Calibration of GafChromic XR-RV3 radiochromic film for skin dose measurement using standardized x-ray spectra and a commercial flatbed scanner.

Bradley P McCabe1, Michael A Speidel, Tina L Pike, Michael S Van Lysel.   

Abstract

PURPOSE: In this study, newly formulated XR-RV3 GafChromic film was calibrated with National Institute of Standards and Technology (NIST) traceability for measurement of patient skin dose during fluoroscopically guided interventional procedures.
METHODS: The film was calibrated free-in-air to air kerma levels between 15 and 1100 cGy using four moderately filtered x-ray beam qualities (60, 80, 100, and 120 kVp). The calibration films were scanned with a commercial flatbed document scanner. Film reflective density-to-air kerma calibration curves were constructed for each beam quality, with both the orange and white sides facing the x-ray source. A method to correct for nonuniformity in scanner response (up to 25% depending on position) was developed to enable dose measurement with large films. The response of XR-RV3 film under patient backscattering conditions was examined using on-phantom film exposures and Monte Carlo simulations.
RESULTS: The response of XR-RV3 film to a given air kerma depended on kVp and film orientation. For a 200 cGy air kerma exposure with the orange side of the film facing the source, the film response increased by 20% from 60 to 120 kVp. At 500 cGy, the increase was 12%. When 500 cGy exposures were performed with the white side facing the x-ray source, the film response increased by 4.0% (60 kVp) to 9.9% (120 kVp) compared to the orange-facing orientation. On-phantom film measurements and Monte Carlo simulations show that using a NIST-traceable free-in-air calibration curve to determine air kerma in the presence of backscatter results in an error from 2% up to 8% depending on beam quality. The combined uncertainty in the air kerma measurement from the calibration curves and scanner nonuniformity correction was +/- 7.1% (95% C.I.). The film showed notable stability. Calibrations of film and scanner separated by 1 yr differed by 1.0%.
CONCLUSIONS: XR-RV3 radiochromic film response to a given air kerma shows dependence on beam quality and film orientation. The presence of backscatter slightly modifies the x-ray energy spectrum; however, the increase in film response can be attributed primarily to the increase in total photon fluence at the sensitive layer. Film calibration curves created under free-in-air conditions may be used to measure dose from fluoroscopic quality x-ray beams, including patient backscatter with an error less than the uncertainty of the calibration in most cases.

Entities:  

Mesh:

Year:  2011        PMID: 21626925      PMCID: PMC3078021          DOI: 10.1118/1.3560422

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  38 in total

1.  Radiation doses in interventional radiology procedures: the RAD-IR Study. Part III: Dosimetric performance of the interventional fluoroscopy units.

Authors:  Stephen Balter; Beth A Schueler; Donald L Miller; Patricia E Cole; Hollington T Lu; Alejandro Berenstein; Robin Albert; Jeffrey D Georgia; Patrick T Noonan; Eric J Russell; Tim W Malisch; Robert L Vogelzang; Michael Geisinger; John F Cardella; James St George; George L Miller; Jon Anderson
Journal:  J Vasc Interv Radiol       Date:  2004-09       Impact factor: 3.464

2.  Evaluation of risk of deterministic effects in fluoroscopically guided procedures.

Authors:  E Vano; L Gonzalez; E Guibelalde; P Aviles; J M Fernandez; C Prieto; C Galvan
Journal:  Radiat Prot Dosimetry       Date:  2006-02-03       Impact factor: 0.972

3.  Radiochromic film dosimetry: reflection vs transmission scanning.

Authors:  J Kalef-Ezra; K Karava
Journal:  Med Phys       Date:  2008-06       Impact factor: 4.071

4.  X-ray burns--painful, protracted, and preventable.

Authors:  Ronald E Vlietstra; Louis K Wagner
Journal:  Clin Cardiol       Date:  2008-04       Impact factor: 2.882

5.  Analysis of uncertainties in Gafchromic EBT film dosimetry of photon beams.

Authors:  Mária Martisíková; Benjamin Ackermann; Oliver Jäkel
Journal:  Phys Med Biol       Date:  2008-11-18       Impact factor: 3.609

6.  GafChromic EBT film dosimetry with flatbed CCD scanner: a novel background correction method and full dose uncertainty analysis.

Authors:  Sigrun Saur; Jomar Frengen
Journal:  Med Phys       Date:  2008-07       Impact factor: 4.071

7.  The characterization of a commercial MOSFET dosimeter system for use in diagnostic x ray.

Authors:  M W Bower; D E Hintenlang
Journal:  Health Phys       Date:  1998-08       Impact factor: 1.316

8.  Total entrance skin dose: an effective indicator of maximum radiation dose to the skin during percutaneous coronary intervention.

Authors:  Koichi Chida; Yutaka Kagaya; Haruo Saito; Yoshihiro Takai; Shoki Takahashi; Shogo Yamada; Masahiro Kohzuki; Masayuki Zuguchi
Journal:  AJR Am J Roentgenol       Date:  2007-10       Impact factor: 3.959

9.  Use of GAFCHROMIC XR type R films for skin-dose measurements in interventional radiology: Validation of a dosimetric procedure on a sample of patients undergone interventional cardiology.

Authors:  S Delle Canne; A Carosi; A Bufacchi; T Malatesta; R Capperella; R Fragomeni; N Adorante; S Bianchi; L Begnozzi
Journal:  Phys Med       Date:  2006 July - September       Impact factor: 2.685

10.  A study of GafChromic XR Type R film response with reflective-type densitometers and economical flatbed scanners.

Authors:  G Thomas; R Y L Chu; Frank Rabe
Journal:  J Appl Clin Med Phys       Date:  2003       Impact factor: 2.102

View more
  8 in total

1.  How accurately can the peak skin dose in fluoroscopy be determined using indirect dose metrics?

Authors:  A Kyle Jones; Joe E Ensor; Alexander S Pasciak
Journal:  Med Phys       Date:  2014-07       Impact factor: 4.071

2.  Monte Carlo simulation of inverse geometry x-ray fluoroscopy using a modified MC-GPU framework.

Authors:  David A P Dunkerley; Michael T Tomkowiak; Jordan M Slagowski; Bradley P McCabe; Tobias Funk; Michael A Speidel
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2015-02-21

3.  High throughput film dosimetry in homogeneous and heterogeneous media for a small animal irradiator.

Authors:  L Wack; W Ngwa; E Tryggestad; P Tsiamas; R Berbeco; S K Ng; J Hesser; P Zygmanski
Journal:  Phys Med       Date:  2013-03-17       Impact factor: 2.685

4.  Validation of a method for estimating peak skin dose from CT-guided procedures.

Authors:  A Kyle Jones; Meghan E Kisiel; X John Rong; Alda L Tam
Journal:  J Appl Clin Med Phys       Date:  2021-05-06       Impact factor: 2.102

5.  Evaluation of different techniques for CT radiation profile width measurement.

Authors:  Steven R Jackson; Salahuddin Ahmad; Yida Hu; Chun Ruan
Journal:  J Appl Clin Med Phys       Date:  2013-07-08       Impact factor: 2.102

6.  Exposure of the eye lens and brain for interventional cardiology staff.

Authors:  Joanna Kidoń; Kinga Polaczek-Grelik; Przemysław Żurek; Wojciech Wojakowski; Andrzej Ochala
Journal:  Postepy Kardiol Interwencyjnej       Date:  2021-09-28       Impact factor: 1.426

7.  Dose evaluation for skin and organ in hepatocellular carcinoma during angiographic procedure.

Authors:  Daniela D'Alessio; Claudia Giliberti; Antonella Soriani; Livio Carpanese; Giuseppe Pizzi; Giulio Vallati; Lidia Strigari
Journal:  J Exp Clin Cancer Res       Date:  2013-10-25

8.  Measurement of entrance surface dose on an anthropomorphic thorax phantom using a miniature fiber-optic dosimeter.

Authors:  Wook Jae Yoo; Sang Hun Shin; Dayeong Jeon; Seunghan Hong; Hyeok In Sim; Seon Geun Kim; Kyoung Won Jang; Seunghyun Cho; Won Sik Youn; Bongsoo Lee
Journal:  Sensors (Basel)       Date:  2014-04-01       Impact factor: 3.576

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.