Literature DB >> 27023016

Active Middle Ear Implantation: Long-term Medical and Technical Follow-up, Implant Survival, and Complications.

Joost W Zwartenkot1, Jef J S Mulder, Ad F M Snik, Cor W R J Cremers, Emmanuel A M Mylanus.   

Abstract

OBJECTIVE: To evaluate the long-term medical and technical results, implant survival, and complications of the semi-implantable vibrant soundbridge (VSB), otologics middle ear transducer (MET), and the otologics fully implantable ossicular stimulator (FIMOS). STUDY
DESIGN: Retrospective cohort study. PATIENTS: Patients with chronic external otitis and either moderate to severe sensorineural or conductive/mixed hearing loss.
SETTING: Tertiary referral center. INTERVENTION: Implantation with the VSB, MET, or FIMOS. MAIN OUTCOME MEASURES: Medical complications, number of reimplantations, and explantations.
RESULTS: Ninety-four patients were implanted, 12 patients with a round window or stapes application. 28 patients were lost to follow-up. The average follow-up duration was 4.4 years (range, 1 month-15 years). 128 devices were evaluated: (92 VSB, 32 MET, 4 FIMOS). 36 devices (28%) have been explanted or replaced (18 VSB, 14 MET, 4 FIMOS). Device failure was 7% for VSB, 28% for MET, and 100% for FIMOS. In 16 patients (17%) revision surgery (n = 20) was performed. Twenty patients (21%) suffered any medical complication.
CONCLUSION: Medical and technical complications and device failures have mostly occurred in the initial period of active middle ear implants (AMEI) implementation and during clinical trials or experimental procedures. All four FIMOS had technical difficulties. An important decrease in the occurrence of both medical and technical complications was observed. Application in more recent years did not show any complications and the recent device failure rates are acceptable. Magnetic resonance imaging (MRI) incompatibility should be taken into account when indicating AMEI.

Entities:  

Mesh:

Year:  2016        PMID: 27023016     DOI: 10.1097/MAO.0000000000001015

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  15 in total

1.  Complications after round window vibroplasty.

Authors:  Maurizio Barbara; Luigi Volpini; Edoardo Covelli; Martina Romeo; Chiara Filippi; Simonetta Monini
Journal:  Eur Arch Otorhinolaryngol       Date:  2019-03-27       Impact factor: 2.503

2.  Guideline "Implantable hearing aids"-short version : German S2k guideline of the Working Group of German-Speaking Audiologists, Neurootologists and Otologists (ADANO), of the German Society of Oto-Rhino-Laryngology, Head and Neck Surgery (DGHNO) in collaboration with the German Society of Audiology (DGA), the German Society of Phoniatrics and Pediatric Audiology (DGPP), and patient representatives.

Authors:  D Beutner; W Delb; H Frenzel; U Hoppe; K B Hüttenbrink; R Mlynski; A Limberger; R Schönweiler; B Schwab; I Todt; M Walger; T Wesarg; T Zahnert; R Zeh
Journal:  HNO       Date:  2018-07       Impact factor: 1.284

3.  [Guideline "Implantable hearing aids"-short version : German S2k guideline of the Working Group of German-speaking Audiologists, Neurootologists and Otologists (ADANO), of the German Society of Oto-Rhino-Laryngology, Head and Neck Surgery (DGHNO) in collaboration with the German Society of Audiology (DGA), the German Society of Phoniatrics and Pediatric Audiology (DGPP), and patient representatives].

Authors:  D Beutner; W Delb; H Frenzel; U Hoppe; K B Hüttenbrink; R Mlynski; A Limberger; R Schönweiler; B Schwab; I Todt; M Walger; T Wesarg; T Zahnert; R Zeh
Journal:  HNO       Date:  2018-09       Impact factor: 1.284

4.  Minimal Reporting Standards for Active Middle Ear Hearing Implants.

Authors:  Hannes Maier; Uwe Baumann; Wolf-Dieter Baumgartner; Dirk Beutner; Marco D Caversaccio; Thomas Keintzel; Martin Kompis; Thomas Lenarz; Astrid Magele; Torsten Mewes; Alexander Müller; Tobias Rader; Torsten Rahne; Sebastian P Schraven; Burkard Schwab; Georg Mathias Sprinzl; Bernd Strauchmann; Ingo Todt; Thomas Wesarg; Barbara Wollenberg; Stefan K Plontke
Journal:  Audiol Neurootol       Date:  2018-09-07       Impact factor: 1.854

5.  Air- and Bone-Conducted Sources of Feedback With an Active Middle Ear Implant.

Authors:  Renee M Banakis Hartl; James R Easter; Mohamed A Alhussaini; Daniel J Tollin; Herman A Jenkins
Journal:  Ear Hear       Date:  2019 May/Jun       Impact factor: 3.570

6.  Importance of adhesiolysis in revision surgery for vibrant soundbridge device failures at the short incus process.

Authors:  Yoon Ah Park; Tae Hoon Kong; Jin Soon Chang; Young Joon Seo
Journal:  Eur Arch Otorhinolaryngol       Date:  2017-08-22       Impact factor: 2.503

7.  The first results of a totally implanted active middle ear device.

Authors:  Maria da Conceição Peixoto; Cristina Miranda; Mafalda Bento; Susana Oliveira; Rui Pratas; Victor Correia da Silva
Journal:  Eur Arch Otorhinolaryngol       Date:  2019-07-24       Impact factor: 2.503

8.  The development of active middle ear implants: A historical perspective and clinical outcomes.

Authors:  Melodi Koşaner Kließ; Arne Ernst; Jan Wagner; Philipp Mittmann
Journal:  Laryngoscope Investig Otolaryngol       Date:  2018-10-06

9.  Middle Ear Transducer: Long Term Stability of the Latest Generation T2.

Authors:  Nils Kristian Prenzler; Eugen Kludt; Thomas Giere; Rolf Salcher; Thomas Lenarz; Hannes Maier
Journal:  Biomed Res Int       Date:  2019-01-06       Impact factor: 3.411

10.  Medical, Technical and Audiological Outcomes of Hearing Rehabilitation with the Bonebridge Transcutaneous Bone-Conduction Implant: A Single-Center Experience.

Authors:  Faris F Brkic; Dominik Riss; Katharina Scheuba; Christoph Arnoldner; Wolfgang Gstöttner; Wolf-Dieter Baumgartner; Erich Vyskocil
Journal:  J Clin Med       Date:  2019-10-03       Impact factor: 4.241

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.