| Literature DB >> 33848418 |
Chan Il Song1, Hyong-Ho Cho2, Byung Yoon Choi3, Jae Young Choi1, Jin Woong Choi4, Yun-Hoon Choung5, Jong Woo Chung6, Won-Ho Chung7, Sung Hwa Hong8, Yehree Kim6, Byung Don Lee9, Il-Woo Lee10, Jong Dae Lee11, Jun Ho Lee12, Kyu-Yup Lee13, Il Joon Moon7, In Seok Moon1, Seung-Ha Oh12, Hong Ju Park6, Shi Nae Park14, Ji Won Seo8.
Abstract
OBJECTIVES: This study was conducted to evaluate the user satisfaction, efficacy, and safety of round window (RW) vibroplasty using the Vibrant Soundbridge (VSB) in patients with persistent mixed hearing loss after mastoidectomy.Entities:
Keywords: Active Middle Ear Implant; Mixed Hearing Loss; Round Window; Vibrant Soundbridge; Vibroplasty
Year: 2021 PMID: 33848418 PMCID: PMC8901952 DOI: 10.21053/ceo.2020.01851
Source DB: PubMed Journal: Clin Exp Otorhinolaryngol ISSN: 1976-8710 Impact factor: 3.372
Patient demographics and characteristics
| Patient No. | Sex | History of HA | Age at RW | Surgery ear | Previous ear surgery | ||
|---|---|---|---|---|---|---|---|
| Diagnosis | Type of mastoidectomy | Ossiculoplasty | |||||
| 1 | M | No | 52 | Right | Chole | Down | 1 |
| 2 | F | Yes | 54 | Right | COM | Down | 1 |
| 3 | F | Yes | 69 | Left | COM | Down | 1 |
| 4 | M | Yes | 56 | Right | Chole | Up | 1 |
| 5 | M | Yes | 59 | Left | COM | Down | 1 |
| 6 | M | Yes | 65 | Left | Chole | Down | 1 |
| 7 | M | Yes | 50 | Left | COM | Down | 1 |
| 8 | F | No | 70 | Left | COM | Down | 1 |
| 9 | F | Yes | 60 | Left | COM | Down | 2 |
| 10 | F | Yes | 63 | Right | Chole | Down | 1 |
| 11 | M | No | 47 | Right | COM | Down | 1 |
| 12 | F | Yes | 65 | Left | COM | Down | 2 |
| 13 | F | No | 28 | Left | COM | Up | 3 |
| 14 | F | No | 67 | Left | COM | Up | 1 |
| 15 | F | No | 58 | Left | COM | Down | 1 |
| 16 | F | No | 48 | Left | Chole | Down | 2 |
| 17 | F | Yes | 76 | Right | Chole | Down | 3 |
| 18 | F | Yes | 53 | Left | Chole | Down | 2 |
| 19 | M | Yes | 61 | Left | Chole | Up | 1 |
| 20 | F | No | 56 | Left | COM | Down | 3 |
| 21 | M | Yes | 57 | Right | COM | Down | 1 |
| 22 | M | No | 47 | Left | COM | Up | 1 |
| 23 | F | No | 72 | Left | COM | Down | 2 |
| 24 | F | No | 62 | Left | COM | Down | 1 |
| 25 | F | Yes | 58 | Left | COM | Up | 2 |
| 26 | M | Yes | 56 | Left | COM | Down | 2 |
| 27 | M | Yes | 76 | Right | COM | Down | 2 |
HA, hearing aids; RW, round window; Chole, chronic otitis media with cholesteatoma; COM, chronic otitis media without cholesteatoma; Down, canal wall–down mastoidectomy; Up, canal wall–up mastoidectomy.
Fig. 1.Scores on the Korean translation of the Abbreviated Profile of Hearing Aid Benefit questionnaire, assessed preoperatively and at 3 and 6 months postoperatively. (A) Ease of communication subscale. (B) Reverberation subscale. (C) Background noise subscale. (D) Aversiveness subscale. Preop, preoperative; VSB, Vibrant Soundbridge; VSB 3 mo, VSB-aided thresholds at 3 months after surgery; VSB 6 mo, VSB-aided thresholds at 6 months after surgery.
Fig. 2.Scores on the Korean version of the International Outcome Inventory for Hearing Aids (K-IOI-HA) questionnaire assessed at three different time points: preoperatively and at 3 and 6 months postoperatively. The K-IOI-HA questionnaire was completed before and 3 and 6 months after round window (RW) vibroplasty by 16 patients who used hearing aids before RW vibroplasty. HA, hearing aid; HA (preop), preoperative HA-aided thresholds; VSB, Vibrant Soundbridge; VSB 3 mo, VSB-aided thresholds at 3 months after surgery; VSB 6 mo, VSB-aided thresholds at 6 months after surgery.
Fig. 3.Pure tone average measured preoperatively and 1 month postoperatively. There was no significant difference between mean preoperative (Preop BC) and postoperative bone conduction thresholds (Postop BC 1 mo) (P=0.444). A comparison between preoperative (Preop AC) and 1-month postoperative air conduction thresholds (Postop AC 1 mo) revealed no significant difference (P=0.232). AC, air conduction; BC, bone conduction.
Comparison between the preoperative pure tone audiometry threshold and VSB-aided thresholds at different frequencies (n=27)
| Frequency (Hz) | Preop PT (dB HL) | VSB 1 mo (dB HL) | VSB 3 mo (dB HL) | VSB 6 mo (dB HL) | |
|---|---|---|---|---|---|
| 250 | 71.7±14.0 | 53.7±15.4 | 55.0±14.0 | 51.5±19.6 | <0.001 |
| 500 | 73.7±16.3 | 50.2±15.9 | 46.9±14.9 | 46.7±16.2 | <0.001 |
| 1,000 | 77.2±17.3 | 39.1±11.6 | 38.1±9.1 | 36.5±11.0 | <0.001 |
| 2,000 | 70.0±14.6 | 35.6±9.5 | 33.3±7.7 | 34.8±10.9 | <0.001 |
| 3,000 | 75.6±17.4 | 48.1±15.7 | 44.1±15.1 | 42.2±14.8 | <0.001 |
| 4,000 | 81.9±19.5 | 62.0±17.9 | 54.3±16.3 | 50.6±15.1 | <0.001 |
Values are presented as mean±standard deviation.
VSB, Vibrant Soundbridge; Preop PT, preoperative pure tone audiometry threshold; VSB 1 mo, VSB-aided thresholds at 1 month after surgery; VSB 3 mo, VSB-aided thresholds at 3 months after surgery; VSB 6 mo, VSB-aided thresholds at 6 months after surgery.
Fig. 4.Mean sound field thresholds for ears treated with round window vibroplasty, assessed using warble tones at frequencies between 250 Hz and 4k Hz. (A) Comparison between preoperative unaided thresholds and postoperative VSB-aided thresholds (n=27). (B) Comparison between preoperative HA-aided thresholds and postoperative VSB-aided thresholds (n=16). AC, air conduction; Preop AC, preoperative unaided thresholds; VSB, Vibrant Soundbridge; VSB 1 mo, VSB-aided thresholds at 1 month after surgery; VSB 3 mo, VSB-aided thresholds at 3 months after surgery; VSB 6 mo, VSB-aided thresholds at 6 months after surgery; HA, hearing aids; HA (preop), preoperative HA-aided thresholds.
Comparison between the HA-aided threshold and VSB-aided thresholds at different frequencies (n=16)
| Frequency (Hz) | Preop HA (dB HL) | VSB 1 mo (dB HL) | VSB 3 mo (dB HL) | VSB 6 mo (dB HL) | |
|---|---|---|---|---|---|
| 250 | 54.0±13.8 | 59.4±14.8 | 58.8±13.1 | 54.7±21.3 | 0.714 |
| 500 | 53.3±15.0 | 55.0±15.6 | 49.1±16.7 | 49.4±19.4 | 0.692 |
| 1,000 | 48.0±11.6 | 40.9±9.9 | 39.4±10.5 | 39.4±12.1 | 0.068 |
| 2,000 | 50.0±12.7 | 37.5±9.1 | 34.4±8.1 | 36.9±12.2 | 0.002 |
| 3,000 | 51.0±10.9 | 51.3±13.6 | 46.6±16.4 | 42.8±15.9 | 0.307 |
| 4,000 | 57.9±10.3 | 64.4±18.2 | 55.0±18.2 | 49.1±15.6 | 0.068 |
Values are presented as mean±standard deviation.
HA, hearing aids; VSB, Vibrant Soundbridge-aided threshold; Preop HA, preoperative hearing aid-aided threshold; VSB 1 mo, VSB-aided thresholds at 1 month after surgery; VSB 3 mo, VSB-aided thresholds at 3 months after surgery; VSB 6 mo, VSB-aided thresholds at 6 months after surgery.