Evaluation of the Root ZX apex locator in primary teeth.

Sara Ghaemmaghami1, Jason Eberle, Donald Duperon.   

Abstract

PURPOSE: The purpose of this study was to evaluate the accuracy of the Root ZX electronic apex locator in determining canal length in primary incisors.
METHODS: Root ZX was utilized to measure the canal lengths of 150 primary incisors in vivo. After the teeth were extracted, a standard ruler was used to measure the canal lengths to the nearest 0.5 mm. A paired t test was used to compare the lengths obtained in vivo with Root ZX and the lengths measured in vitro.
RESULTS: In 98 (65%) of the canals, the length measured by Root ZX was determined to be equal to the length measured in vitro. Root ZX measurements were found to be less than in vitro measurements by 0.5 mm in 24 canals (16%), and were greater than in vitro measurements by 0.5 mm in 21 canals (14%). Root ZX measurements exceeded in vitro measurements by 1.0 to 1.5 mm in 5 canals (3%) and were less than in vitro measurements by 1.0 mm in 2 canals (1%). Paired t test analysis indicated no significant difference between the 2 measurement methods (P=.67).
CONCLUSIONS: Root ZX can accurately assess the length of canals in primary incisors.

Mesh:

Year:  2008        PMID: 19186775

Source DB:  PubMed          Journal:  Pediatr Dent        ISSN: 0164-1263            Impact factor:   1.874


  6 in total

1.  Clinical evaluation of Root ZX II electronic apex locator in primary teeth.

Authors:  S Saritha; K S Uloopi; C Vinay; R Chandra Sekhar; V V Rao
Journal:  Eur Arch Paediatr Dent       Date:  2012-02

2.  Formulating a regression equation for determination of working length in primary molars using apex locators: a clinical study.

Authors:  A Dandempally; R Muppa; M K Duddu; P Bhupatiraju; S Nallanchakrava
Journal:  Eur Arch Paediatr Dent       Date:  2013-06-18

3.  Comparison of Accuracy of Apex Locator with Tactile and Conventional Radiographic Method for Working Length Determination in Primary and Permanent Teeth.

Authors:  Khushboo Rathore; Sandeep Tandon; Meenakshi Sharma; Garima Kalia; Tejshree Shekhawat; Yogendra Chundawat
Journal:  Int J Clin Pediatr Dent       Date:  2020 May-Jun

4.  Accuracy of an Electronic Apex Locator for Working Length Determination in Primary Anterior Teeth.

Authors:  Zahra Bahrololoomi; Ali Asghar Soleymani; Jalil Modaresi; Mahsa Imanian; Malihe Lotfian
Journal:  J Dent (Tehran)       Date:  2015-04

5.  A comparative evaluation of electronic and radiographic determination of root canal length in primary teeth: An in vitro study.

Authors:  Iyer Satishkumar Krishnan; Sheela Sreedharan
Journal:  Contemp Clin Dent       Date:  2012-10

6.  A Comparative Evaluation of Efficacy of Electronic Apex Locator, Digital Radiography, and Conventional Radiographic Method for Root Canal Working Length Determination in Primary Teeth: An In Vitro Study.

Authors:  Anchal Sahni; Rishabh Kapoor; Kapil Gandhi; Dipanshu Kumar; Geetika Datta; Ritika Malhotra
Journal:  Int J Clin Pediatr Dent       Date:  2020 Sep-Oct
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.