Anchal Sahni1, Rishabh Kapoor1, Kapil Gandhi1, Dipanshu Kumar1, Geetika Datta1, Ritika Malhotra1.
Abstract
AIM: Pulpectomy in the primary tooth has unique challenges due to morphological variations in root pattern and physiological root resorption. The electronic apex locator (EAL) is one of the recent methods to determine the estimation of working length without much radiation exposure to the patient as well as the operator. The present study was undertaken for evaluating and comparing the efficacy of EAL, conventional radiography, digital radiography, and actual visual method for the estimation of in the root canal working length (RCL) in extracted primary teeth.Entities:
Keywords: Digital radiograph; Electronic apex locator; Primary tooth; Working length
Year: 2020 PMID: 33623342 PMCID: PMC7887167 DOI: 10.5005/jp-journals-10005-1821
Source DB: PubMed Journal: Int J Clin Pediatr Dent ISSN: 0974-7052
Fig. 1Working length estimation with electronic apex locator
Fig. 2File length measurement using a digital Vernier caliper
Fig. 3Jig design
Fig. 4Working length estimation with digital radiograph
Fig. 5Working length estimation with conventional radiograph using the jig
Sample distribution
| Method I | Electronic apex locators (EALs) | 90 |
| Method II | Digital radiography (DR) | |
| Method III | Conventional radiograph (IOPAR) | |
| Control method | Actual visual method (AVM) |
Comparison of mean root length measurements among methods
| Electronic apex locator | 10.10 | 1.78 | 13.83 | 5.23 | 0.426 | 0.235 |
| Digital radiograph | 10.08 | 2.10 | 14.80 | 5.62 | ||
| Conventional radiograph | 9.78 | 1.81 | 13.70 | 5.41 | ||
| Actual visual method | 10.36 | 1.80 | 14.21 | 5.31 |
Fig. 6Comparison of root length measurements among groups
Intergroup comparison of root canal measurement
| Apex locator — digital radiograph | 10.104 ± 1.782 | 3.099 | 0.066 |
| Apex locator — IOPAR | 10.07 ± 1.752 | 10.329 | 0.085 |
| Digital radiograph — IOPAR | 10.08 ± 2.107 | 2.458 | 0.07 |
Fig. 7Intergroup comparison among different groups
Correlation of electronic apex locator with actual working length
| 0.999 | 0.997 | 0.997 | 0.53072 |
Correlation of digital radiography with actual working length
| 0.991 | 0.981 | 0.981 | 1.44607 |
Correlation of conventional radiography with actual working length
| 0.990 | 0.961 | 0.961 | 0.6260 |