| Literature DB >> 30284273 |
Faris F Brkic1, Dominik Riss1, Alice Auinger1, Barbara Zoerner1, Christoph Arnoldner1, Wolf-Dieter Baumgartner1, Wolfgang Gstoettner1, Erich Vyskocil1.
Abstract
OBJECTIVE: To assess the audiological and long-term medical and technical follow-up outcomes of an active middle ear implant.Entities:
Keywords: Active middle ear implant; hearing loss; long-term outcome
Mesh:
Year: 2018 PMID: 30284273 PMCID: PMC6585801 DOI: 10.1002/lary.27513
Source DB: PubMed Journal: Laryngoscope ISSN: 0023-852X Impact factor: 3.325
Patients Demographics and Characteristics.
| Hearing Loss | Mean Age, Years | Male/Female, n | Male/Female, % | FMT Coupling | Total | |
|---|---|---|---|---|---|---|
| Sensorineural | 50.4 | 41/36 | 53.2/46.8 | Long process | 73 | 77/65.3 |
| Short process | 4 | |||||
| Mixed and conductive | 43.9 | 22/19 | 53.7/46.3 | RW | 19 | 41/34.7 |
| OW | 8 | |||||
| Stapes | 9 | |||||
| Promontorium | 5 | |||||
| Total | 48.3 | 63/55 | 53.4/46.6 | 118/100 |
FMT = floating mass transducer; n = number of cases; OW = oval window; RW = round window.
Device Explantations and Revision Surgeries.
| Indication | Complication | Type of Surgery | n | Total | % |
|---|---|---|---|---|---|
| Medical | Infection | Explantation | 4 | 14 | 11.9 |
| Wire extrusion in the outer ear canal | Explantation | 3 | |||
| Pain complaints | Explantation | 3 | |||
| FMT displacement | FMT repositioning | 4 | |||
| Technical | Device failure | Explantation | 2 | 5 | 4.2 |
| Device failure | Reimplantation | 2 | |||
| Poor benefit (unknown reason) | Explantation | 1 | |||
| Progressive hearing loss | Explantation/CI implantation | 13 | 13 | 11 | |
| No second surgery | 86 | 86 | 72.9 | ||
| Total | 118 | 100 |
CI = cochlear implant; FMT = floating mass transducer; n = number of cases.
Figure 1Percentage of Vibrant Soundbridge (MED‐EL, Innsbruck, Austria) implantations requiring an explantation or revision per 3 years.
Complications for Individual FMT Coupling Approaches.
| Hearing loss | FMT coupling | Complications, n | Total | Complications, % |
|---|---|---|---|---|
| Sensorineural | Long process | 7 | 73 | 9.6 |
| Short process | 0 | 4 | 0 | |
| Mixed | RW | 10 | 19 | 52.6 |
| OW | 2 | 8 | 25 | |
| Stapes | 0 | 9 | 0 | |
| Promontorium | 0 | 5 | 0 |
FMT = floating mass transducer; OW = oval window; RW = round window.
Figure 2Kaplan‐Meier implant survival graph. m, months; VSB = Vibrant Soundbridge (MED‐EL, Innsbruck, Austria). [Color figure can be viewed in the online issue, which is available at www.laryngoscope.com.]
Mean FHG in Free‐Field Audiometry in Different Hearing Loss Types.
| Type of Hearing Loss | Mean FHG (dB) |
|
|---|---|---|
| Sensorineural | 13.1 ± 8.9 | < .0001 |
| Mixed/conductive | 20.9 ± 8.8 | < .0001 |
dB = decibel; FHG = functional hearing gain.
Figure 3Mean postoperative unaided and thresholds with the VSB in free‐field audiometry across frequencies in patients with sensorineural hearing loss. unaided: postoperative thresholds without the VSB; VSB: postoperative thresholds with the VSB. dB = decibel; Hz = Hertz; VSB = Vibrant Soundbridge (MED‐EL, Innsbruck, Austria). [Color figure can be viewed in the online issue, which is available at www.laryngoscope.com.]
Figure 4Mean postoperative unaided and thresholds with the VSB in free‐field audiometry across frequencies in patients with mixed/conductive hearing loss. unaided: postoperative thresholds without the VSB; VSB: postoperative thresholds with the VSB. dB = decibel; Hz = Hertz; m = months; VSB = Vibrant Soundbridge (MED‐EL, Innsbruck, Austria). [Color figure can be viewed in the online issue, which is available at www.laryngoscope.com.]