Clinical evaluation of the measuring accuracy of ROOT ZX in primary teeth.

Andrej M Kielbassa1, Ursula Muller, Inka Munz, Jürgen Schulte Monting.   

Abstract

OBJECTIVE: The aim of this study was to assess the accuracy of an electronic device (Root ZX; Morita, Tokyo, Japan) for measuring the root canal length in primary teeth. STUDY
DESIGN: The pulp tissue was removed from 71 nonrestorable teeth scheduled to be extracted under general anesthesia, and the root canals (n = 105) were irrigated (H(2)O(2), 3%; NaOCl, 1%). Subsequently, the length was determined clinically with the electronic device prior to extraction. Treatments were performed by 2 dentists (42 and 63 root canals). After extraction, the real length was recorded and the 2 measurements were compared.
RESULTS: Measurements were affected significantly by the dentists (P < .01; chi(2)). However, regression analysis revealed sufficient accuracy of the device, with a tendency to estimate the root canal length just short (x = -0.98 +/- 1.75 mm) of the apex. These results were not influenced by tooth type, root canal type, status of the periapex, or clinical condition (P > .05; chi(2)).
CONCLUSION: Root ZX can be strongly recommended for clinical implementation of endodontics in primary teeth, particularly when treating fidgety children.

Entities:  

Mesh:

Year:  2003        PMID: 12539034     DOI: 10.1067/moe.2003.99

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod        ISSN: 1079-2104


  17 in total

1.  Clinical evaluation of Root ZX II electronic apex locator in primary teeth.

Authors:  S Saritha; K S Uloopi; C Vinay; R Chandra Sekhar; V V Rao
Journal:  Eur Arch Paediatr Dent       Date:  2012-02

2.  Accuracy of two different apex locators in primary teeth with and without root resorption.

Authors:  Haluk Bodur; Mesut Odabaş; Ozlem Tulunoğlu; Ali Cemal Tinaz
Journal:  Clin Oral Investig       Date:  2007-11-20       Impact factor: 3.573

3.  Evaluation of the multifrequency electronic apex locator Joypex 5 in primary teeth.

Authors:  E J N L Silva; D R Herrera; E J Souza-Júnior; T P Rosa
Journal:  Eur Arch Paediatr Dent       Date:  2013-07-24

4.  In vitro comparison between apex locators, direct and radiographic techniques for determining the root canal length in primary teeth.

Authors:  L Z Adriano; J C Barasuol; M Cardoso; M Bolan
Journal:  Eur Arch Paediatr Dent       Date:  2019-01-02

Review 5.  [Advanced research on root canal therapy for primary teeth].

Authors:  Yi Yu; Xin Zhou; Li-Wei Zheng
Journal:  Hua Xi Kou Qiang Yi Xue Za Zhi       Date:  2020-04-01

6.  Digital radiography for determination of primary tooth length: in vivo and ex vivo studies.

Authors:  Maria D Basso; Fabiano Jeremias; Rita C L Cordeiro; Lourdes Santos-Pinto
Journal:  ScientificWorldJournal       Date:  2015-02-23

7.  Accuracy of an Electronic Apex Locator for Working Length Determination in Primary Anterior Teeth.

Authors:  Zahra Bahrololoomi; Ali Asghar Soleymani; Jalil Modaresi; Mahsa Imanian; Malihe Lotfian
Journal:  J Dent (Tehran)       Date:  2015-04

8.  Use of artificial primary teeth for endodontic laboratory research: experiments related to canal length determination.

Authors:  Anna Carolina V Mello-Moura; Carmela R Bresolin; Cacio Moura-Netto; André Ito; Angela T Araki; José Carlos P Imparato; Fausto M Mendes
Journal:  BMC Oral Health       Date:  2017-11-17       Impact factor: 2.757

9.  A comparative evaluation of electronic and radiographic determination of root canal length in primary teeth: An in vitro study.

Authors:  Iyer Satishkumar Krishnan; Sheela Sreedharan
Journal:  Contemp Clin Dent       Date:  2012-10

10.  An in vitro comparison of root canal length determination by DentaPort ZX and iPex apex locators.

Authors:  Nikhil Puri; Rupali Chadha; Pragya Kumar; Komal Puri
Journal:  J Conserv Dent       Date:  2013-11
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.