Accuracy of a new apex locator: an in vitro study.

A Y Kaufman1, S Keila, M Yoshpe.   

Abstract

AIM: The purpose of this study was to test in an in vitro model the accuracy of a Bingo 1020 electronic apex locator, to compare the results to those of a well known apex locator, Root ZX, as well as to those of the radiographic method of tooth length determination.
METHODOLOGY: A total of 120 extracted teeth, preserved in Thymol solution and kept refrigerated, was used for the study. The experiment was performed on single-rooted teeth and on one-root canal, chosen randomly, in multirooted teeth. The teeth were randomly divided into 12 groups of 10 teeth each. After access preparation, the actual length (AL) was measured. The teeth were embedded in an alginate model specially developed for testing apex locators. Electronic tooth length measurements (EL) were carried out prior to root canal preparation using the two electronic apex locators (EAL) - Root ZX and Bingo 1020; three measurements were taken and an average computed. After the third measurement, the file was left in the root canal and a periapical radiograph was taken. The radiographic length (RL) was recorded by measuring the file length from the coronal reference point to the tip of the file. Each root canal was then prepared to a no. 40 K-file diameter using a standardized technique; saline was used for irrigation. Upon completion of the root canal preparation, EL measurements were taken by each EAL in dry conditions and with different irrigation solutions. Each measurement was repeated three times. The RL was recorded according to the last EL measurement. Results were subjected to statistical analysis.
RESULTS: In all parameters tested, a significant statistical difference was found between Bingo 1020 and the Root ZX. Measurements obtained using the Bingo 1020 were consistently closer to the AL (0.08 mm) than those obtained using the Root ZX. Both EALs measured the tooth length with great accuracy and a positive correlation of 0.76 (P = 0.00) existed between the two devices. No significant difference was found between the two apex locators when measurements were taken with the different irrigants (P = 0.34) and the content of the root canal did not affect the accuracy of the measurements. Lengths obtained by calculations from the radiographs were longer than the AL as well as the length obtained by both EALs (P = 0.00).
CONCLUSIONS: The Bingo 1020 proved to be as reliable as Root ZX and was user friendly. Under the experimental conditions, electronic measurements were more reliable than radiographs in the process of root length determination.

Entities:  

Mesh:

Year:  2002        PMID: 11843975     DOI: 10.1046/j.1365-2591.2002.00468.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  26 in total

1.  Clinical evaluation of Root ZX II electronic apex locator in primary teeth.

Authors:  S Saritha; K S Uloopi; C Vinay; R Chandra Sekhar; V V Rao
Journal:  Eur Arch Paediatr Dent       Date:  2012-02

2.  Accuracy of two different apex locators in primary teeth with and without root resorption.

Authors:  Haluk Bodur; Mesut Odabaş; Ozlem Tulunoğlu; Ali Cemal Tinaz
Journal:  Clin Oral Investig       Date:  2007-11-20       Impact factor: 3.573

3.  Electronic root canal length measurement before and after experimentally induced pulpitis and apical periodontitis in dogs.

Authors:  Maja Kovacevic; Tomislav Tamarut; Snjezana Glavicić; Nives Jonjic; Sanja Zoricić-Cvek; Dragica Bobinac
Journal:  Med Biol Eng Comput       Date:  2006-07-13       Impact factor: 2.602

4.  An ex vivo comparison of working length determination by three electronic root canal length measurement devices integrated into endodontic rotary motors.

Authors:  Manal Maree Ali; Ronald Wigler; Shaul Lin; Arieh Y Kaufman
Journal:  Clin Oral Investig       Date:  2016-07-09       Impact factor: 3.573

5.  In vitro evaluation of the accuracy of two electronic apex locators.

Authors:  Viresh Chopra; Shibani Grover; S Datta Prasad
Journal:  J Conserv Dent       Date:  2008-04

6.  Comparative evaluation of accuracy of two electronic apex locators in the presence of various irrigants: An in vitro study.

Authors:  Saru Jain; Ravi Kapur
Journal:  Contemp Clin Dent       Date:  2012-09

7.  Accuracy of Root ZX in teeth with simulated root perforation in the presence of gel or liquid type endodontic irrigant.

Authors:  Hyeong-Soon Shin; Won-Kyung Yang; Mi-Ri Kim; Hyun-Jung Ko; Kyung-Mo Cho; Se-Hee Park; Jin-Woo Kim
Journal:  Restor Dent Endod       Date:  2012-08-29

8.  In vitro detection of simulated apical root perforation with two electronic apex locators.

Authors:  Kiumars Nazari Moghaddam; Shahrzad Nazari; Leila Shakeri; Kiamars Honardar; Farshid Mirmotalebi
Journal:  Iran Endod J       Date:  2010-02-20

9.  Comparison Between the Accuracy of NovApex Apex Locator and Radiographs in Determining Radiographic Apex.

Authors:  Shahriar Shahi; Saeed Rahimi; Amin Salem Milani; Mohammad Asghari Jafarabadi; Gholam Reza Djoorabbaf Naghsh
Journal:  Iran Endod J       Date:  2011-05-15

10.  A comparative evaluation of electronic and radiographic determination of root canal length in primary teeth: An in vitro study.

Authors:  Iyer Satishkumar Krishnan; Sheela Sreedharan
Journal:  Contemp Clin Dent       Date:  2012-10
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.