Comparative evaluation of corneal incision enlargement after intraocular lens delivery of new preloaded and manual implantation systems.

Lixin Wang1, Patricia Wolfe2, Sumit Paliwal2, Alexis Chernosky2, Thomas Kohnen3.   

Abstract

PURPOSE: To determine whether total corneal incision enlargement after implantation of an intraocular lens with a new preloaded delivery system is comparable to a standard-of-care manual delivery system using an in vitro human cadaver eye model, despite having a smaller initial incision size.
METHODS: Human cadaver phakic whole eye globes were used for these studies (n = 16 per group). Each pair of eyes was randomly assigned to a new preloaded delivery system (UltraSertTM) or a manual delivery system (MONARCH® III D). The surgical procedure included creating a 2.2- and 2.4-mm corneal incision for the preloaded and manual delivery systems, respectively, measuring intraocular pressure and pre-implantation corneal incision size, delivering the intraocular lens into the anterior chamber, and measuring the post-implantation incision size.
RESULTS: The final corneal incision enlargement after intraocular lens delivery using the preloaded delivery system was 2.33 ± 0.04 mm, compared to 2.54 ± 0.05 mm after intraocular lens delivery with the manual delivery system. The mean corneal incision enlargement was comparable between the two systems, being 0.13 ± 0.04 mm using the preloaded delivery system and 0.14 ± 0.05 mm using the manual delivery system (p = 0.432).
CONCLUSION: In a human cadaver eye model, the preloaded delivery system demonstrated an intraocular lens delivery performance on cornea incision enlargement was noninferior to the manual, standard-of-care intraocular lens delivery system despite a smaller initial incision size. TRANSLATIONAL RELEVANCE: Smaller incision sizes for cataract surgery improve patient outcomes via faster visual and wound recovery and decreased risk of complications such as postoperative inflammation and surgically induced astigmatism.

Entities:  

Keywords:  Intraocular lens implantation; cataract surgery; corneal incision size; intraocular lens delivery; preloaded intraocular lens delivery system

Mesh:

Year:  2019        PMID: 31718281     DOI: 10.1177/1120672119882334

Source DB:  PubMed          Journal:  Eur J Ophthalmol        ISSN: 1120-6721            Impact factor:   2.597


  2 in total

1.  Clinical Evaluation of a Novel Preloaded Intraocular Lens Delivery System During Routine Cataract Surgery.

Authors:  Daniel Black; Dean Corbett; Timothy V Roberts; Brendan Cronin; Pamela J Smith; D Priya Janakiraman; Beth E Jackson
Journal:  Clin Ophthalmol       Date:  2020-08-12

2.  Multicenter Evaluation of Time, Operational, and Economic Efficiencies of a New Preloaded Intraocular Lens Delivery System versus Manual Intraocular Lens Delivery.

Authors:  Javier Mendicute; Lucia Bascarán; Luis Pablo; Cédric Schweitzer; Laurent Velasque; Christine Bouchet; Aldo A Martinez
Journal:  Clin Ophthalmol       Date:  2021-02-16
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.