Literature DB >> 33955655

Validation of a method for estimating peak skin dose from CT-guided procedures.

A Kyle Jones1, Meghan E Kisiel1, X John Rong1, Alda L Tam2.   

Abstract

A method for estimating peak skin dose (PSD) from CTDIvol has been published but not validated. The objective of this study was to validate this method during CT-guided ablation procedures. Radiochromic film was calibrated and used to measure PSD. Sixty-eight patients were enrolled in this study, and measured PSD were collected for 46 procedures. CTDIvol stratified by axial and helical scanning was used to calculate an estimate of PSD using the method [1.2 × CTDIvol (helical) + 0.6 × CTDIvol (axial)], and both calculated PSD and total CTDIvol were compared to measured PSD using paired t-tests on the log-transformed data and Bland-Altman analysis. Calculated PSD were significantly different from measured PSD (P < 0.0001, bias, 18.3%, 95% limits of agreement, -63.0% to 26.4%). Measured PSD were not significantly different from total CTDIvol (P = 0.27, bias, 3.97%, 95% limits of agreement, -51.6% to 43.7%). Considering that CTDIvol is reported on the console of all CT scanners, is not stratified by axial and helical scanning modes, and is immediately available to the operator during CT-guided interventional procedures, it may be reasonable to use the scanner-reported CTDIvol as an indicator of PSD during CT-guided procedures. However, further validation is required for other models of CT scanner.
© 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  CT dosimetry; CT-guided intervention; CTDIvol; ablation; peak skin dose

Year:  2021        PMID: 33955655      PMCID: PMC8200428          DOI: 10.1002/acm2.13261

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


INTRODUCTION

The scope and complexity of computed tomography (CT)‐guided interventions is increasing, and CT is now used in combination with angiography in hybrid imaging systems. , , , Patient dose indices vary depending on the complexity of the procedure, and patients can be exposed to high peak skin doses (PSD) during the most complex procedures, as the same skin site is scanned repeatedly. This is very similar to the pattern of irradiation during dynamic imaging, including CT perfusion. The potential effects of high PSD delivered during such procedures were demonstrated in the CT perfusion incidents that occurred in the mid‐2000’s, with patients experiencing transient and permanent alopecia and an incident during which a young child experienced erythema after repeated CT of the head. Radiation cataractogenesis is a related concern if the area of repeated scanning includes the eye lenses. Given the potential complications associated with high PSD experienced during CT, it is useful to understand how scanner‐reported dose indices relate to patient skin and surface doses, as these indices are available to the physician during the procedure, reported in real‐time. Soon after the CT perfusion incidents, studies examining patient doses resulting from CT brain perfusion were published, including a Monte Carlo study of patient lens and skin doses, a phantom study of organ doses, and a phantom study of skin dose. While these studies examined only CT brain perfusion imaging using axial scanning mode, an expression relating patient skin dose to the volume computed tomography dose index (CTDIvol) for interventional CT procedures was published around the same time, based on previous evaluation of the relationship between CTDIvol and surface dose for different scanning modes. This method was based on a linear weighted sum of CTDIvol(helical) and CTDIvol(axial) . The weighting factors were determined from the definitions of CTDI and CTDIvol, the relationship between peripheral and central CTDI measurements in the 32 cm body phantom, and the relationship between peripheral CTDI and surface dose reported by Bauhs et al. The method is as follows: This method, while incorporated into existing recommendations regarding CT‐guided interventional procedures, has never been validated. The purpose of this study was to validate this existing method for calculating patient PSD during CT‐guided interventional procedures. We hypothesized that the method would predict skin doses that were not significantly different from skin doses measured using radiochromic film.

METHODS

This study was conducted in compliance with the Healthcare Insurance Portability and Accountability Act and received prospective Institutional Review Board approval. Informed consent was obtained from all 68 patients enrolled in this study during a five month period. All procedures were performed on a single Definition AS128+ CT (Siemens Healthineers, Malvern, PA) located in the interventional radiology department. Procedures performed included radiofrequency, microwave, and cryo‐ ablation. PSD were measured using Gafchromic XR‐RV3 film (Ashland, New Jersey). The film was cut into 35 cm × 5 cm strips. All film was from a single lot, and the film was calibrated using the same CT scanner used to perform the ablation procedures using standard methods. During calibration, strips of film were placed immediately beneath a 32 cm CTDI phantom, on top of the CT slicker and table pad, with the orange side facing down towards the table. Exposure was measured using a 0.6 cc thimble ionization chamber (10x5‐0.6, Radcal, Monrovia, CA) located adjacent to the film. The 0.6 cc chamber had a National Institute of Standards and Technology (NIST) traceable calibration at 120 kVp (HVL = 7.65 mm Al). Calibration films were exposed to air kerma ranging from 0 mGy to 4,009 mGy at 120 kV. The films were stored in a dark location for 2 weeks prior to scanning, as were films irradiated during patient procedures. All films were scanned using an Epson V700 Professional flatbed scanner (Epson, Long Beach, CA) in reflective mode in 48 bit color at 150 dpi with all corrections disabled. The long axis of the film was oriented parallel to the scan direction, which centered the narrow strips in the scanner. After scanning, film darkening was measured from the red channel data. A four parameter logistic model Rodbard fit was applied to the calibration data in ImageJ (NIH, Bethesda, MD) to derive the calibration function. During CT‐guided interventions, one strip of film was placed in the same location and orientation as for calibration, immediately under the patient. The technologist was instructed to identify the lesion to be treated during the procedure on prior cross‐sectional imaging and then use anatomical landmarks to center the film, oriented along the long axis of the patient, as close as possible to the region of interest. As PSD was the dosimetric quantity of interest, it mattered only that the film was irradiated during a single rotation during each scanning event, i.e., capturing the entire length of long helical scans was not required. All images from every procedure were reviewed to ensure that the film was irradiated during each scan event. The area of maximum film darkening was identified visually and a region of interest containing at least 6,750 pixels was placed to measure the film darkening with the same methods used for calibration. PSD was calculated using the method of Leng et al, listed previously in the introduction. This was accomplished by multiplying CTDIvol from helical scans acquired during clinical procedures by 1.2 and adding this to the CTDIvol from axial scans multiplied by 0.6 for all acquisitions during the procedure. The CDTIvol reported for topograms was not included. CTDIvol reported by the CT scanner used in this study was within 5% of the CTDIvol measured during routine quality assurance testing performed during the study period. CT‐guided procedures at our institution generally follow the workflow outlined in the Best Practice Guidelines for CT‐Guided Interventional Procedures from the Society of Interventional Radiology, including, in order, a topogram, a helical preprocedure planning scan (PPS), axial or helical imaging during the intervention phase, and an optional postprocedure helical scan. Most procedures also included a limited multiphase contrast‐enhanced CT to assess ablation margin adequacy. Otherwise, most helical scanning was performed in i‐Spiral mode using a collimated beam width of 38.4 mm (detector configuration 32 x 1.2 mm) and a pitch of 1.0, and most axial scanning was performed using a collimated beam width of 14.4 mm (detector configuration 3 x 4.8 mm) in i‐Sequence mode. CT fluoroscopy was used occasionally, also at a beam width of 14.4 mm, and was treated the same as axial scanning. Most procedures used 120 kV, with occasional use of other kV depending on patient dimensions, and mAs was adapted to patient size using technique charts. Measured PSD were compared to calculated PSD and to total CTDIvol alone using paired t‐tests on the log‐transformed data. A simple Bonferroni correction was used to correct for multiple comparisons. Five statistical tests were performed during data analysis, resulting in a corrected alpha of 0.05/5 = 0.01. Bland‐Altman analysis was used to examine bias, and correlation was calculated using the Spearman correlation coefficient.

RESULTS

The film calibration data are plotted in Figure 1. One procedure was cancelled, the technologist forgot to place the film for 11 procedures, and no exposure was recorded on the film (i.e., the film was placed outside the imaged field of view) for 10 procedures, leaving 46 procedures available for analysis. Measured PSD ranged from 30.5 mGy to 1,303.9 mGy, with a median of 263.1 mGy.
Fig. 1

Radiochromic film calibration data plotted as air kerma versus reflective density.

Radiochromic film calibration data plotted as air kerma versus reflective density. The fraction of CTDIvol contributed by axial scanning varied widely (Fig. 2). Calculated PSD were significantly different from measured PSD (P < 0.0001). Calculated PSD underestimated the measured PSD by 18.3% (Fig. 3a, 95% limits of agreement, −63.0% to 26.4%). The standard deviation of the differences between calculated PSD and measured PSD across all procedures was 20.5 mGy. Total CTDIvol was not significantly different from measured PSD (P = 0.27) and underestimated the measured PSD by 3.97% (Fig. 3b, 95% limits of agreement, −51.6% to 43.7%). The standard deviation of the differences between CTDIvol and measured PSD across all procedures was 24.1 mGy.
Fig. 2

Histogram of the fraction of total procedural CTDIvol contributed by axial scanning across all 46 procedures for which film dosimetry was available.

Fig. 3

Bland‐Altman plots of (a) percent difference between peak skin dose (PSD) calculated using an existing method and measured PSD and (b) percent difference between total CTDIvol and measured PSD. The dashed line indicates bias and the dotted lines indicate the 95% limits of agreement (LoA), and the x‐axis title “Average (mGy)” is the average of the estimates produced by the two measurement methods.

Histogram of the fraction of total procedural CTDIvol contributed by axial scanning across all 46 procedures for which film dosimetry was available. Bland‐Altman plots of (a) percent difference between peak skin dose (PSD) calculated using an existing method and measured PSD and (b) percent difference between total CTDIvol and measured PSD. The dashed line indicates bias and the dotted lines indicate the 95% limits of agreement (LoA), and the x‐axis title “Average (mGy)” is the average of the estimates produced by the two measurement methods.

DISCUSSION

In the specific context of this study, CTDIvol alone was a better predictor of PSD during CT‐guided interventions than PSD calculated using an existing method. The 95% limits of agreement for the difference between CTDIvol and measured PSD indicated that CTDIvol can predict the PSD within 50%, a reasonable benchmark for patient dosimetry with indirect dose indices. While the existing method also, in general, predicted PSD within 50%, bias was greater, and use of the existing method required that CTDIvol be stratified into contributions from axial and helical scanning, which is somewhat cumbersome and difficult to do during the procedure at the scanner console. It has been reported previously that surface dose as a percentage of CTDIvol for axial scanning increases with decreasing phantom size, from 49% to 65% for phantom lateral widths ranging from 50 cm to 25 cm. There was a weak, non‐significant correlation ( ρ 0.30, P = 0.041) between percent error in calculated PSD compared to measured PSD and patient effective diameter (Fig. 4a). There was no correlation ( ρ 0.25, P = 0.093) between percent error in CTDIvol compared to measured PSD and patient effective diameter (Fig. 4b). There was a weak, significant correlation ( ρ 0.44, P = 0.0023) between percent error in CTDIvol compared to measured PSD and the fraction of CTDIvol contributed by axial scanning (Fig. 4c). These findings likely reflect the complex interactions among the factors that affect the relationship between patient surface dose and CTDIvol, including scanning mode, patient size, and the position of the patient relative to isocenter, as patient access and probe placement considerations often mean that the patient is not centered in the gantry.
Fig. 4

(a) Percent error in PSD was calculated using an existing method compared to measured PSD versus patient effective diameter; (b) percent error in CTDIvol compared to measured PSD versus patient effective diameter; and (c) percent error in CTDIvol compared to measured PSD versus fraction of CTDIvol contributed by axial scanning.

(a) Percent error in PSD was calculated using an existing method compared to measured PSD versus patient effective diameter; (b) percent error in CTDIvol compared to measured PSD versus patient effective diameter; and (c) percent error in CTDIvol compared to measured PSD versus fraction of CTDIvol contributed by axial scanning. There are several uncertainties associated with radiochromic film dosimetry, these have been discussed extensively in a review by McCabe et al and summarized elsewhere. Considering that calibration of the film for the present study was done using backscatter conditions, the accuracy limits on measured PSD in this study were approximately ±10%, as calculated using the summary information in Table II of Ref. 18. While this study included a variety of CT‐guided ablations performed on patients of different sizes, all procedures were performed using a single modern CT scanner at a single site, a limitation of this study. The relative contributions of primary and scatter radiation to CTDIvol vary as a function of beam width , and it has been reported that surface dose as a percentage of CTDIvol for axial scanning increases with increasing beam width, from 68% to 89% for beam widths ranging from 5 mm to 28.8 mm. However, the beam widths used for most scanning in this study, 14.4 mm for axial scanning and 38.4 mm for helical scanning, are very common in modern CT. Even so, our findings are still somewhat scanner‐specific, as other aspects such as beam quality, determined by X‐ray tube and generator design and bowtie filter shapes and materials, affect the relationship between peripheral and central CTDI100 and the relationship between surface dose and CTDIvol, as reported by las Heras et al. las Heras found that for the 16 cm CTDI head phantom the ratio of PSD to CTDIvol varied by up to 25% among 4 CT scanner models. However, the scan protocols used in their study used 1, 2, or 4 gantry rotations to cover the imaged volume, leading to substantial variations in surface dose across the imaged volume.

CONCLUSION

Peak skin doses (PSD) resulting from CT‐guided procedures, calculated using an existing method, were significantly different from measured PSD for procedures performed using a single make and model of CT scanner. Total CTDIvol was not significantly different from measured PSD in the specific context of this study.

CONFLICT OF INTEREST

Alda L. Tam has the following disclosures: Research grant: BTG, Guerbet; Consulting: Boston Scientific, Endocare. None of the other authors has anything to disclose. This study was conducted in compliance with the Healthcare Insurance Portability and Accountability Act and received prospective Institutional Review Board approval.
  15 in total

Review 1.  CT dosimetry: comparison of measurement techniques and devices.

Authors:  John A Bauhs; Thomas J Vrieze; Andrew N Primak; Michael R Bruesewitz; Cynthia H McCollough
Journal:  Radiographics       Date:  2008 Jan-Feb       Impact factor: 5.333

2.  Experimental estimates of peak skin dose and its relationship to the CT dose index using the CTDI head phantom.

Authors:  Hugo de las Heras; Ronaldo Minniti; Sean Wilson; Chad Mitchell; Marlene Skopec; Claudia C Brunner; Kish Chakrabarti
Journal:  Radiat Prot Dosimetry       Date:  2013-07-16       Impact factor: 0.972

3.  Estimation of radiation exposure for brain perfusion CT: standard protocol compared with deviations in protocol.

Authors:  Jenny K Hoang; Chu Wang; Donald P Frush; David S Enterline; Ehsan Samei; Greta Toncheva; Carolyn Lowry; Terry T Yoshizumi
Journal:  AJR Am J Roentgenol       Date:  2013-09-24       Impact factor: 3.959

4.  Calibration of GafChromic XR-RV3 radiochromic film for skin dose measurement using standardized x-ray spectra and a commercial flatbed scanner.

Authors:  Bradley P McCabe; Michael A Speidel; Tina L Pike; Michael S Van Lysel
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

5.  How accurately can the peak skin dose in fluoroscopy be determined using indirect dose metrics?

Authors:  A Kyle Jones; Joe E Ensor; Alexander S Pasciak
Journal:  Med Phys       Date:  2014-07       Impact factor: 4.071

6.  Best Practice Guidelines for CT-Guided Interventional Procedures.

Authors:  A Kyle Jones; Robert G Dixon; Jeremy D Collins; Eric M Walser; Boris Nikolic
Journal:  J Vasc Interv Radiol       Date:  2018-02-15       Impact factor: 3.464

7.  Comparative Analysis of Intra-arterial Cone-Beam Versus Conventional Computed Tomography During Hepatic Arteriography for Transarterial Chemoembolization Planning.

Authors:  Ethan Y Lin; A Kyle Jones; Gouthami Chintalapani; Zachary S Jeng; Joe Ensor; Bruno C Odisio
Journal:  Cardiovasc Intervent Radiol       Date:  2018-11-09       Impact factor: 2.740

8.  Targeted transarterial oily chemoembolization for small foci of hepatocellular carcinoma using a unified helical CT and angiography system: analysis of factors affecting local recurrence and survival rates.

Authors:  K Takayasu; Y Muramatsu; T Maeda; R Iwata; H Furukawa; Y Muramatsu; N Moriyama; T Okusaka; S Okada; H Ueno
Journal:  AJR Am J Roentgenol       Date:  2001-03       Impact factor: 3.959

9.  Procedure-specific CT Dose and Utilization Factors for CT-guided Interventional Procedures.

Authors:  Kai Yang; Suvranu Ganguli; Matthew C DeLorenzo; Hui Zheng; Xinhua Li; Bob Liu
Journal:  Radiology       Date:  2018-07-17       Impact factor: 11.105

10.  Comparison of radiation dose and image quality between flat panel computed tomography and multidetector computed tomography in a hybrid CT-angiography suite.

Authors:  Aaron K Jones; Bruno C Odisio
Journal:  J Appl Clin Med Phys       Date:  2020-01-10       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.