| Literature DB >> 33623360 |
Javier Mendicute1, Lucia Bascarán1, Luis Pablo2, Cédric Schweitzer3, Laurent Velasque4, Christine Bouchet5, Aldo A Martinez5.
Abstract
PURPOSE: To evaluate intraoperative intraocular lens (IOL) delivery time and total surgical case time using the UltraSert preloaded delivery system (System U) during routine cataract surgeries and to compare with the manually loaded Monarch delivery system (System M). Physician satisfaction with System U was also assessed. PATIENTS AND METHODS: In this prospective observational study, subjects ≥18 years old underwent cataract surgery in 1 eye and received the AcrySof IQ IOL via the manually loaded System M (n=103) or the AcrySof IQ IOL model AU00T0 via the preloaded System U (n=93). Procedures were digitally recorded by an external camera or by a camera within the operating microscope. Device preparation, IOL delivery, and IOL positioning times were evaluated by 2 independent graders. Pearson χ2 test or Fisher exact test was used for categorical variables and Student's t-test or Wilcoxon rank-sum test for continuous variables (all tests were 2-sided and performed at a 5% α-level). Physician satisfaction levels were assessed using questionnaires.Entities:
Keywords: Monarch; UltraSert; cataract surgery; intraocular lens delivery time
Year: 2021 PMID: 33623360 PMCID: PMC7896803 DOI: 10.2147/OPTH.S263658
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Figure 1UltraSert preloaded IOL delivery system.
Figure 2Lens delivery process.
Figure 3Mean lens delivery times with System U and System M.
Figure 4Mean total intraoperative surgical times with System U and System M (device preparation + lens delivery).
Figure 5Comparison of System U and System M. (A) Device preparation time, lens delivery time, and lens unfolding time. (B) Total surgical case time includes device preparation, lens delivery, and lens positioning/unfolding.
Device Preparation Time with System U and System M
| System U, n=93 | System M, n=103 | |
|---|---|---|
| Mean device preparation time ± SD (range), s | 30.3±6.6 (20−60) | 59.8±31.0 (31−202) |
| Mean lens delivery time ± SD (range), s | 12.9±5.1 (4−28) | 12.2±6.3 (4−44) |
| Mean lens unfolding time ± SD (range), s | 13.8±7.0 (40−50) | 17.6±11.0 (4−58) |
| Mean total surgical case time | 56.6±12.6 (35−98) | 89.6±34.6 (41–232) |
Figure 6Physician satisfaction levels: *1=not satisfied at all; 7=extremely satisfied; †1=manual better, 5=neutral, 9=System U better; ‡1=other preloaded device better, 5=neutral, 9=System U better.