Literature DB >> 18812128

Incision sizes before and after implantation of SN60WF intraocular lenses using the Monarch injector system with C and D cartridges.

Thomas Kohnen1, Oliver K Klaproth.   

Abstract

PURPOSE: To evaluate incision sizes for 6.0 mm optic hydrophobic acrylic intraocular lenses (IOLs) implanted with different injector systems and surgical techniques.
SETTING: Department of Ophthalmology, Johann Wolfgang Goethe-University, Frankfurt, Germany.
METHODS: In this prospective randomized trial of 100 cataract patients, hydrophobic acrylic IOLs (SN60WF) were implanted using 2 injector systems with 2 cartridges (Monarch II and Monarch III with C and D cartridges, respectively; Alcon) and 2 surgical techniques (direct implantation and wound-assisted docking implantation). Incision sizes were measured before and after phacoemulsification as well as before and after IOL implantation using a Tsuneoka microincision gauge. The Saphiro-Wilks test was performed to detect nonparametric distribution and the Kruskal-Wallis test, to test for significant differences between groups.
RESULTS: The mean incision sizes (mm) at the 4 points of treatment with the Monarch III with the D cartridge were 1.98+/-0.07, 2.01+/-0.05, 2.42+/-0.07, and 2.51+/-0.05, respectively, with direct implantation and 1.92+/-0.06, 1.98+/-0.04, 2.02+/-0.11, and 2.09+/-0.11, respectively, with wound-assisted implantation. With the Monarch II with the C cartridge, the mean incision sizes were 2.17+/-0.06, 2.2+/-0.16, 2.77+/-0.11, and 2.86+/-0.11, respectively, with direct implantation and 2.12+/-0.10, 2.17+/-0.19, 2.19+/-0.09, and 2.28+/-0.08, respectively, with wound-assisted implantation. Postoperative incision sizes differed significantly from preoperative incision sizes and from each other (P<.05).
CONCLUSION: With new injector systems and an appropriate docking implantation technique, foldable hydrophobic acrylic IOLs can be implanted through 2.0 mm tunnel incisions.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18812128     DOI: 10.1016/j.jcrs.2008.06.031

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  12 in total

1.  Comparison of clear corneal incision injuries between torsional and conventional phacoemulsification.

Authors:  Xu Chen; Yinghong Ji; Yi Lu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-07-20       Impact factor: 3.117

Review 2.  [Incisions for biaxial and coaxial microincision cataract surgery].

Authors:  M Müller; T Kohnen
Journal:  Ophthalmologe       Date:  2010-02       Impact factor: 1.059

3.  [Intraocular lenses for microincisional cataract surgery].

Authors:  T Kohnen; O K Klaproth
Journal:  Ophthalmologe       Date:  2010-02       Impact factor: 1.059

4.  [Technologies, techniques and tactics for micro-incision cataract surgery].

Authors:  E Fabian; M Maier
Journal:  Ophthalmologe       Date:  2010-02       Impact factor: 1.059

5.  Safety of hydroimplantation in cataract surgery in patients with pseudoexfoliation syndrome.

Authors:  Tevfik Oğurel; Reyhan Oğurel; Zafer Onaran; Kemal Örnek
Journal:  Int J Ophthalmol       Date:  2017-05-18       Impact factor: 1.779

6.  In vivo and in vitro results of an automated preloaded delivery system for IOL implantation in cataract surgery.

Authors:  Leonardo Mastropasqua; Lisa Toto; Erminia D'Ugo; Manuela Lanzini; Peter A Mattei; Gennaro Falconio; Emanuele Doronzo; Francesca Diomede; Oriana Trubiani
Journal:  Int Ophthalmol       Date:  2019-08-26       Impact factor: 2.031

Review 7.  [The basics of refractive lens exchange].

Authors:  T Kohnen; O K Klaproth
Journal:  Ophthalmologe       Date:  2008-11       Impact factor: 1.059

Review 8.  Cataract surgery with implantation of an artificial lens.

Authors:  Thomas Kohnen; Martin Baumeister; Daniel Kook; Oliver K Klaproth; Christian Ohrloff
Journal:  Dtsch Arztebl Int       Date:  2009-10-23       Impact factor: 5.594

9.  Comparison of new motorized injector vs manual injector for implantation of foldable intraocular lenses on wound integrity: an ASOCT study.

Authors:  S Khokhar; R Sharma; B Patil; N Aron; S Gupta
Journal:  Eye (Lond)       Date:  2014-07-18       Impact factor: 3.775

10.  Wound Integrity of 2.0 mm Transconjunctival Single-Plane Sclerocorneal Incision: A Comparison between a Motorized Injector with/without Pause Time and a Manual Injector.

Authors:  Hisanori Imai; Ayaka Fujii; Emiko Tani; Atsushi Azumi
Journal:  J Ophthalmol       Date:  2016-02-17       Impact factor: 1.909

View more