In vitro delivery performance assessment of a new preloaded intraocular lens delivery system.

Lixin Wang1, Patricia Wolfe2, Alexis Chernosky2, Sumit Paliwal2, Khiun Tjia2, Stephen Lane2.   

Abstract

PURPOSE: To compare the delivery performance of a new preloaded intraocular lens (IOL) delivery system (Ultrasert [system U]) with that of 2 commercially available preloaded systems (iSert [system S] and Tecnis iTec [system T]), and a manual system (Monarch III D).
SETTING: Alcon Laboratories, Fort Worth, Texas, USA.
DESIGN: Experimental study.
METHODS: Freshly excised porcine eyes were randomly assigned to 5 groups of 10 eyes each as follows: preloaded systems U, S, T 2.2 mm incision, T 2.4 mm incision, and manual system. Corneal incision size was measured before and after delivery. The rate of successful in-the-bag IOL delivery, IOL adherence to the plunger tip, trapped trailing haptic, and presence of nozzle tip splitting were assessed. Statistical analysis was performed to determine the significance of differences between groups.
RESULTS: All systems successfully placed the IOL in the capsular bag. System U had the least corneal incision enlargement (mean 0.07 mm ± 0.05 [SD]) and the smallest final incision size (mean 2.31 ± 0.06 mm) (P < .05). An associated split in the nozzle tip was observed for system S devices (P < .05). Furthermore, in more than 50% of deliveries with system T, the IOL remained adhered to the plunger tip after exiting the nozzle (P < .05). Systems S and T each had 1 occurrence of a trapped trailing haptic during IOL delivery.
CONCLUSION: The new design of the system U with a depth-guard nozzle tip allowed for the smallest incision enlargement with an absence of nozzle tip splitting and uneventful IOL delivery compared with the other preloaded systems tested in this study. FINANCIAL DISCLOSURE: Drs. Tjia and Lane are consultants to Alcon Laboratories, Inc. Drs. Wang, Wolfe, Paliwal, and Miss Chernosky are employees of Alcon Laboratories, Inc.
Copyright © 2016 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2016        PMID: 28007114     DOI: 10.1016/j.jcrs.2016.10.014

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  6 in total

1.  Comparison of incision size and intraocular lens performance after implantation with three preloaded systems and one manual delivery system.

Authors:  Javier Mendicute; Thierry Amzallag; Lixin Wang; Aldo A Martinez
Journal:  Clin Ophthalmol       Date:  2018-08-21

2.  Clinical Evaluation of a Novel Preloaded Intraocular Lens Delivery System During Routine Cataract Surgery.

Authors:  Daniel Black; Dean Corbett; Timothy V Roberts; Brendan Cronin; Pamela J Smith; D Priya Janakiraman; Beth E Jackson
Journal:  Clin Ophthalmol       Date:  2020-08-12

3.  Experimental Study on Delivery Performance of an Automated Preloaded Intraocular Lens Injector System for Corneal and Sclerocorneal Incisions.

Authors:  Tetsuro Oshika; Noriyuki Sasaki
Journal:  J Ophthalmol       Date:  2021-03-31       Impact factor: 1.909

4.  A laboratory evaluation of nozzle tip damage in four generations of intraocular lens injector systems using a self-developed damage scale.

Authors:  Hui Fang; Lu Zhang; Sonja Schickhardt; Patrick R Merz; Weijia Yan; Mélanie Leroux; Gerd U Auffarth
Journal:  Sci Rep       Date:  2022-02-17       Impact factor: 4.379

5.  Multicenter Evaluation of Time, Operational, and Economic Efficiencies of a New Preloaded Intraocular Lens Delivery System versus Manual Intraocular Lens Delivery.

Authors:  Javier Mendicute; Lucia Bascarán; Luis Pablo; Cédric Schweitzer; Laurent Velasque; Christine Bouchet; Aldo A Martinez
Journal:  Clin Ophthalmol       Date:  2021-02-16

6.  Comparison of a new IOL injector system against 3 standard IOL injector systems with different incision sizes: Miyake-Apple view experimental laboratory study.

Authors:  Lu Zhang; Sonja Schickhardt; Hui Fang; Florian Auerbach; Perfecto Cagampang; Patrick R Merz; Gerd U Auffarth
Journal:  J Cataract Refract Surg       Date:  2022-02-01       Impact factor: 3.351

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.