| Literature DB >> 23864642 |
Hugo de las Heras1, Ronaldo Minniti, Sean Wilson, Chad Mitchell, Marlene Skopec, Claudia C Brunner, Kish Chakrabarti.
Abstract
A straightforward method is presented to estimate peak skin doses (PSDs) delivered by computed tomography (CT) scanners. The measured PSD values are related to the well-known volume CT dose index (CTDI(vol)), displayed on the console of CT scanners. PSD measurement estimates were obtained, in four CT units, by placing radiochromic film on the surface of a CTDI head phantom. Six different X-ray tube currents including the maximum allowed value were used to irradiate the phantom. PSD and CTDI(vol) were independently measured and later related to the CTDI(vol) value displayed on the console. A scanner-specific relationship was found between the measured PSD and the associated CTDI(vol) displayed on the console. The measured PSD values varied between 27 and 136 mGy among all scanners when the routine head scan parameters were used. The results of this work allow relating the widely used CTDI(vol) to an actual radiation dose delivered to the skin of a patient.Entities:
Mesh:
Substances:
Year: 2013 PMID: 23864642 PMCID: PMC3853653 DOI: 10.1093/rpd/nct171
Source DB: PubMed Journal: Radiat Prot Dosimetry ISSN: 0144-8420 Impact factor: 0.972