figure 3Instrument fracture.
Descriptive statistics analysis of both groups.
| Groups | Usage time |
| Mean ± SD | 95% confidence interval for mean | |
|---|---|---|---|---|---|
| Lower bound | Upper bound | ||||
| AF F-ONE | New | 10 | 3758.650 ± 1074.677 | 2989.872 | 4527.428 |
| One use | 10 | 1547.700 ± 640.416 | 1089.574 | 2005.826 | |
| Two use | 10 | 1653.050 ± 360.143 | 1395.419 | 1910.681 | |
| Three use | 10 | 1754.900 ± 293.125 | 1545.211 | 1964.589 | |
| 2Shape | New | 10 | 766.200 ± 119.509 | 680.708 | 851.692 |
| One use | 10 | 714.300 ± 81.147 | 656.251 | 772.349 | |
| Two use | 10 | 700.200 ± 89.924 | 635.872 | 764.528 | |
| Three use | 10 | 648.900 ± 135.934 | 551.659 | 746.141 | |
Normality, homogeneity, and ANOVA test outcome for instruments in AF F-one group.
| Groups | Instrument type | Shapiro–Wilk ( | Homogeneity test | ANOVA test |
|---|---|---|---|---|
| ( | ( | |||
| AF F-one | New | 0.852 (0.062) | 4.783 (0.007) | 25.086 (0.000) |
| One use | 0.782 (0.065) | |||
| Two use | 0.948 (0.645) | |||
| Three use | 0.920 (0.358) |
Pairwise multiple comparison result for AF F-one group.
| AF F-one group | Mean difference (I–J) | Sig. | 95% confidence interval | ||
|---|---|---|---|---|---|
| Lower bound | Upper bound | ||||
| New | One use | 2210.950* | 0.000 | 1982.838 | 4105.862 |
| Two use | 2105.600* | 0.001 | 1026.864 | 3184.336 | |
| Three use | 2003.750* | 0.001 | 932.312 | 3075.188 | |
| One use | Two use | −105.750* | 0.000 | −1296.485 | −581.015 |
| Three use | −207.600* | 0.000 | −1332.950 | −748.250 | |
| Two use | Three use | −101.850* | 0.898 | −518.540 | 314.840 |
∗= mean difference of number of cycles to fracture between two group of instruments.
- = number of cycles to fracture in this group was increased.
Normality, homogeneity and Kruskal–Wallis test for 2Shape/TS2 group.
| Groups | Instrument type | Shapiro–Wilk | Homogeneity test | Kruskal–Wallis |
|---|---|---|---|---|
| ( | ( | |||
| 2Shape/TS2 | New | 0.766 (0.006) | 0.879 (0.461) | 6.526 (0.089) |
| One use | 0.854 (0.065) | |||
| Two use | 0.753 (0.004) | |||
| Three use | 0.928 (0.429) |
Between groups non-parametric pairwise comparison results.
| Pairwise comparison | Mean difference | Mean rank difference | Mann–Whitney |
|---|---|---|---|
| New/AF F-one–New/2Shape/TS2 | 2992.450 | 10.000 | 0.000 (0.000) |
| One use/AF F-one–one use/2Shape/TS2 | 833.400 | 9.200 | 4.000 (0.001) |
| Two use/AF F-one–two use/2Shape/TS2 | 952.850 | 10.000 | 0.000 (0.000) |
| Three use/AF F-one–three use/2Shape/TS2 | 1106.000 | 10.000 | 0.000 (0.000) |